Empowering Scientific Discovery

JINTENG GM-0.5A Dual-Mode Oil-Free Diaphragm Vacuum Pump

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand JINTENG
Origin Tianjin, China
Pump Type Oil-Free Diaphragm Pump
Dual-Mode Operation Vacuum & Pressure (Positive/Negative)
Ultimate Vacuum ≤ 0.075 MPa (75 kPa absolute)
Maximum Positive Pressure ≥ 30 psi (207 kPa)
Air Flow Rate (Vacuum/Pressure Mode) 30 L/min
Motor Power 120 W, Single-Phase
Noise Level < 55 dB(A)
Operating Temperature Range 7–40 °C
Pump Head Material Anodized Aluminum Alloy with Corrosion-Resistant Coating
Diaphragm Material Fluoroelastomer (FKM)
Inlet/Outlet Port Diameter Ø6 mm
Dimensions (W×D×H) 215 × 165 × 270 mm
Net Weight 8 kg
Continuous Duty Rating Yes (Equipped with Integrated Air-Cooling System)
Compliance Designed for GLP-compliant laboratory environments

Overview

The JINTENG GM-0.5A is a dual-mode, oil-free diaphragm pump engineered for precision laboratory applications requiring both vacuum generation and positive-pressure air delivery in a single compact unit. Unlike traditional rotary vane or piston pumps, this device operates on a reciprocating diaphragm principle—where a fluorocarbon elastomer (FKM) membrane displaces gas volume without lubricants, mechanical contact, or hydrocarbon contamination. Its sealed, dry compression mechanism ensures Class 0 air purity per ISO 8573-1, making it suitable for sensitive analytical workflows including HPLC sample preparation, solid-phase extraction (SPE), vacuum filtration, lyophilization support, and controlled-pressure drying. The pump delivers stable vacuum down to 0.075 MPa absolute (equivalent to ~75 kPa or ~22 inHg) and positive pressure up to 30 psi (207 kPa), enabling seamless transition between aspiration and pressurization tasks—eliminating the need for separate vacuum and air compressor units.

Key Features

  • Dual-mode functionality: Simultaneous vacuum and positive-pressure operation via internal valving architecture—no external switching hardware required.
  • Fuel-free, maintenance-light design: FKM diaphragm resists chemical degradation from organic solvents, acids, and bases; no oil changes, seals, or lubrication intervals.
  • Thermally optimized housing: Die-cast aluminum body with corrosion-resistant coating dissipates heat efficiently; integrated passive airflow cooling allows uninterrupted 24-hour duty cycles.
  • Low-noise acoustics: Precision-balanced motor and vibration-dampened mounting reduce operational sound to <55 dB(A) at 1 m distance—ideal for shared lab spaces and noise-sensitive environments.
  • Stable pressure control: Adjustable pressure regulator enables fine-tuned output across the full 0–30 psi range, supporting method-specific filtration rates or controlled gas flow in inert atmosphere chambers.
  • Compact footprint and portability: Weighing only 8 kg with ergonomic top handle and non-slip suction feet for secure benchtop placement—optimized for modular lab layouts and mobile analytical setups.

Sample Compatibility & Compliance

The GM-0.5A is compatible with aqueous, mildly acidic, and low-concentration organic solvent vapors commonly encountered during filtration, degassing, and sample concentration procedures. Its FKM diaphragm and stainless-steel valve components provide resistance to ethanol, acetone, methanol, dilute HCl/HNO₃ (<10%), and phosphate-buffered saline solutions. While not rated for aggressive halogenated solvents (e.g., chloroform, DCM) or strong oxidizers (e.g., concentrated H₂O₂), it meets standard safety thresholds for routine QC/QA labs operating under ISO/IEC 17025 or cGMP-aligned protocols. The pump’s oil-free operation supports compliance with USP , FDA 21 CFR Part 11 (when paired with validated software interfaces), and GLP documentation requirements for audit-trail-capable instrumentation.

Software & Data Management

The GM-0.5A operates as a standalone analog device with manual pressure/vacuum regulation and does not include embedded digital controls or firmware. However, its consistent performance profile and repeatable flow characteristics make it highly compatible with external data acquisition systems. Users may integrate the pump into automated workflows via 0–10 V or 4–20 mA analog signal inputs (with optional third-party controller modules) for remote start/stop, pressure ramping, or event-triggered operation. All mechanical adjustments—including inlet/outlet port orientation (standard G1/8″ BSP threads) and pressure setpoint calibration—are accessible without tools, facilitating rapid reconfiguration between SOP-defined methods.

Applications

  • Vacuum-assisted filtration of biological samples, environmental water extracts, and pharmaceutical suspensions using glass or polypropylene filter holders.
  • Positive-pressure purging of GC autosampler vials, SPE cartridges, and microplate manifolds prior to elution.
  • Supporting vacuum ovens and desiccators in materials science labs where residual oil vapor must be excluded from thermal treatment zones.
  • Feeding gas chromatography carrier gas lines in resource-constrained settings where nitrogen generators are unavailable.
  • Providing clean, dry air for pneumatic actuators in custom-built lab automation rigs and microfluidic test benches.

FAQ

Is the GM-0.5A certified for use in regulated pharmaceutical laboratories?
It is designed to meet baseline requirements for oil-free air supply in GLP and cGMP environments but does not carry formal CE, UL, or ATEX certification. Validation documentation (IQ/OQ/PQ templates) is available upon request for integration into quality-controlled workflows.
Can this pump handle corrosive gases such as HCl or Cl₂?
No. It is intended for ambient air and non-aggressive vapors only. Exposure to chlorine, hydrogen sulfide, ammonia, or high-concentration acid fumes will degrade the FKM diaphragm and compromise long-term reliability.
What is the expected service life of the diaphragm under continuous operation?
Under typical lab usage (≤8 hrs/day, ambient temperature ≤30°C), the FKM diaphragm maintains functional integrity for ≥12,000 hours. Replacement kits—including diaphragm, valves, and gaskets—are available directly from JINTENG authorized distributors.
Does the pump include an exhaust filter?
Yes—a particulate and activated carbon composite filter is factory-installed on the exhaust path to capture trace moisture, aerosols, and volatile organics before atmospheric release.
Is there a warranty or technical support available internationally?
JINTENG provides a 24-month limited warranty covering material and workmanship defects. Technical documentation, spare parts catalogs, and application engineering support are delivered in English via regional partners in EMEA, APAC, and North America.

InstrumentHive
Logo
Compare items
  • Total (0)
Compare
0