| Brand | ANJIE |
|---|---|
| Origin | Shanghai, China |
| Model | AJ-500 / AJ-900 |
| Sample Tray Configurations | 90-position (86 × 15 mL + 4 × 50 mL) or 51-position (51 × 50 mL) |
| Detector Interface | Photomultiplier Tube (PMT) Compatible |
| Brand | ANJIE |
|---|---|
| Origin | Shanghai, China |
| Model | AJ-5100 |
| Measurement Principle | Potassium Permanganate Oxidation with ORP Endpoint Detection |
| Instrument Type | Benchtop Laboratory Analyzer |
| Range | 0–4.5 mg/L as O₂ |
| Repeatability | RSD ≤ 5.0% |
| Accuracy | ±10% of reading |
| Limit of Detection | < 0.07 mg/L |
| Resolution | 1 µL |
| Sample Analysis Time | 3 min 50 s per sample |
| Heating Method | Infrared Rapid Boiling (reaches boiling within 100 s) |
| Reaction Control | User-defined oxidation holding time |
| Fluidic Architecture | Dedicated reagent addition station, titration analysis station, and waste discharge station |
| Dispensing System | Single-arm precision liquid handler for all reagents |
| Brand | Anjie |
|---|---|
| Origin | Shanghai, China |
| Model | AJ-1000 |
| Product Type | Flow Injection Analyzer |
| Wavelength Range | 340–1100 nm |
| Sample Capacity | 162 positions |
| Reagent Positions | 3–8 |
| Flow Cell Path Length | 10 mm |
| Light Source | Tungsten-Halogen Lamp |
| Brand | ANJIE |
|---|---|
| Origin | Shanghai, China |
| Manufacturer Type | Authorized Distributor |
| Country of Origin | China |
| Model | AJ-200 (0102) |
| Pricing | Available Upon Request |
| Digestion Principle | UV-Catalyzed Ambient-Pressure Oxidation |
| Sample Throughput | ≤2.5 min per sample |
| Dilution Capability | Automated 1–40× dilution for high-concentration samples |
| Integration | Compatible with AJ-3700 series analyzers |
| Fluidic Architecture | Clamp-free peristaltic tubing system |
| Reagent Management | Modular reagent rack support included |
| Compliance | Designed to support ISO 11905-1, EPA Method 351.2, and GB/T 11894–1989 workflows |
| Brand | Anjie |
|---|---|
| Origin | Shanghai, China |
| Model | AJ-3000 Plus |
| Detector Type | Photomultiplier Tube (PMT) |
| Wavelength Range | 190–900 nm |
| Spectral Bandwidth | 0.1 nm, 0.2 nm, 0.4 nm, and 1.0 nm (selectable) |
| Wavelength Accuracy | ±0.2 nm |
| Wavelength Repeatability | 0.1 nm |
| Brand | ANJIE |
|---|---|
| Origin | Shanghai, China |
| Manufacturer Type | Authorized Distributor |
| Country of Manufacture | China |
| Model | AJ-600 |
| Pricing | Available Upon Request |
| Dispensing Volume Range | 50–1000 mL |
| Dispensing Accuracy | ≤ ±2% |
| Temperature Control Range | 4–60 °C |
| Temperature Stability | ±1 °C |
| Full-Volume Preparation Time | <15 min |
| Onboard Protocol Storage | ≥10 preloaded methods |
| Integrated Log System | Audit-trail-capable user activity logging |
| Touchscreen Interface | Color LCD with intuitive GUI |
| Built-in Functions | Automated stirring, temperature regulation, system self-cleaning, headspace liquid-path purging, method recall and execution |
| Brand | Anjie |
|---|---|
| Origin | Shanghai, China |
| Manufacturer Type | Authorized Distributor |
| Instrument Category | Domestic |
| Model | AJ-510 |
| Quotation | Upon Request |
| Detector Type | Photomultiplier Tube (PMT) |
| Wavelength Range | 190–900 nm |
| Spectral Bandwidth | 2.0 nm |
| Wavelength Accuracy | ±0.2 nm |
| Wavelength Repeatability | ±0.1 nm |
| Brand | ANJIE |
|---|---|
| Origin | Shanghai, China |
| Model | AJ-3400 Series |
| Detector Type | Photomultiplier Tube (PMT) |
| Wavelength Accuracy | ±0.2 nm |
| Wavelength Repeatability | 0.1 nm |
| Brand | Anjie |
|---|---|
| Origin | Shanghai, China |
| Model | AJ-6400 |
| Target Analytes | Thallium (Tl), Lithium (Li) |
| Response Time | ≤2 min |
| Power Supply | Integrated Rechargeable Li-ion Battery (≥12 h continuous operation) |
| Operating Pressure | Ambient Atmospheric Pressure |
| Sample Introduction | Direct Liquid Injection (No Pretreatment Required) |
| Data Storage Capacity | >2000 Records with Real-time Spectral Display |
| Connectivity | USB for Data Export & Firmware Update |
| Compliance | Designed for Field Deployment under ISO/IEC 17025-aligned QA/QC Practices |
| Brand | ANJIE |
|---|---|
| Origin | Shanghai, China |
| Model | AJ-3150 Plus |
| Instrument Type | Laboratory Analyzer |
| Measurement Principle | Gas-Phase Chemiluminescence (GPC) with Onboard Oxidation |
| Measurement Range | 0.08–100 mg/L NH₃–N |
| Accuracy | ≤ ±5% of reading |
| Detection Limit | 0.02 mg/L NH₃–N |
| Analysis Time per Sample | 1–3 minutes |
| Batch Capacity | 80 samples |
| Air Supply | Integrated Clean Air Generator (Atmospheric Pressure, No High-Pressure Cylinders Required) |
| Onboard Functions | Integrated Oxidation Module, Heating Module (for Nitrate & Total Nitrogen), Homogenizing Sparging System, Refrigerated Water Trap, Sintered-Glass Gas–Liquid Separator, Real-Time Graphical Monitoring, Auto-Dilution (≥40×), Adjustable Lamp Current & Automatic Light Gain Control |
| Brand | ANJIE |
|---|---|
| Origin | Shanghai, China |
| Model | APA-500 |
| Principle | Potassium Permanganate Oxidation with ORP Endpoint Detection |
| Type | Online Analyzer |
| Range | 0–5 mg/L |
| Repeatability | ±2% RSD |
| Accuracy | ±5% of reading |
| LOD | 0.05 mg/L |
| Resolution | 1 µL |
| Cycle Time | 4 min per sample |
| Temperature Control | 40–100 °C (adjustable water bath) |
| Environmental Monitoring | Integrated ambient temperature, humidity, and atmospheric pressure sensors |
| Sample Handling | Motorized adaptive gripper (pneumatic-free) |
| Endpoint Detection | RGB-spectral colorimetric endpoint simulation with real-time video recording |
| Compliance | Designed for ISO 8467:2021, EPA Method 4500-PERM, and GB/T 5750.7–2023 |
| Brand | Anjie |
|---|---|
| Origin | Shanghai, China |
| Model | AJ-5750 |
| Instrument Type | Laboratory Benchtop Analyzer |
| Measurement Principle | Rapid Digestion Method (Closed-Tube, Reflux-Based) |
| Measurement Range | 16–700 mg/L COD |
| Accuracy | ≤5% (vs. Certified Reference Materials) |
| Repeatability | RSD ≤2.0% |
| Analysis Time | 14 minutes per sample |
| Batch Capacity | ≥21 samples |
| Safety Design | Fully Enclosed Digestion Chamber with Spherical Condenser |
| Operational Alert | Real-Time Reagent Vacuum Loss Detection |
| Compliance Context | Designed to support ISO 6060, ASTM D1252–06 (Method B), and HJ 828–2017 (Chinese EPA Standard) |
| Brand | Anjie |
|---|---|
| Origin | Shanghai, China |
| Model | AJ-6400A |
| Detection Target | Thallium (Tl) in aqueous samples |
| Response Time | ≤2 min |
| Power Supply | Integrated rechargeable lithium battery (≥12 h continuous operation) |
| Data Storage | >2000 measurement records with real-time spectral display |
| Interface | Touchscreen GUI, USB for data export & firmware update |
| Operating Environment | Ambient atmospheric pressure, no external gas supply or AC power required |
| Sample Handling | Direct liquid injection, no pretreatment |
| Calibration | Built-in standard curves + on-site curve generation capability |
| Remote Diagnostics | Supported |
| Brand | ANJIE |
|---|---|
| Origin | Shanghai, China |
| Model | AJ-6100 |
| Instrument Type | Automated Iodine-Specific Elemental Analyzer for Urine, Water and Salt Matrices |
| Regulatory Context | Designed for compliance with WS/T 107.1–2016 (China), ISO 8468:1996 (Water Quality – Determination of Iodide), and USP <232>/<233> elemental impurities guidance frameworks |
| Cooling Method | Integrated Peltier-based water bath refrigeration |
| Digestion Architecture | Graphite block digestion with reflux condensation |
| Reagent Consumption | Optimized for ≤1.5 mL per sample (standard protocol) |
| Throughput | Up to 48 samples per batch (with optional autosampler) |
| Safety Compliance | Meets GB/T 18204.1–2013 (Public Health Inspection Standards) for laboratory ventilation integration |
| Brand | Anjie |
|---|---|
| Origin | Shanghai, China |
| Manufacturer Type | Authorized Distributor |
| Instrument Type | Photomultiplier Tube (PMT) Detector |
| Wavelength Range | 190–900 nm |
| Spectral Bandwidth | 0.1 nm / 0.2 nm / 0.4 nm / 1.0 nm (selectable) |
| Wavelength Accuracy | ±0.2 nm |
| Wavelength Repeatability | ±0.1 nm |