| Model | PX04006 |
| Combustion Chamber Dimensions | 750(W) × 300(D) × 900(H) mm |
| Control Cabinet Dimensions | 650(W) × 650(D) × 1600(H) mm |
| Power Supply | AC 220 V, 50 Hz, 16 A |
| Weight | ~180 kg |
| Fuel Source | Propane (≥95% purity) |
| Heat Flux Calibration Range | 0–100 kW/m² |
| Heat Flux Accuracy | <±3% |
| Response Time | 0.2 s |
| Emissivity | >0.95 |
| Radiant Heat Source | Nine T150 quartz infrared tubes (13 kW/m² ±4 kW/m²) |
| Convective Heat Source | Two 45°-inclined Meker burners (38 mm outer diameter, 5/16″ inner orifice), 25 kW total output |
| Total Incident Heat Flux | 84 kW/m² ±2 kW/m² |
| Temperature Measurement | OMEGA thermocouples with stainless-steel calibration disk, 0–1100 °C range, 0.1 °C display resolution |
| Cooling System | Integrated chiller (10 L/min flow rate), water-cooled aluminum heat sink plate (340 × 286 mm), portable closed-loop water source |
| Automation | PLC-controlled electromechanical positioning modules for specimen and heat flux sensor actuation |
| Software Platform | NI LabVIEW-based HMI with real-time curve visualization, data logging, audit-trail-enabled reporting, and GLP-compliant export formats |