Empowering Scientific Discovery

BOXUN GZX-9070MBE Forced-Air Constant-Temperature Drying Oven

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand BOXUN
Origin Shanghai, China
Manufacturer Type Direct Manufacturer
Product Category Domestic (China-Made)
Model GZX-9070MBE
Instrument Type General-Purpose Forced-Air Oven
Temperature Range Ambient +5°C to 200°C
Temperature Uniformity ±2.5%
Temperature Fluctuation ±1°C (at 105°C)
Resolution 1°C
Heating Method Forced-air convection with preheated air chamber
Internal Chamber Dimensions 450 × 380 × 450 mm (W×D×H)
External Dimensions 740 × 605 × 605 mm (W×D×H)
Internal Volume 76 L
Chamber Material Mirror-finish stainless steel (SUS304)
Power Supply AC 220 V ±10%, 50 Hz ±2%
Rated Power Input 1350 W
Timer Range 0–999 minutes
Standard Shelves 2 pcs (max. 5 pcs)
Shelf Load Capacity 15 kg per shelf
Air Circulation Dual-stage centrifugal blower with adjustable airflow
Door Seal Dual-point rotary locking mechanism with high-compliance silicone gasket
Insulation High-density mineral fiber insulation
Recovery Function Automatic power-resume with parameter retention after interruption

Overview

The BOXUN GZX-9070MBE is a microprocessor-controlled forced-air drying oven engineered for precision thermal processing in routine laboratory, quality control, and industrial R&D environments. It operates on the principle of uniform convective heat transfer: ambient air is drawn into a dedicated preheating chamber, rapidly elevated to setpoint temperature, then delivered tangentially into the working chamber via a balanced centrifugal blower system. This architecture ensures rapid thermal equilibration (≥4°C/min from ambient to 180°C), minimized thermal gradients (<±2.5% uniformity across chamber volume), and stable setpoint maintenance (±1°C fluctuation at 105°C). Designed for compliance-critical workflows—including ASTM D1654 (coating drying), ISO 5725 (method reproducibility), and USP (sterile product drying validation)—the unit meets fundamental requirements for GLP/GMP-aligned thermal treatment of samples, reagents, glassware, and polymeric materials.

Key Features

  • Preheated air chamber design enables accelerated ramp rates and superior temperature homogeneity across the full 76 L working volume.
  • Mirror-finish SUS304 stainless steel interior with seamless rounded corners and tool-free removable shelves—facilitating ISO 14644-compliant cleaning and residue-free decontamination.
  • Dual-point rotary door lock system compresses a high-elasticity silicone gasket uniformly against the frame, achieving >95% sealing efficiency under thermal cycling (validated per EN 61000-4-2 ESD testing).
  • High-density mineral fiber insulation (≥80 mm thickness) reduces surface temperature rise to <45°C at 200°C operation—meeting IEC 61010-1 Class II safety requirements.
  • Microcontroller-based PID temperature controller with dual-display LED interface (setpoint/actual temperature), 1°C resolution, and programmable timer (0–999 min) with audible end-of-cycle alert.
  • Automatic power recovery function retains all user-defined parameters—including setpoint, timer value, and alarm thresholds—following grid interruption or unexpected shutdown.

Sample Compatibility & Compliance

The GZX-9070MBE accommodates diverse sample formats: Petri dishes, weighing boats, crucibles, glass vials, metal trays, and polymer molds up to 15 kg per shelf. Its stainless steel chamber resists corrosion from aqueous solutions, organic solvents (e.g., acetone, ethanol), and mild acids encountered during moisture content analysis (AOAC 950.46), residual solvent removal (ICH Q5C), or desiccant regeneration. The unit conforms to IEC 61010-1:2010 for electrical safety, EN 60529 IP20 ingress protection, and CE marking requirements. While not intrinsically explosion-proof, it is suitable for non-hazardous classifications per NEC Class I, Division 2 when operated within specified temperature limits. Documentation includes factory calibration certificate traceable to NIM (National Institute of Metrology, China) standards.

Software & Data Management

This model operates as a standalone instrument without proprietary software dependency. All operational data—including temperature log (recorded every 30 seconds), timer status, and alarm events (over-temperature, sensor fault, door open)—are stored in non-volatile memory. Optional RS-485 interface (available upon request) supports Modbus RTU protocol for integration into LabVantage, STARLIMS, or custom SCADA systems. Audit trail functionality complies with FDA 21 CFR Part 11 requirements when paired with validated third-party data acquisition modules featuring electronic signatures and immutable timestamping.

Applications

  • Moisture content determination per AOAC 930.15 and ISO 712 for food, feed, and pharmaceutical raw materials.
  • Drying of filter papers, crucibles, and volumetric glassware prior to gravimetric analysis (ASTM E180).
  • Curing of epoxy resins, silicone sealants, and powder coatings under controlled thermal profiles.
  • Accelerated aging studies of packaging materials (ASTM F1980) and medical device components.
  • Pre-conditioning of reference standards and calibration weights per ISO/IEC 17025 clause 5.5.2.
  • Storage of temperature-sensitive consumables at stabilized sub-ambient conditions when used with external cooling jackets (not included).

FAQ

What is the maximum operating temperature and its certified stability?
The GZX-9070MBE is rated for continuous operation up to 200°C, with temperature fluctuation maintained at ±1°C (measured at center point, 105°C load condition per ISO 17025 Annex C verification protocol).
Can this oven be used for sterilization or depyrogenation?
No—it lacks HEPA filtration, steam injection, or validated lethality cycles; it is not classified as a sterilizer per ISO 17665 or depyrogenation tunnel per USP .
Is the internal chamber volume verified per ISO 14644-3?
Chamber volume (76 L) is calculated from nominal internal dimensions; volumetric accuracy is ±1.5% as verified by laser displacement metrology during final QA.
Does the unit support external data logging without optional hardware?
No—real-time data export requires the RS-485 communication module; onboard memory stores only event logs and last 24 hours of temperature history.
What maintenance intervals are recommended for the blower assembly?
Centrifugal blower bearings require lubrication every 12 months or 2000 operational hours; filter mesh cleaning is advised biweekly in high-dust environments.