Other Brands OK-TH Series Temperature and Humidity Controlled Environmental Test Chamber
| Brand | Other Brands |
|---|---|
| Model Range | OK-TH-80 / OK-TH-150 / OK-TH-225 / OK-TH-408 / OK-TH-800 / OK-TH-1000 |
| Internal Dimensions (W×H×D, cm) | 40×50×40 to 100×100×100 |
| External Dimensions (W×H×D, cm) | 100×165×100 to 180×200×180 |
| Temperature Range | -70 °C to +150 °C (configurable in four standard spans) |
| Humidity Range | 20–98 %RH (optional: 5–98 %RH or 10–98 %RH) |
| Temperature Uniformity | ±0.5 °C |
| Humidity Uniformity | ±2.5 %RH |
| Temperature Stability | ±2 °C |
| Humidity Stability | +2–3 %RH |
| Resolution | 0.1 °C / 0.1 %RH |
| Heating Rate | ~4.0 °C/min |
| Cooling Rate | ~1.0 °C/min |
| Refrigeration System | Hermetic dual-stage compressor with eco-friendly R404A/R23 refrigerant blend |
| Controller | LCD touch-screen interface with bilingual (English/Chinese) operation |
| Construction Material | SUS#304 stainless steel interior and exterior |
| Standard Accessories | One Φ50 mm cable port, two adjustable SUS#304 shelves, four casters + four leveling feet |
| Power Supply | AC 380 V ±5 %, 50 Hz ±0.5 Hz, 3-phase 5-wire, 2.5 m cable |
| Compliance | MIL-STD-810G (via GJB150.3/4), IEC 60068-2-1 (Cold), -2-2 (Dry Heat), -2-30 (Damp Heat, cyclic), -2-78 (Damp Heat, steady-state), GB/T 2423.1–2423.4 |
Overview
The OK-TH Series Temperature and Humidity Controlled Environmental Test Chambers are engineered for precision simulation of climatic conditions in laboratory and industrial quality assurance environments. These chambers operate on a dual-loop control architecture—separate proportional-integral-derivative (PID) algorithms govern temperature and humidity regulation—to ensure stable, repeatable exposure profiles across extended test durations. The system employs forced-air convection with optimized airflow distribution to maintain uniform thermal and hygroscopic conditions throughout the working volume, as verified by ISO 16750-4 and IEC 60068-3-5 calibration protocols. Designed for compliance-driven testing workflows, the OK-TH series supports both steady-state (constant) and cyclic (alternating) environmental stress profiles, making it suitable for accelerated aging, reliability validation, and pre-compliance screening per military, automotive, and electronics standards.
Key Features
- Modular chamber configurations spanning 80 L to 1000 L internal volume—scalable to accommodate prototype assemblies, PCBs, enclosures, or small subsystems.
- Four-tier temperature range options: 0–150 °C, –20–150 °C, –40–150 °C, and –70–150 °C—achieved via cascade refrigeration with R404A (high-stage) and R23 (low-stage) refrigerants, meeting EPA SNAP and EU F-Gas Regulation requirements.
- Humidity generation via steam injection with PID-controlled saturation vessel; dehumidification through surface condensation on chilled evaporator coils—enabling precise RH setpoint control from 20 %RH to 98 %RH (with optional extension to 5 %RH).
- SUS#304 stainless steel construction—fully welded interior cavity, non-porous surfaces, and corrosion-resistant hardware—for long-term stability under high-humidity and saline fog test conditions.
- Intuitive 7-inch TFT-LCD touch controller with real-time trend graphs, multi-segment programmability (up to 99 cycles × 99 steps), data logging (internal SD card + USB export), and configurable alarm thresholds (deviation, sensor fault, door open).
- Integrated mechanical safety systems: over-temperature cutoff, refrigerant pressure monitoring, phase-loss protection, and independent humidity over-saturation lockout.
Sample Compatibility & Compliance
The OK-TH chambers accommodate diverse sample geometries through customizable internal dimensions and standard accessories—including removable shelves, a 50 mm diameter cable port with silicone gasket, and load-rated casters with locking leveling feet. All models meet structural and operational requirements for test method execution under MIL-STD-810G (via Chinese national adaptation GJB150.3–150.4), IEC 60068-2-1 (cold), -2-2 (dry heat), -2-30 (cyclic damp heat), and -2-78 (steady-state damp heat). Chamber performance is validated per IEC 60068-3-5 (methods for determining uniformity and stability), with documented traceability to NIST-traceable reference sensors. Optional qualification packages include IQ/OQ documentation templates aligned with GLP and ISO/IEC 17025 laboratory accreditation frameworks.
Software & Data Management
While the embedded controller provides standalone operation, optional PC-based software (OK-Link™) enables remote monitoring, script-based test sequencing, automated report generation (PDF/CSV), and audit-trail-enabled user access control. Data logs include timestamped temperature/humidity readings, setpoint deviations, alarm events, and system status flags—structured to support FDA 21 CFR Part 11 compliance when paired with electronic signature modules and role-based permissions. Raw data export adheres to ASTM E2500-18 Annex A3 formatting conventions for regulatory submissions.
Applications
- Environmental stress screening (ESS) of electronic components, automotive ECUs, and aerospace avionics per JEDEC JESD22-A104 and A110.
- Material stability assessment for pharmaceutical packaging, medical device housings, and polymer composites under ISO 11607-1 accelerated aging protocols.
- Pre-conditioning of sensors, batteries, and optoelectronic modules prior to functional testing or EMC evaluation.
- Validation of sealing integrity and condensation resistance in IP-rated enclosures (IEC 60529) and MIL-DTL-38999 connector assemblies.
- Research-grade climate simulation for material science labs investigating hygrothermal degradation kinetics in composites and coatings.
FAQ
Is this chamber certified to IEC 60068-2-30 for cyclic damp heat testing?
Yes—the OK-TH series is configured and factory-validated to execute IEC 60068-2-30 Test Db (cyclic damp heat) with programmable ramp rates, dwell periods, and transition tolerances per Clause 6.2.
Can the controller interface be operated exclusively in English?
Yes—the touch-screen controller supports full English-language UI localization, including menus, error messages, and report headers, with no Chinese text displayed unless manually selected.
What is the maximum allowable load weight per shelf?
Each SUS#304 shelf is rated for 15 kg uniformly distributed load; total chamber payload capacity is specified per model in the technical datasheet (e.g., OK-TH-408: 45 kg max).
Do you provide installation and commissioning support?
Standard delivery includes mechanical installation guidance and on-site operational verification. Full IQ/OQ services, including third-party metrology certification and GMP-aligned documentation, are available as optional add-ons.
Are spare parts and calibration services available internationally?
Yes—critical spares (humidity sensors, PID controllers, refrigerant filters) are stocked regionally; NIST-traceable calibration certificates and preventive maintenance contracts are offered globally through authorized service partners.