Empowering Scientific Discovery

ThermoTwin SA Automatic Fat Crystallization Point Analyzer by Linetronic

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand Linetronic
Origin Switzerland
Model ThermoTwin SA
Temperature Range +80 °C to −50 °C
Bath Temperature Range −10 °C to +50 °C
Heating Rate 3 °C/min
Cooling Rate 1 °C/min
Temperature Sensor PT100 Class A
Compliance CE Marked
Instrument Form Factor Benchtop with Dual Sample Positions
Cooling System CFC-free Air-Cooled Compressor
Sample Chamber Dry, Jacketed Cooling Environment
Data Output Real-time Slope Calculation (Q = ΔT/Δt), Shukoff & Tempering Curve Generation

Overview

The ThermoTwin SA Automatic Fat Crystallization Point Analyzer is a precision benchtop instrument engineered for the quantitative determination of crystallization onset temperature (COT), polymorphic transition temperatures, and thermal stability profiles in fats, oils, cocoa butter equivalents, and fat-based food matrices. Based on controlled-rate cooling and high-fidelity thermal monitoring, the system applies standardized thermal profiling protocols aligned with AOCS Cd 16b-93 (Crystallization Point of Fats and Oils) and ISO 6321 (Determination of Crystallization Point of Fats). Its core measurement principle relies on real-time detection of latent heat release during phase transition—captured via dual high-accuracy PT100 Class A sensors—one immersed directly in the sample, the other monitoring the aluminum thermal bath. The resulting thermogram enables precise identification of critical inflection points including minimum temperature (minT), maximum temperature (maxT), and slope-derived kinetic parameters (Q = ΔT/Δt), essential for predicting storage behavior, tempering response, and mouthfeel performance in chocolate and confectionery applications.

Key Features

  • Benchtop dual-position design enables parallel analysis of two independent samples without cross-contamination or thermal interference.
  • Integrated air-cooled compressor system—CFC-free and compliant with EU Regulation (EC) No 1005/2009—ensures stable, maintenance-free operation across extended duty cycles.
  • Aluminum alloy thermal bath with active temperature control (−10 °C to +50 °C) provides uniform thermal transfer and rapid equilibration prior to programmed ramping.
  • Dry-jacketed sample chamber eliminates condensation, oil migration, or frost formation—critical for reproducible crystallization kinetics in low-moisture lipid systems.
  • Onboard industrial-grade touchscreen PC runs proprietary Linetronic acquisition software, supporting fully automated Shukoff curve generation, tempering cycle simulation, and derivative-based peak detection.
  • PT100 Class A temperature probes (IEC 60751) deliver ±0.1 °C accuracy over the full operational range (−50 °C to +80 °C), traceable to national metrology institutes.
  • No consumables or calibration gases required; routine verification performed using certified reference materials (e.g., pure stearic acid, palmitic acid).

Sample Compatibility & Compliance

The ThermoTwin SA accommodates standard 10–25 mL glass vials or disposable polypropylene cups—eliminating solvent use, sample pre-treatment, or filtration steps. It is validated for use with edible oils (soybean, palm, sunflower), hydrogenated fats, cocoa butter, cocoa butter replacers (CBR), margarines, shortening, and bakery fat systems. All hardware and firmware comply with CE Directive 2014/30/EU (EMC) and 2014/35/EU (LVD). Data acquisition meets ALCOA+ principles for electronic records, supporting audit-ready workflows under GLP and GMP environments. Optional 21 CFR Part 11 compliance package available upon request—including user authentication, electronic signatures, and immutable audit trails.

Software & Data Management

The embedded Linetronic Analytical Suite provides intuitive workflow navigation, multi-curve overlay, automatic baseline correction, and export of raw time-temperature datasets in CSV and ASTM E1382-compliant formats. Advanced features include batch reporting with statistical summary (mean, SD, RSD%), pass/fail threshold logic per product specification, and customizable report templates aligned with internal QA documentation standards. All data are stored locally with optional network backup; no cloud dependency or third-party licensing required. Software updates are delivered via secure USB firmware loader, preserving full regulatory traceability.

Applications

  • Quality control of cocoa butter and cocoa butter equivalents in chocolate manufacturing (ISO 21708:2018).
  • Polymorph screening and β-V stabilization assessment in confectionery fats.
  • Shelf-life prediction through crystallization onset correlation with storage temperature history.
  • Supplier qualification and incoming raw material verification for edible oil refineries.
  • R&D support for fat structuring studies, emulsifier efficacy testing, and low-trans fat reformulation.
  • Regulatory submission dossiers requiring thermally resolved crystallization data (e.g., EFSA, FDA GRAS notifications).

FAQ

What standards does the ThermoTwin SA support for crystallization point determination?

It aligns with AOCS Cd 16b-93, ISO 6321, and ASTM D3921 for fat crystallization profiling, with method validation documentation provided.
Can the instrument generate both cooling and heating ramps within a single run?

Yes—programmable bidirectional thermal protocols allow sequential cooling/heating cycles to characterize hysteresis and recrystallization behavior.
Is external PC connectivity required for operation?

No—the integrated touchscreen PC handles all acquisition, analysis, and reporting; optional Ethernet port enables remote monitoring and data export.
How is temperature calibration verified?

Using NIST-traceable reference substances (e.g., stearic acid, mp = 69.3 °C ± 0.1 °C) and ice-point verification at 0.00 °C.
Does the system require daily maintenance or refrigerant servicing?

No—air-cooled compressor design eliminates refrigerant handling; only periodic cleaning of the aluminum bath and sensor ports is recommended.