| Brand | ANJIE |
|---|---|
| Origin | Shanghai, China |
| Model | AJ-5100 |
| Measurement Principle | Potassium Permanganate Oxidation with ORP Endpoint Detection |
| Instrument Type | Benchtop Laboratory Analyzer |
| Range | 0–4.5 mg/L as O₂ |
| Repeatability | RSD ≤ 5.0% |
| Accuracy | ±10% of reading |
| Limit of Detection | < 0.07 mg/L |
| Resolution | 1 µL |
| Sample Analysis Time | 3 min 50 s per sample |
| Heating Method | Infrared Rapid Boiling (reaches boiling within 100 s) |
| Reaction Control | User-defined oxidation holding time |
| Fluidic Architecture | Dedicated reagent addition station, titration analysis station, and waste discharge station |
| Dispensing System | Single-arm precision liquid handler for all reagents |
| Brand | ANJIE |
|---|---|
| Origin | Shanghai, China |
| Model | APA-500 |
| Principle | Potassium Permanganate Oxidation with ORP Endpoint Detection |
| Type | Online Analyzer |
| Range | 0–5 mg/L |
| Repeatability | ±2% RSD |
| Accuracy | ±5% of reading |
| LOD | 0.05 mg/L |
| Resolution | 1 µL |
| Cycle Time | 4 min per sample |
| Temperature Control | 40–100 °C (adjustable water bath) |
| Environmental Monitoring | Integrated ambient temperature, humidity, and atmospheric pressure sensors |
| Sample Handling | Motorized adaptive gripper (pneumatic-free) |
| Endpoint Detection | RGB-spectral colorimetric endpoint simulation with real-time video recording |
| Compliance | Designed for ISO 8467:2021, EPA Method 4500-PERM, and GB/T 5750.7–2023 |