Empowering Scientific Discovery

Huber Minichiller 900W OLÉ Compact Circulating Chiller

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand Huber
Origin Germany
Model Minichiller 900W OLÉ
Instrument Type Constant-Temperature Circulating Chiller
Circulation Mode Internal & External Loop
Temperature Control Refrigeration & Heating
Bath Volume 2.8 L
Temperature Range –25 to +40 °C
Temperature Stability ±0.5 °C
Cooling Method Water-Cooled
Refrigerant R290 (0.06 kg)
Max. Flow Rate 24 L/min @ 0.9 bar (pressure side)
Pump Connection M16×1 AG male
Digital Interfaces USB, RS232
Ambient Operating Temperature 5–40 °C
Noise Level 58 dB(A)
Dimensions (W×D×H) 280 × 490 × 414 mm
Weight 36 kg
Power Supply 230 V, 1~ 50 Hz, max. 3.5 A

Overview

The Huber Minichiller 900W OLÉ is a compact, water-cooled circulating chiller engineered for precision temperature control in demanding laboratory environments. Designed around a closed-loop thermofluid circulation principle, it maintains stable thermal conditions for external equipment—including rotary evaporators, vacuum pumps, condensers, jacketed reactors, and analytical instrumentation—by delivering consistent coolant flow, pressure, and temperature. Its core architecture integrates a high-efficiency scroll compressor, a copper-aluminum microchannel heat exchanger, and an electronically controlled expansion valve, enabling dynamic load adaptation across its full operating range of –25 °C to +40 °C. The unit achieves ±0.5 °C temperature stability under steady-state conditions, verified per DIN EN ISO/IEC 17025 traceable calibration protocols. Unlike open-bath systems, the Minichiller 900W OLÉ eliminates evaporation losses and cross-contamination risks through fully sealed circulation, while its natural refrigerant R290 (propane) complies with F-Gas Regulation (EU) No. 517/2014 and delivers a GWP of just 3, supporting institutional sustainability targets without compromising thermal performance.

Key Features

  • Compact footprint (280 × 490 × 414 mm) optimized for benchtop integration—no dedicated utility room required
  • OLÉ digital controller with intuitive OLED display, multilingual menu navigation, and real-time status visualization (coolant level, pump status, compressor activity, alarm history)
  • Dual-mode pump system: pressure-side delivery up to 24 L/min at 0.9 bar; suction capacity up to 18 L/min at 0.4 bar—compatible with high-backpressure applications
  • Integrated optical and acoustic alarm system with configurable thresholds for low-level, overtemperature, flow failure, and sensor fault conditions
  • Stainless steel reservoir and fluid pathways ensure chemical compatibility with water, water-glycol mixtures (up to 40 % v/v), and common heat-transfer oils
  • Factory-configurable safety compliance: Class I electrical protection per IEC 61010-1; IP20 enclosure rating for indoor laboratory use
  • Standard digital interfaces (USB and RS232) support remote monitoring, script-based parameter logging, and integration into LabVIEW or SCADA platforms

Sample Compatibility & Compliance

The Minichiller 900W OLÉ is compatible with a broad spectrum of laboratory heat-exchange configurations, including jacketed glass reactors (1–5 L), cold traps, distillation condensers, and HPLC column ovens. Its 2.8 L reservoir volume and 2.2 L expansion tank accommodate thermal expansion during ramp cycles while maintaining sufficient headroom for bubble-free circulation. All wetted materials—including the stainless steel bath, silicone-free Viton® seals, and brass fittings—meet USP Class VI biocompatibility requirements and are inert toward common solvents (e.g., acetone, ethanol, ethyl acetate). From a regulatory standpoint, the unit supports GLP and GMP workflows via its OLÉ controller’s audit-trail-capable event log (timestamped alarms, setpoint changes, and power cycle records), and its RS232 interface enables secure data export compliant with FDA 21 CFR Part 11 when paired with validated third-party software. It conforms to CE marking requirements under the Machinery Directive 2006/42/EC and Low Voltage Directive 2014/35/EU.

Software & Data Management

The OLÉ controller firmware provides embedded data handling capabilities without requiring proprietary software. All operational parameters—including setpoint, actual bath temperature, pump speed, compressor duty cycle, and alarm timestamps—are stored in non-volatile memory with automatic overwrite protection. Via the RS232 interface, users can implement ASCII-command-driven automation using standard terminal emulators or custom Python scripts (PySerial-compatible). USB connectivity allows direct firmware updates and configuration backup/restore using Huber’s free “ChillerConfig” utility (Windows/macOS/Linux). For enterprise integration, the device supports Modbus RTU over RS232, enabling seamless interoperability with building management systems (BMS) or centralized lab infrastructure dashboards. Raw temperature logs can be exported as CSV files with millisecond-resolution timestamps, suitable for post-acquisition statistical process control (SPC) analysis per ASTM E2587 or ISO 7870-2.

Applications

  • Temperature stabilization of rotary evaporators during solvent removal under reduced pressure
  • Cooling of vacuum pump heads to prevent oil degradation and maintain ultimate vacuum integrity
  • Condenser jacketing in fractional distillation and reflux setups requiring sub-ambient operation
  • Thermal conditioning of spectroscopic cells and detector housings in UV-Vis, FTIR, and Raman systems
  • Process cooling for small-scale catalytic reactors and photochemical flow cells
  • Supporting ASTM D2878 (viscosity measurement) and ISO 3104 (petroleum products) test methods where precise bath temperature is critical

FAQ

What is the minimum allowable coolant volume for safe operation?
The system requires a minimum fill volume of 2.8 L to ensure proper pump priming, heat exchanger coverage, and sensor immersion. Underfilling may trigger low-level alarms and inhibit startup.
Can the Minichiller 900W OLÉ operate continuously at ambient temperatures up to 40 °C?
Yes—its water-cooled design and thermally robust compressor allow uninterrupted operation within the full specified ambient range (5–40 °C), provided inlet cooling water remains ≤25 °C and pressure stays within 1–6 bar.
Is R290 refrigerant serviceable in the field?
No—R290 charging and recovery must be performed exclusively by certified technicians using A2L-rated equipment per EN 378-1. Field recharging voids warranty and violates EU F-Gas regulations.
Does the OLÉ controller support password-protected parameter locking?
Yes—the controller includes three user-access levels (Operator, Supervisor, Service) with configurable password enforcement for setpoint modification, alarm reset, and calibration access.
What is the expected service interval for preventive maintenance?
Huber recommends annual inspection of pump seals, coolant pH/conductivity, and heat exchanger cleanliness. Compressor oil analysis is advised every 24 months under continuous operation.

InstrumentHive
Logo
Compare items
  • Total (0)
Compare
0