Beifen Sanpu BS-H2 Dual-Channel Ambient Air Sampler

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand Beifen Sanpu
Origin Beijing, China
Manufacturer Type Manufacturer
Product Category Domestic
Model BS-H2
Instrument Type Conventional Gas Sampler
Instrument Configuration Dual-Channel Gas Sampler
Flow Range 0–1000 mL/min
Maximum Vacuum Pressure 26 kPa
Flow Stability ±1%
Flowmeter Accuracy ±2.5%
Service Life at Max Flow & Backpressure 5000 h
Continuous Operation Time at Max Flow & Backpressure 8 h
Standard Flow Range (for Compliance) 100–500 sccm (0 °C / 101.3 kPa or 20 °C / 101.3 kPa)
Flow Repeatability ±2%
Timing Accuracy ≤±0.2%
Dimensions (L×W×H) 26×20×16 cm
Weight 3.65 kg
Battery Rechargeable Li-ion, 14.8 V / 7800 mAh
Charging Time <10 h (auto-cutoff with LED indicator)
Operating Temperature −5 to +45 °C
Battery Runtime ≥6 h

Overview

The Beifen Sanpu BS-H2 Dual-Channel Ambient Air Sampler is a precision-engineered, microprocessor-controlled instrument designed for compliant, trace-level sampling of volatile organic compounds (VOCs), formaldehyde, benzene homologues, and other airborne contaminants in indoor and ambient environments. It operates on the principle of constant-flow, low-pressure aspiration through sorbent-packed tubes—primarily Tenax-TA and activated carbon cartridges—enabling quantitative collection under standardized thermodynamic conditions. Engineered specifically to meet the operational and metrological requirements of GB 50325–2010 “Code for Control of Indoor Environmental Pollution in Civil Building Engineering”, the BS-H2 eliminates flow drift and calibration dependency associated with mechanical rotameters by integrating high-stability electronic mass flow sensors. All displayed flow values are automatically referenced to standard temperature and pressure (STP: 0 °C, 101.3 kPa) or standard ambient temperature and pressure (SATP: 20 °C, 101.3 kPa), eliminating manual environmental correction and reducing volumetric uncertainty in downstream thermal desorption–GC/MS analysis.

Key Features

  • Dual independent sampling channels with fully programmable flow (100–500 sccm) and time settings (0–99 h 59 min 59 s), each featuring real-time closed-loop feedback control;
  • High-torque diaphragm pump capable of sustaining ≥26 kPa vacuum against typical sorbent tube backpressures (e.g., Tenax-TA at 500 sccm), ensuring stable operation across variable resistance profiles;
  • Electronic thermal mass flow sensors (±2.5% full-scale accuracy, ±1% stability) replace legacy float-type meters, delivering digital resolution to 1 mL/min and eliminating parallax and operator reading error;
  • Integrated STP/SATP compensation algorithm—no external thermometer or barometer required; flow and elapsed time are continuously calculated and displayed in standard volume units;
  • Rechargeable lithium-ion battery (14.8 V / 7800 mAh) supports ≥6 hours of continuous dual-channel operation; intelligent charging circuit provides visual status indication (red → green LED);
  • Rugged portable enclosure (26 × 20 × 16 cm, 3.65 kg) rated for field deployment across −5 °C to +45 °C ambient conditions;
  • Automatic flow regulation maintains setpoint within ±2% repeatability even when load impedance changes dynamically during extended sampling campaigns.

Sample Compatibility & Compliance

The BS-H2 is validated for use with widely adopted solid-phase sorbent media, including but not limited to Tenax-TA, Carbopack™ series, and coconut-shell activated carbon tubes—commonly specified in Chinese national standards (GB/T 18883, GB 50325), ISO 16017-1, and EPA TO-17 methodologies. Its robust flow control architecture ensures consistent breakthrough volumes across varying cartridge resistances, supporting method-defined sampling durations from 15 minutes to 8 hours. The instrument’s metrological design aligns with GLP-aligned laboratory practices: all flow and timing parameters are digitally logged with timestamped records, enabling audit-ready data traceability. While not inherently 21 CFR Part 11 compliant out-of-the-box, its deterministic control logic and non-volatile memory support integration into validated workflows requiring electronic record integrity.

Software & Data Management

The BS-H2 operates via an embedded real-time control system with no external PC dependency. All operational parameters—including channel-specific flow setpoints, sampling duration, elapsed time, and instantaneous flow—are presented on a dual-line LCD display with 1 mL/min and 1 s resolution. The device stores runtime logs (start/stop timestamps, average flow per channel, total sampled volume in Lstd) in non-volatile memory, accessible via front-panel navigation. Optional RS-232 or USB-to-serial interface modules (sold separately) enable export of ASCII-formatted log files for integration into LIMS or custom QA/QC dashboards. No proprietary software installation is required; exported data conforms to CSV-compatible schema for direct import into Excel, Python pandas, or statistical validation packages.

Applications

  • Indoor air quality (IAQ) assessment per GB 50325–2010 for formaldehyde, TVOC, benzene, toluene, ethylbenzene, and xylene (BTEX);
  • Ambient air monitoring networks for urban VOC baseline studies and source apportionment;
  • Occupational hygiene surveys in manufacturing, printing, coating, and composite fabrication facilities;
  • Method development and validation for thermal desorption–gas chromatography–mass spectrometry (TD-GC-MS) workflows;
  • Field-deployable sampling in epidemiological exposure studies where portability, battery autonomy, and STP-referenced quantification are critical;
  • Calibration verification of fixed-site air monitors using co-located grab-sampling protocols.

FAQ

Does the BS-H2 require manual temperature or pressure input to calculate standard volume?
No. The instrument internally references all flow measurements to either 0 °C/101.3 kPa or 20 °C/101.3 kPa using factory-calibrated thermal sensor characteristics—no user input or field instrumentation is needed.

Can both channels operate at different flow rates simultaneously?
Yes. Each channel has independent flow and timer controls; users may configure Channel A at 200 sccm for 4 h and Channel B at 400 sccm for 2 h without cross-interference.

What is the maximum allowable backpressure for reliable operation?
The pump maintains rated flow stability up to 26 kPa static vacuum, accommodating high-resistance Tenax-TA tubes (e.g., 1/4″ OD × 3.5″ length, packed at 200 mg) at 500 sccm without flow deviation exceeding ±2%.

Is the battery replaceable in the field?
Yes. The sealed Li-ion pack is user-serviceable via four screws on the rear panel; replacement batteries (14.8 V, ≥7800 mAh, UL-certified) are available as OEM spares.

How is flow accuracy verified during routine maintenance?
A certified dry calibrator (e.g., Gilian Gilibrator-2 or Mesa Labs DryCal DC-Lite) can be connected upstream of the inlet; the BS-H2’s real-time flow display permits direct comparison against NIST-traceable reference values at multiple points across the 100–500 sccm range.