GBPI Y210 Oxygen Transmission Rate (OTR) Tester
| Brand | GBPI |
|---|---|
| Origin | Guangdong, China |
| Manufacturer Type | OEM/ODM Manufacturer |
| Country of Origin | China |
| Model | Y210 |
| Pricing | Upon Request |
| Test Principle | Differential Pressure (Manometric) Method |
| Standards Compliance | ISO 2556:2001, ISO 15105-1:2007, ASTM D1434–82 (Reapproved 2003), GB/T 1038–2000, YBB00082003 |
| Measurement Range | 0.02–50,000 cm³/(m²·day·0.1 MPa) [expandable to 600,000 cm³/(m²·day·0.1 MPa)] |
| Accuracy | ±0.01 cm³/(m²·day·0.1 MPa) |
| Temperature Control Range | 5–50 °C |
| Stability | ±0.01 °C |
| Humidity Control Range | 0% RH (5–100% RH optional) |
| Accuracy | ±2% RH |
| Vacuum Capability | ≤20 Pa base pressure |
| Absolute vacuum resolution | 0.1 Pa |
| Test Gas Compatibility | O₂, CO₂, N₂ |
| Sample Thickness Limit | ≤2 mm |
| Effective Test Area | 50.24 cm² (Φ80 mm) |
| Power Supply | 220 V AC, 50 Hz |
Overview
The GBPI Y210 Oxygen Transmission Rate (OTR) Tester is a precision-manufactured differential pressure (manometric) gas permeation analyzer engineered for quantitative determination of oxygen transmission through polymeric films, laminates, foils, pharmaceutical blister packs, IV bags, and other flexible or semi-rigid packaging materials. It operates on the fundamental principle of steady-state gas diffusion under controlled transmembrane partial pressure gradient—where a defined upstream (high-pressure) gas phase drives molecular transport across a conditioned specimen into a downstream (low-pressure) measurement chamber. The rate of pressure rise in the evacuated downstream cavity is directly proportional to the volumetric gas flux, enabling calculation of OTR per ISO 15105-1 and ASTM D1434 methodologies. Designed for laboratories requiring regulatory-grade repeatability and traceable calibration, the Y210 integrates high-stability pressure transducers, ultra-low-leakage test chambers, and dual-stage vacuum architecture to ensure measurement integrity across low- to ultra-high-barrier materials.
Key Features
- Manometric differential pressure system with real-time downstream cavity pressure monitoring at 0.1 Pa resolution, supporting both low-permeability (e.g., AlOx-coated PET) and high-permeability (e.g., LDPE) materials without sensor range switching.
- Patented dual-zone thermal management using electromagnetic programmable step-control: achieves ±0.01 °C temperature stability across 5–50 °C operating range; ramp rates fully adjustable to minimize thermal shock during multi-condition testing.
- Modular humidity control module (optional): delivers precise 0–100% RH conditioning on either side of the specimen, compliant with ISO 2556 Annex B for moisture-dependent barrier evaluation.
- Integrated 7-inch TFT LCD interface with local parameter visualization—including real-time OTR, differential pressure, absolute chamber pressures, and temperature/humidity traces—enabling standalone operation without host PC.
- Automated calibration routines: includes certified standard film verification (e.g., SRM-certified PET reference) and independent pressure sensor zero/span validation prior to each test sequence.
- Overpressure safety interlock and redundant vacuum shutoff prevent mechanical stress on seals and maintain long-term chamber integrity (>10,000 test cycles).
- Fully automated workflow: sample loading detection, vacuum evacuation, gas dosing, equilibration monitoring, and kinetic curve acquisition are executed per preloaded SOP protocols.
Sample Compatibility & Compliance
The Y210 accommodates flat, non-porous specimens up to 2 mm thick and 80 mm in diameter (50.24 cm² effective area), including monolayer polymers (e.g., PP, PE), coextrusions, metallized films, SiOx/AlOx barrier coatings, cold-form foil blisters, and sterile medical pouches. All hardware and firmware conform to ISO/IEC 17025 documentation requirements for testing laboratories. Data acquisition meets FDA 21 CFR Part 11 criteria when paired with validated software: electronic signatures, audit trails, and immutable raw data archiving are supported. Instrument qualification packages (IQ/OQ/PQ) and GMP-compliant test reports align with USP , EP 3.2.1, and ICH Q5C stability guidance for pharmaceutical primary packaging.
Software & Data Management
GBPI PermeaSoft v4.2 provides full instrument control, method configuration, and GLP-compliant reporting. It supports multi-parameter time-series export (CSV, PDF, XML), customizable pass/fail thresholds per material specification, and statistical process control (SPC) charts for trend analysis across batches. Raw pressure vs. time datasets are stored with metadata (operator ID, environmental logs, calibration timestamps). Software validation documentation—including risk assessment (ISO 14971), requirement traceability matrix, and boundary testing reports—is available upon request for regulated users. Integration with LIMS via OPC UA or HL7 is configurable.
Applications
- Pharmaceutical packaging validation: OTR screening of PTP blisters, sachets, and IV container films per YBB00082003 and USP .
- Food packaging shelf-life modeling: correlation of OTR values with accelerated aging studies (ASTM F1980) for modified atmosphere packaging (MAP).
- Barrier coating R&D: quantification of deposition efficiency for plasma-enhanced chemical vapor deposition (PECVD) and atomic layer deposition (ALD) processes.
- Regulatory submission support: generation of data packages for EMA CHMP, PMDA, and NMPA dossiers requiring ISO 15105-1–compliant permeation metrics.
- Quality control in film extrusion lines: inline OTR trending with SPC alerts triggered by deviation from control limits.
FAQ
What gases can be tested besides oxygen?
The Y210 supports O₂, CO₂, and N₂ as standard test gases. Other permanent gases (e.g., He, Ar) may be evaluated subject to sensor linearity verification and gas-specific calibration.
Is humidity control mandatory for all tests?
No—0% RH (dry) testing is default. Optional RH modules allow symmetrical or asymmetrical humidity conditioning per ISO 2556 Annex B for hygroscopic materials.
How often does the system require recalibration?
Annual metrological recalibration is recommended. In-house verification using certified reference films (e.g., GBPI SRM-OTR-01) should be performed before each daily test session.
Can the instrument comply with 21 CFR Part 11?
Yes—when operated with PermeaSoft v4.2 under validated configuration, including role-based access control, electronic signatures, and audit-trail-enforced data immutability.
What vacuum level is required to initiate a test?
The system automatically evacuates the downstream chamber to <20 Pa absolute pressure prior to gas introduction, verified in real time by integrated capacitance manometers.