mini-BUCK M-5 Digital Soap Film Flow Calibrator
| Origin | USA |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | M-5 |
| Price | ¥15,800 |
| Product Type | Positive Displacement |
| Flow Range | 1–6000 mL/min |
| Accuracy | ±0.5% of reading |
| Operating Temperature | 0–40 °C |
| Power Supply | Rechargeable NiCd battery (4.8 V), 8 h continuous operation |
| Dimensions | 14 × 15 × 6 cm |
| Weight | 0.9 kg |
| Compliance | NIST-traceable calibration certificate included |
Overview
The mini-BUCK M-5 Digital Soap Film Flow Calibrator is a portable, microprocessor-controlled positive displacement flow measurement instrument engineered for high-reproducibility volumetric flow verification in field and laboratory environments. It operates on the soap film principle—a gravimetric, time-of-transit method where a precisely formed soap bubble traverses a calibrated glass tube under controlled pressure differential. This technique eliminates reliance on orifice coefficients, temperature compensation algorithms, or altitude-based corrections, delivering inherently stable measurements across global operating conditions (0–40 °C; sea level to >5000 m elevation). Unlike thermal mass or turbine-based calibrators, the M-5 provides true volumetric flow rate (mL/min) independent of gas composition, density, or viscosity—making it ideal for calibrating sampling pumps, rotameters, and panel-mounted flow indicators used in environmental monitoring, occupational hygiene, and industrial hygiene applications.
Key Features
- True volumetric flow measurement via soap film transit time across a precision-bore, NIST-traceable glass tube
- Wide dynamic range: 1–6000 mL/min with ±0.5% full-scale accuracy (as certified per ISO/IEC 17025-accredited calibration)
- Zero-altitude correction requirement: validated for consistent performance from sea level to high-elevation field sites
- Temperature-resilient design: minimal drift over 0–40 °C ambient range; no internal temperature sensor required for compensation
- One-button operation with backlit LCD display—optimized for readability under direct sunlight and low-light conditions
- Robust mechanical architecture: inverted-tube configuration prevents fluid leakage even at 180° orientation
- Integrated rechargeable NiCd battery (4.8 V) supports up to 8 hours of continuous use without external power
- Compact footprint (14 × 15 × 6 cm) and lightweight construction (0.9 kg) enable single-hand operation and extended field deployment
Sample Compatibility & Compliance
The M-5 is compatible with clean, dry, non-corrosive gases—including ambient air, nitrogen, oxygen, and synthetic calibration mixtures—commonly used in personal and area sampling systems. It is not intended for use with condensable vapors, particulate-laden streams, or aggressive chemicals that may degrade the silicone-based soap solution or glass bore surface. The instrument meets ASTM D3195 (Standard Test Method for Rotameters) for secondary calibration verification and serves as a primary reference standard for flowmeter validation per EPA Compendium Method TO-11A and NIOSH Manual of Analytical Methods (NMAM) Chapter 5. All units ship with a NIST-traceable calibration certificate issued by an ISO/IEC 17025-accredited laboratory. Additionally, each device bears certification from the China National Institute of Metrology (CNIM), confirming compliance with JJG 640–2001 (Verification Regulation for Soap Film Flowmeters).
Software & Data Management
The M-5 operates as a standalone, embedded-system calibrator with no PC interface or proprietary software dependency. Measurement data—including flow rate, elapsed time, and directional sign (± for suction vs. pressure-driven flow)—are displayed in real time on the integrated LCD. While no digital data logging is built-in, the unit’s repeatability (<0.3% RSD over 10 consecutive measurements) supports manual recordkeeping aligned with GLP and ISO 17025 documentation requirements. Users may integrate M-5 verification results into LIMS or electronic lab notebooks using standardized templates compliant with FDA 21 CFR Part 11 when paired with audit-trail-enabled entry protocols.
Applications
- Field verification and daily calibration of personal sampling pumps (e.g., SKC, Gilian, Apex) prior to air sampling campaigns
- Secondary calibration of variable-area flowmeters (rotameters) in environmental laboratories and manufacturing QC settings
- Validation of flow settings on gas chromatography carrier lines, dilution systems, and purge-and-trap apparatus
- Performance auditing of fixed-site air quality monitors requiring traceable flow assurance
- Training and proficiency testing for industrial hygienists and laboratory technicians under ISO/IEC 17025 competency frameworks
FAQ
Does the M-5 require recalibration after transportation or altitude change?
No. Its soap film principle is physically independent of local gravity and atmospheric pressure variations; no user adjustment or firmware correction is needed.
Can the M-5 be used with humid or particulate-laden air streams?
It is designed for clean, dry gases only. Moisture or aerosols may disrupt soap film integrity and compromise measurement validity.
Is the NIST-traceable certificate valid for regulatory submissions?
Yes—provided the certificate is current (within 12 months), issued by an ILAC-MRA signatory lab, and includes uncertainty budgeting per GUM guidelines.
What maintenance is required between calibrations?
Routine cleaning of the glass tube with isopropyl alcohol and replacement of soap solution every 3–6 months (or sooner if film formation becomes inconsistent).
How does the M-5 compare to thermal mass flow calibrators for low-flow applications?
Unlike thermal devices, the M-5 delivers absolute volumetric accuracy below 10 mL/min without zero-drift or gas-specific calibration factors—critical for respirable dust and VOC sampling compliance.