BrandHORIBA
OriginJapan
ModelTCA-51d / TCA-51p
Explosion ProtectionEx d IIB+H2 T4 / Ex px d IIC T4X
Detection PrincipleThermal Conductivity
MeasurandsH₂, O₂
Range0–10 to 0–100 vol% (H₂), 0–5 to 0–25 vol% (O₂)
Repeatability±0.5% F.S. (standard range)
Linearity±1.0% F.S.
Zero Drift±1.0% F.S./week (standard range)
Span Drift±2.0% F.S./week (standard range)
Response Time (T90)≤20 s
Sample Flow Rate300–1500 mL/min
Operating Pressure≥1.98 kPa (TCA-51d), ≥14.7 kPa (TCA-51p)
Wetted MaterialsSUS304, SUS316, FKM, Pt, Au, SiO₂, Borosilicate Glass
ComplianceIEC 60079-0, IEC 60079-1 (Ex d), IEC 60079-2 (Ex p), ATEX Directive 2014/34/EU, UKCA