- All
- Favorite
- Popular
- Most rated
| Brand | Autobio |
|---|---|
| Model | LUMO |
| Detection Method | Chemiluminescence |
| Detector Type | Imported Photomultiplier Tube (PMT) |
| Luminescence Range | 10⁰–10⁹ RLU |
| Plate Format | 96-well and 48-well microplates |
| Read Time per Well | 0.1–10 s |
| Full 96-Well Plate Read Time | <32 s |
| Sensitivity | 10⁻² mol/plate |
| Precision (CV) | <2% |
| Inter-well Cross-Talk | <5×10⁻⁷ |
| Plate Shaking Modes | 3 programmable patterns (speed & duration adjustable) |
| Power Input | 220 V ± 22 V AC |
| Power Consumption | <32.4 VA |
| Dimensions (L×W×H) | 470 × 335 × 205 mm |
| Weight | 13 kg |
| Safety Compliance | CE-marked, 24 V DC internal supply, ISO 13485 certified |
| Brand | Autobio |
|---|---|
| Model | PHOMO |
| Origin | Henan, China |
| Instrument Type | Multifunctional Automated Microplate Reader |
| Assay Mode | Endpoint |
| Calibration Curve | Linear |
| Detector | Dual-LED Laser-Based Photodiode Array |
| Wavelength Range | 200–1000 nm |
| Absorbance Range | 0–4.5 OD |
| Absorbance Linearity | 0.000–2.000 Abs ±0.01% |
| Measurement Speed | 12 s (single wavelength), 20 s (dual wavelength), 6 s (kinetic single-wavelength) |
| Precision (CV) | <0.5% |
| Repeatability (CV) | ≤0.15% |
| Light Source | Dual-LED Cold Light Source (Lifetime ≥100,000 h) |
| Optical Architecture | RSE (Real-time Spectral Enhancement) Silicon Photodetector Platform |
| Scan Mode | Multi-point per well (up to 9 points) |
Show next