Empowering Scientific Discovery

BOXUN GZX-9140MBE Forced-Air Constant-Temperature Drying Oven

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand BOXUN
Origin Shanghai, China
Manufacturer Type Direct Manufacturer
Product Category Domestic (China-Made)
Model GZX-9140MBE
Instrument Type General-Purpose Forced-Air Oven
Temperature Range Ambient +5°C to 200°C
Temperature Uniformity ±2.5%
Temperature Fluctuation ±1°C (at 105°C)
Resolution 1°C
Heating Rate >4°C/min (from ambient to 180°C)
Chamber Volume 142 L
Interior Dimensions 550 × 470 × 550 mm
Exterior Dimensions 840 × 695 × 705 mm
Interior Material Mirror-Finish Stainless Steel (SUS304)
Heating Method Forced-Air Circulation with Dual-Speed Fan Motor
Power Supply AC 220 V ±10%, 50 Hz ±2%
Input Power 1750 W
Timer Range 0–999 min
Standard Shelves 3 pcs (up to 7 optional)
Shelf Load Capacity 15 kg per shelf
Insulation High-Density Mineral Wool
Door Seal Dual-Stage Rotational Locking System with Silicone Gasket
Viewing Window Large-Area Tempered Glass
Safety Features Over-Temperature Alarm, Auto Restart After Power Failure, Real-Time Digital Display

Overview

The BOXUN GZX-9140MBE Forced-Air Constant-Temperature Drying Oven is an engineered solution for precise thermal processing in laboratory and industrial quality control environments. Designed around the principle of forced convection heating, this oven circulates preheated air uniformly throughout the chamber using a dual-speed blower motor and optimized airflow ducting—ensuring rapid thermal equilibration and minimal spatial temperature gradients. Its operational range spans from ambient +5°C to 200°C, making it suitable for routine drying, sterilization validation support, moisture content determination (per ASTM E121–22), polymer curing, pharmaceutical stability testing (ICH Q1A), and pre-conditioning of samples prior to analytical measurement. The chamber’s mirror-finish stainless steel (SUS304) interior resists corrosion from solvents and acidic vapors, while the high-density mineral wool insulation minimizes thermal loss and improves energy efficiency during extended operation cycles.

Key Features

  • Microprocessor-based PID temperature controller with real-time digital display (dual LED screens) and intuitive touch-key interface for precise setpoint adjustment and timer programming
  • Pre-heating cavity design ensures heated air is fully mixed before entering the working chamber—enhancing thermal uniformity and reducing warm-up time to <15 minutes (to 105°C)
  • Dual-stage rotational door lock mechanism ensures consistent compression of the silicone gasket across the full perimeter, achieving reliable sealing performance and minimizing heat leakage
  • Large-view tempered glass window enables continuous visual monitoring without opening the door—critical for observing sample behavior during thermal treatment
  • Rounded internal corners and tool-free removable shelves simplify cleaning and reduce contamination risk—supporting GLP-compliant lab hygiene protocols
  • Auto-restart function preserves programmed parameters after power interruption, eliminating manual reconfiguration and preventing experimental discontinuity
  • Over-temperature safety cut-off with audible/visual alarm meets IEC 61010-1 requirements for Class II laboratory equipment

Sample Compatibility & Compliance

The GZX-9140MBE accommodates a broad spectrum of sample formats—including Petri dishes, weighing boats, crucibles, glassware, and standard ISO/ASTM test specimens—within its 142 L chamber. Its stainless steel construction and non-porous interior surface are compatible with aqueous, organic, and mildly corrosive media encountered in chemical analysis, materials science, and food testing workflows. The unit complies with GB/T 5170.2–2017 (Chinese national standard for temperature uniformity testing of drying ovens) and incorporates design elements aligned with ISO/IEC 17025:2017 clause 6.4.3 on environmental condition control. While not certified for Class I Div 1 hazardous locations, it is routinely deployed in non-explosive laboratory zones operating under ISO 14644-1 cleanroom classifications up to ISO 8.

Software & Data Management

This model operates as a standalone instrument without embedded data logging or network connectivity. All operational parameters—including set temperature, elapsed time, and alarm status—are displayed locally via dual high-brightness LED indicators. For traceable recordkeeping in regulated environments (e.g., pharmaceutical QC labs adhering to FDA 21 CFR Part 11), external validation-grade data loggers (e.g., compliant with NIST-traceable thermocouple inputs) may be integrated via the oven’s accessible thermistor output port (optional accessory). Routine calibration verification follows manufacturer-recommended procedures using calibrated reference thermometers (Class A platinum RTDs) placed at defined geometric positions within the chamber, per ISO/IEC 17025 Annex A.3 guidelines.

Applications

  • Drying of filter papers, soil samples, and biological tissues prior to gravimetric analysis (ASTM D2234, ISO 565)
  • Moisture loss-on-drying (LOD) assays for raw materials in pharmaceutical manufacturing (USP <731>)
  • Conditioning of polymer films and composites before tensile or DMA testing (ASTM D618, ISO 291)
  • Baking of ceramic substrates and printed circuit boards during R&D prototyping
  • Stabilization of reference standards and calibration weights in metrology laboratories
  • Pre-heating of glassware and metal components in preparation for vacuum or inert-atmosphere processing

FAQ

What is the maximum recommended continuous operating temperature?
The GZX-9140MBE is rated for continuous operation up to 200°C. Prolonged use above this threshold may accelerate degradation of insulation and seal integrity.
Does the oven support programmable multi-step temperature profiles?
No—this model features single-setpoint temperature control with independent timer functionality. Multi-segment ramp-soak profiles require external programmable controllers.
Can the shelving configuration be customized for oversized samples?
Yes—up to seven adjustable shelves can be installed; load distribution must remain ≤15 kg per shelf to maintain structural stability and airflow integrity.
Is the unit supplied with calibration documentation?
Factory calibration certificate (as-built) is included. Full IQ/OQ documentation packages and third-party calibration services are available upon request.
How often should temperature uniformity be verified?
Per ISO/IEC 17025, verification should occur at least annually—or after relocation, major maintenance, or whenever process-critical results exhibit unexplained variability.

InstrumentHive
Logo
Compare items
  • Total (0)
Compare
0