GreenPrima Aqualysis 800 Online Water Hardness Analyzer (Calcium & Magnesium, Colorimetric Titration Method)
| Brand | GreenPrima |
|---|---|
| Origin | Shanghai, China |
| Manufacturer Type | Manufacturer |
| Model | TH7001 |
| Instrument Type | Online Water Quality Analyzer |
| Measured Parameter | Calcium and Magnesium Hardness (as CaCO₃) |
| Measurement Principle | Automated Colorimetric Titration |
| Measurement Range (with TH7001 reagent) | 0.447–625.8 ppm CaCO₃ |
| Accuracy | ±2% of selected reagent upper limit |
| Repeatability | ±1% of selected reagent upper limit |
| Analysis Time | ~3 minutes (hardness-dependent) |
| Display | Color LCD with bilingual (English/Chinese) interface |
| Dimensions | 400 × 300 × 180 mm (W×H×D) |
| Weight | 4 kg |
| Enclosure Rating | IP65 |
| Power Supply | 85–265 VAC, 47–63 Hz |
| Power Consumption | 25 VA (operational), 3.5 VA (standby) |
| Output Interfaces | 4 programmable relay outputs (250 VAC, 4 A), 1 isolated 4–20 mA output (max. 750 Ω), 1 RS485 Modbus RTU port |
| Digital Inputs | 2 programmable (IN1: start analysis |
| IN2 | stop analysis) |
| Data Storage | Internal non-volatile memory for measurement logs and events |
| Sample Requirements | pH 4–10 |
| Sample Temp. | 5–40 °C |
| Operating Ambient Temp. | 10–40 °C |
| Sample Pressure | 0.2–6 bar (recommended: 1–2 bar) |
| Mounting | Wall-mounted |
| Flush Time Programmable | 15–1800 s |
| Measurement Interval Programmable | 0–360 min |
Overview
The GreenPrima Aqualysis 800 Online Water Hardness Analyzer is an industrial-grade, fully automated instrument engineered for continuous, real-time monitoring of total calcium and magnesium hardness in water streams. It operates on the principle of standardized colorimetric titration—specifically, the reaction between a metallochromic indicator (e.g., Eriochrome Black T or Calmagite) and divalent cations (Ca²⁺ and Mg²⁺) under controlled pH conditions, followed by endpoint detection via photometric absorbance measurement (spectrophotometric extinction). This method complies with foundational principles outlined in ISO 6059 (water quality — determination of hardness — EDTA titrimetric method) and ASTM D1126 (Standard Test Method for Hardness in Water), adapted for unattended online operation. The system delivers traceable, reproducible results without manual intervention, making it suitable for environments where regulatory compliance, process continuity, and long-term stability are critical—particularly in steam generation, pharmaceutical water systems, and closed-loop cooling circuits.
Key Features
- Automated colorimetric titration with dual-wavelength photometric endpoint detection ensures high specificity for Ca²⁺/Mg²⁺ complexes, minimizing interferences from transition metals within specified limits (Fe < 3 ppm, Cu < 0.2 ppm, etc.).
- Modular reagent cartridge design supports six interchangeable hardness ranges (TH7001 to TH7100), enabling application-specific calibration without hardware modification.
- Bilingual color LCD interface (English/Chinese) provides intuitive navigation, real-time trend visualization, alarm status, and diagnostic feedback—including reagent level, valve actuation history, and optical path integrity checks.
- IP65-rated enclosure ensures reliable operation in humid, dusty, or splash-prone mechanical rooms, boiler houses, and utility corridors.
- Low-maintenance architecture: self-cleaning fluidic path, wear-resistant peristaltic pumps, and temperature-compensated photodiode array reduce scheduled servicing to annual verification only.
- Flexible I/O configuration includes four programmable relays (for pump control, alarm staging, or soft-start sequencing), two digital inputs for external trigger synchronization (e.g., with sampling valves or PLC cycles), and native Modbus RTU over RS485 for SCADA integration.
Sample Compatibility & Compliance
The Aqualysis 800 is validated for use with potable water, softened effluent, reverse osmosis permeate, boiler feedwater, and low-conductivity pharmaceutical waters (PW, WFI). Its operational constraints—pH 4–10, absence of suspended solids (>5 µm), and defined trace metal thresholds—are aligned with USP “Water for Pharmaceutical Purposes” and EP 2.2.42 “Water for Injection” guidance on pre-treatment compatibility. While not intrinsically 21 CFR Part 11 compliant, the device supports audit-trail-ready data export (via RS485 or periodic SD card dump) when integrated with validated LIMS or MES platforms. All reagents—including TH7001—meet REACH Annex XIV exclusion criteria and carry SDS documentation compliant with GHS Rev. 9.
Software & Data Management
Onboard firmware implements cyclic data logging with timestamped hardness values, reagent consumption counters, error codes (e.g., “low signal-to-noise”, “valve timeout”), and calibration event records. Measurement data is retained in non-volatile flash memory for ≥12 months at default 15-minute intervals. The RS485 interface supports Modbus RTU register mapping for seamless ingestion into Historian systems (e.g., OSIsoft PI, Ignition, WinCC). Optional GreenPrima DataBridge software enables remote configuration, firmware updates, and CSV report generation—including daily max/min/avg, deviation alerts, and reagent expiry forecasting. No cloud dependency or proprietary drivers are required.
Applications
- Boiler feedwater monitoring to prevent scale formation in high-pressure steam generators (ASME B31.1 / EN 12952-12).
- Softener breakthrough detection in ion exchange trains—enabling predictive regeneration scheduling.
- Pharmaceutical purified water (PW) and water-for-injection (WFI) loop surveillance per EU GMP Annex 1 and FDA Guidance for Industry (2022).
- Cooling tower blowdown optimization through real-time hardness trending and conductivity correlation.
- RO pretreatment validation and post-RO hardness leakage detection in semiconductor ultrapure water (UPW) systems.
- Thermal power plant condensate polishing system performance verification.
FAQ
Does the Aqualysis 800 require daily calibration?
No. Factory calibration remains stable for ≥6 months under normal operating conditions. Verification using NIST-traceable hardness standards (e.g., 100 ppm CaCO₃) is recommended quarterly.
Can TH7001 reagent be used interchangeably with other TH-series kits?
No. Each TH-coded reagent corresponds to a specific concentration range and endpoint pH buffer formulation. Substitution invalidates accuracy claims and voids metrological traceability.
Is the analyzer compatible with hazardous area installations (e.g., Zone 2)?
Not natively. The unit lacks ATEX or IECEx certification. For classified locations, install in a ventilated analyzer shelter with purge air or use as a remote sensor with intrinsically safe signal conditioning.
What maintenance tasks are required every 6 months?
Replacement of the optical cuvette seal, inspection of peristaltic pump tubing for cracking, and cleaning of the flow cell with 10% citric acid solution to remove carbonate deposits.
How does the system handle turbid or colored samples?
The instrument applies dual-wavelength compensation (reference + analytical wavelength) to correct for baseline absorbance drift caused by sample color or suspended particulates—provided turbidity remains below 5 NTU and no oil emulsion is present.