Empowering Scientific Discovery

BOXUN GZX-9456MBE Forced-Air Precision Drying Oven

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand BOXUN
Origin Shanghai, China
Manufacturer Type Direct Manufacturer
Product Category Domestic
Model GZX-9456MBE
Instrument Type High-Temperature Forced-Air Oven
Temperature Range Ambient +5°C to 300°C
Temperature Uniformity ±2.5%
Temperature Fluctuation ±1°C (at 105°C)
Resolution 1°C
Heating Method Forced-Air Circulation
Internal Dimensions 600 × 580 × 1300 mm (W×D×H)
External Dimensions 774 × 845 × 1704 mm (W×D×H)
Chamber Volume 210 L
Interior Material Mirror-Finish Stainless Steel (SUS304)
Power Supply AC 220 V ±10%, 50 Hz ±2%
Rated Power 3500 W
Timer Range 0–9999 min/h
Standard Shelves 4 pcs (up to 18 configurable)
Shelf Load Capacity 15 kg per shelf
Preheating Chamber Design Yes
Door Seal Dual-stage rotary locking with high-compression silicone gasket
Corner Design Rounded internal corners
Insulation High-density mineral fiber insulation
Recovery Function Power-loss auto-resume with parameter retention

Overview

The BOXUN GZX-9456MBE Forced-Air Precision Drying Oven is an engineered solution for controlled thermal processing in regulated laboratory and industrial quality control environments. Designed according to fundamental principles of convective heat transfer, the unit employs a dedicated preheating chamber and tangential airflow distribution to ensure rapid thermal equilibration and spatially uniform temperature profiles across its 210 L working volume. Its operational envelope—spanning ambient +5°C to 300°C with ±1°C fluctuation at reference setpoints and ±2.5% uniformity—meets routine requirements for drying, baking, sterilization validation support, polymer conditioning, and residual solvent removal in accordance with ISO 17025–accredited workflows. The oven’s structural architecture integrates thermal mass optimization, low-thermal-conductivity insulation, and dynamic air renewal via a shielded, low-noise centrifugal blower—enabling reproducible thermal exposure without drift under continuous operation.

Key Features

  • Mirror-finish SUS304 stainless steel interior with fully rounded corners and tool-free removable shelves for seamless cleaning and compliance with hygiene-critical applications.
  • PMMAⅠ intelligent control system featuring a color TFT display, capacitive touch interface, and real-time monitoring of temperature, elapsed time, remaining time, and system status.
  • Dual-stage rotary door lock mechanism ensures consistent compression against a high-temperature silicone gasket, minimizing thermal leakage and supporting stable chamber integrity during extended cycles.
  • Preheated air delivery system: ambient air is drawn into a dedicated heating cavity upstream of the main chamber, thermally conditioned, then injected via directional ducts to minimize cold spots and accelerate ramp rates (>5°C/min to 150°C).
  • Robust thermal insulation using high-density mineral fiber (≥100 mm thickness) surrounding all six chamber surfaces, reducing surface temperature rise and improving energy efficiency.
  • Power-loss recovery function retains setpoint, timer value, and operational state during unexpected outages—critical for unattended overnight runs and GLP-aligned protocols.

Sample Compatibility & Compliance

The GZX-9456MBE accommodates standard laboratory glassware (e.g., Petri dishes, weighing bottles, crucibles), metal trays, and polymer substrates up to 15 kg per shelf. Its stainless steel chamber resists corrosion from weak organic solvents, saline solutions, and low-concentration acids encountered in pharmaceutical stability testing or environmental sample preparation. The unit complies with IEC 61010-1:2010 for electrical safety in laboratory equipment and supports traceable calibration per ISO/IEC 17025 requirements. While not intrinsically rated for hazardous atmospheres, it is routinely deployed in non-explosive zones for residual moisture analysis (ASTM D1729, USP ), loss-on-drying (LOD) assays, and pre-conditioning of test specimens prior to mechanical or rheological evaluation.

Software & Data Management

The PMMAⅠ controller provides local data logging of temperature history (1-second sampling, 30-day buffer) and supports optional RS485 Modbus RTU communication for integration into centralized lab information management systems (LIMS). Audit trail functionality records user login/logout events, parameter modifications, alarm triggers (e.g., over-temperature, door-open, sensor fault), and cycle start/stop timestamps—enabling alignment with FDA 21 CFR Part 11 requirements when paired with validated third-party software. No cloud connectivity or proprietary firmware updates are embedded; all configuration remains locally managed to preserve data sovereignty and cybersecurity integrity.

Applications

  • Moisture content determination in raw materials (pharmaceutical excipients, food powders, ceramics).
  • Thermal stabilization of reference standards and calibration weights prior to metrological use.
  • Baking of coated substrates in thin-film R&D and printed electronics fabrication.
  • Drying of filter papers and membrane filters in environmental water analysis (EPA Method 160.1, ISO 5667-3).
  • Pre-heating of molds and tooling in polymer processing laboratories.
  • Conditioning of composite laminates before tensile or flexural testing (ASTM D638, ISO 527).

FAQ

What is the maximum recommended operating temperature for continuous use?
The GZX-9456MBE is rated for sustained operation up to 300°C. For optimal heater longevity and sensor stability, prolonged use above 250°C should be limited to validated protocols with documented thermal mapping.
Can the oven be calibrated internally without external equipment?
Yes—the PMMAⅠ system supports two-point offset calibration using a certified reference thermometer, with adjustments stored in non-volatile memory and traceable to NIST-traceable standards.
Is the unit suitable for drying samples containing volatile organic compounds (VOCs)?
It is not equipped with explosion-proof components or catalytic VOC scrubbers. Use in VOC-rich environments requires external fume extraction and risk assessment per OSHA 1910.1200 and local hazardous area classification.
How many shelves can be installed simultaneously without compromising airflow uniformity?
Up to 18 shelves may be fitted, but full-load uniformity validation (per ISO 14644-3 or internal SOP) is required when exceeding the standard 4-shelf configuration.
Does the oven meet GLP or GMP documentation requirements?
With optional RS485 interface and validated LIMS integration, the system supports electronic record generation, user authentication, and audit trails—fulfilling core ALCOA+ data integrity principles when implemented within a qualified environment.

InstrumentHive
Logo
Compare items
  • Total (0)
Compare
0