GreenPrima PM8200CL Chlorine Dioxide Analyzer

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand GreenPrima
Origin Switzerland
Manufacturer GreenPrima AG
Instrument Type Online Analyzer
Measurement Principle Amperometric Electrochemical Sensor (Constant Voltage Mode)
Measured Parameter Chlorine Dioxide (ClO₂)
Range 0–2.000 mg/L and 0–20.00 mg/L
Accuracy ±2% of reading
Repeatability ±0.01 mg/L
Resolution 0.001 mg/L
Detection Limit 0.001 mg/L
Temperature Compensation Automatic or Manual (NTC10K / PT1000)
Operating Temperature 0–70.0 °C
Storage Temperature −20–70.0 °C
Display Backlit Graphic LCD (Chinese/English Switchable)
Data Storage 600,000 records
Power Supply 90–260 VAC, 50/60 Hz or 24 VDC (optional)
Output Signals Dual isolated 4–20 mA (0.1% FS), RS485 Modbus RTU (HART optional)
Relay Outputs 2 programmable Hi/Lo alarms (5 A / 250 VAC / 30 VDC)
Cleaning Control Adjustable interval (0.1–1000 h) and duration (1–1000 s)
Enclosure Rating IP65
Dimensions (Controller) 144 × 144 × 106 mm
Panel Cutout 138 × 138 mm
Weight 0.86 kg
Sensor Model Bsens650 (Ag/AgCl reference, gel electrolyte, dual Pt working electrodes, Φ12 × 120 mm, max. pressure 6 bar, PG13.5 or M12 connector, 3 m cable standard)
Optional Sensor Bsens660 (dual gold-ring ClO₂ sensor for enhanced low-concentration stability)

Overview

The GreenPrima PM8200CL is a high-precision, online chlorine dioxide (ClO₂) analyzer engineered for continuous, real-time monitoring in critical water treatment applications. It operates on the amperometric electrochemical principle using constant-voltage polarography—a method standardized under ASTM D6169 and widely accepted for residual disinfectant measurement in potable water systems. Unlike colorimetric or iodometric methods, this sensor-based approach eliminates reagent consumption, membrane replacement, and sample pretreatment, delivering inherently stable output with minimal drift over extended deployment cycles. The system comprises a microprocessor-controlled transmitter (PM8200CL) and a field-replaceable, flow-through amperometric sensor—either the Bsens650 (dual platinum electrode) or the Bsens660 (dual gold-ring configuration)—mounted in a BAF615 flow cell. Its design adheres to ISO 9001-certified manufacturing protocols and supports compliance with WHO Guidelines for Drinking-water Quality, US EPA Method 334.0, and EU Directive 98/83/EC for disinfectant residual control.

Key Features

  • True reagent-free operation: No chemical dosing, no membrane changes, no electrolyte refills—reducing total cost of ownership and maintenance downtime.
  • Dual-electrode architecture: Bsens650 employs two platinum working electrodes with Ag/AgCl reference and gel-based electrolyte, enabling robust interference rejection in complex matrices (e.g., high chloride, organic load, or variable pH).
  • Bsens660 variant offers enhanced sensitivity below 0.1 mg/L via dual gold-ring geometry—optimized for ultralow ClO₂ monitoring in secondary distribution networks or pharmaceutical process water.
  • Intelligent temperature compensation: Supports NTC10K or PT1000 probes; auto-compensates across −10 to +130 °C (sensor operating range: −5 to +100 °C), ensuring accuracy across seasonal or thermal process variations.
  • Configurable digital I/O: Two isolated 4–20 mA outputs assignable to ClO₂ concentration and temperature; RS485 Modbus RTU interface for SCADA integration; optional HART protocol for legacy DCS compatibility.
  • Secure operational integrity: Password-protected configuration menu, audit-trail-capable event logging, and programmable cleaning cycles (duration and frequency independently adjustable) to maintain sensor surface integrity in fouling-prone environments.

Sample Compatibility & Compliance

The PM8200CL is validated for use in drinking water, hospital wastewater effluent, cooling tower make-up, food & beverage process water, and pharmaceutical purified water (PW) systems. Its sensor exhibits negligible cross-sensitivity to free chlorine (<0.5% signal contribution at 2 mg/L Cl₂), ozone (<1% at 0.5 mg/L O₃), and monochloramine—enabling selective ClO₂ quantification without spectral deconvolution. The analyzer meets EN 61326-1 (EMC), EN 61010-1 (safety), and is designed to support GLP/GMP data integrity requirements: all measurements are time-stamped, stored with sensor ID and calibration history, and exportable via USB or Modbus. While not FDA 21 CFR Part 11–certified out-of-the-box, its audit log structure and user-access controls facilitate validation under regulated environments when deployed with documented SOPs.

Software & Data Management

Data acquisition and configuration are managed locally via the backlit graphic LCD with intuitive icon-driven navigation and bilingual (English/Chinese) firmware. All operational parameters—including range selection, alarm thresholds, relay hysteresis, and cleaning schedule—are configurable without external software. Internal non-volatile memory stores up to 600,000 timestamped records (ClO₂ value, temperature, sensor status, calibration events). Export is supported via USB mass storage mode (CSV format) or real-time streaming over Modbus RTU. For centralized fleet management, GreenPrima’s optional CloudLink gateway enables secure TLS-encrypted telemetry to web-based dashboards, supporting predictive maintenance alerts based on sensor polarization voltage trends and response time degradation.

Applications

  • Drinking water treatment plants: Real-time ClO₂ dosing control at point-of-entry and post-filtration stages.
  • Hospital central sterile supply departments (CSSD): Monitoring ClO₂ concentration in instrument reprocessing water to ensure microbiological efficacy per ISO 15883-5.
  • Pharmaceutical water-for-injection (WFI) loops: Low-range verification (0–2 mg/L) during sanitization cycles where oxidant residuals must be tightly constrained.
  • Food processing facilities: Validation of ClO₂ wash water in produce lines per USDA-FSIS guidelines.
  • Industrial cooling systems: Corrosion-mitigated biocide control where excessive ClO₂ accelerates copper alloy degradation.

FAQ

Does the PM8200CL require daily calibration?
No. Field calibration is recommended every 7–30 days depending on application severity; factory calibration remains stable for ≥6 months under typical municipal water conditions. Zero and span checks can be performed using certified ClO₂ standard solutions (e.g., NIST-traceable KIO₃-based standards).

Can the same sensor measure free chlorine and ozone?
The Bsens650 and Bsens660 sensors are optimized specifically for ClO₂. While minor cross-response to free chlorine and ozone occurs, GreenPrima does not validate or support multi-analyte operation on a single sensor. Dedicated sensors (e.g., Bsens610 for free chlorine) are required for compliant multi-parameter monitoring.

What flow rate is required for accurate measurement in the BAF615 flow cell?
Optimal laminar flow is achieved at 150–300 mL/min. Lower flow rates may induce boundary layer effects; higher flows (>500 mL/min) risk mechanical stress on the sensor tip. Integrated flow switches are available as an accessory for alarm-triggered low-flow detection.

Is the controller suitable for outdoor installation?
Yes—the IP65-rated enclosure permits wall-mount or panel-mount deployment in sheltered outdoor enclosures. Ambient temperature must remain within 0–70 °C; direct solar exposure should be mitigated with sunshades to prevent internal condensation.