- All
- Favorite
- Popular
- Most rated
| Origin | Italy |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | PMM9010/03P |
| Pricing | Upon Request |
| Brand | PMM |
|---|---|
| Origin | Italy |
| Model | EP600 / EP601 |
| Frequency Range | 100 kHz–9.25 GHz (EP600), 10 kHz–9.25 GHz (EP601) |
| Flatness | ±0.4 dB (typ.), ±0.5 dB (typ.) |
| Dynamic Range | 0.14–140 V/m (60 dB), 0.5–500 V/m (60 dB) |
| Linearity | ±0.4 dB (50 MHz, 0.3–100 V/m |
| Resolution | 0.01 V/m |
| Isotropy | ±0.5 dB (±0.3 dB typical) |
| Overload | 300 V/m (EP600), 1000 V/m (EP601) |
| Sampling Rate | 22 samples/s (max), adjustable down to 0.03 samples/s |
| Digital Filter | Adjustable low-pass, 2.3–28 Hz |
| Power Supply | Rechargeable 3 V Li-Mn battery (capacity: mAh not specified) |
| Operating Time | 80 h |
| Charging | Trickle charge, full capacity in 48 h |
| Temperature Sensor Resolution | 0.1 °C |
| Battery Voltage Sensing Resolution | 10 mV |
| Dimensions | ≤53 mm (probe body: 17 mm, sensor core: 17 mm) |
| Weight | 25 g (including 1 m fiber pigtail) |
| Connector | 20-UNF female |
| Fiber Interface | Duplex optical fiber, standard length 10 m (20/40 m optional) |
| PC Interface | USB via fiber-optic transceiver |
| Compliance | CE, RoHS, IEC 61000-4-3 (radiated immunity test support), ISO/IEC 17025 traceable calibration |
| Brand | PMM |
|---|---|
| Origin | Italy |
| Model Series | SHC-Series |
| Impedance | 1500 Ω |
| Attenuation | SHC-1 & SHC-1/1000: 35 dB |
| SHC-2 | 30 dB |
| Max Input Voltage | SHC-1/1000: 1 kV<sub>DC/AC<sub>rms</sub></sub> |
| Compliance | CISPR 16-1-2:2023 Edition |
| Application | Conducted Emission Testing per EN 55032, EN 55022, FCC Part 15, MIL-STD-461G |
| Brand | PMM |
|---|---|
| Origin | Italy |
| Model | Click4E |
| Compliance | CISPR 14-1:2005 (incl. Exceptions E3 & E4) |
| Channel Configuration | 1–4 channels (software-activated, hardware-expandable) |
| Measurement Mode | Fully automatic parallel testing or manual single-channel operation |
| Detection Method | Quasi-Peak (QP) detection per CISPR 16-1-1 |
| Data logging | Per-pulse duration, interval, amplitude, and timestamp per IEC/CISPR requirements |
| Report generation | Auto-generated, audit-ready PDF/CSV with full traceability |
| Storage | High-capacity embedded flash memory for ≥10⁶ pulse events |
| Brand | PMM |
|---|---|
| Origin | Italy |
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | PMM 1008 |
| Quotation | Upon Request |
| Compliance | IEC 61000-4-8, EN 61000-4-8, EU EMC Directive 2014/30/EU (replacing 89/336/EEC), Low Voltage Directive 2014/35/EU (replacing 73/23/EEC) |
| Generator Output | Continuous up to 160 A / Transient up to 1200 A |
| Magnetic Field Levels | Continuous — Class 1 (1 A/m) to Class 5 (100 A/m), Custom |
| Open-Circuit Voltage | ≤4 V |
| Coil Dimensions | 1035 mm × 1035 mm × 35 mm (square loop) |
| Antenna Factor | 0.87 |
| Inductance | ~0.6 µH |
| DC Resistance | 670 µΩ |
| Height Adjustment Range | 0–2000 mm |
| Weight (Net) | 28 kg |
| Dimensions (W×H×D) | 385 × 400 × 215 mm |
| Power Supply | AC 198–264 V, 50/60 Hz |
| Communication Interface | RS-232 / RS-485 (DB9) |
| Safety Features | Thermal cutoff, magnetic overcurrent protection, safety interlock port, fanless design |
| Brand | PMM |
|---|---|
| Origin | Italy |
| Model Range | L2-16A, L3-32, L3-64, L2-D, L3-100, L3-500, L1-150, L5-50-1-01, L5-70-1-01, L5-100-1-01 |
| Current Ratings | 2 A to 350 A |
| Impedance Networks | 50 Ω // (5 Ω + 50 µH), 150 Ω, 50 Ω // (1 Ω + 5 µH) |
| Compliance | CISPR 16-1-2, CISPR 22, CISPR 25, ISO 7637-2, MIL-STD-461G (RS103, CS106), EN 55032/55035 |
| Control Interface | TTL/RS-232 compatible with PMM receivers (e.g., PRM9000, EMI Suite) |
| Brand | PMM |
|---|---|
| Origin | Italy |
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | C106209 |
| Price Range | USD 1,400 – 7,000 |
| Brand | PMM |
|---|---|
| Origin | Italy |
| Product Types | Triaxial Loop Antenna (RF-300), BALUN Transformer, Coupling/Decoupling Network (CDN), Impedance Stabilization Network (ISN), Lamp Simulator (Incandescent, Fluorescent, LED Driver Emulator), V-Network, Conical Metallic Shielding Enclosure |
| Import Status | Imported |
| Brand | PMM |
|---|---|
| Origin | Italy |
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | BC-01 |
| Frequency Range | 30–200 MHz |
| Impedance | 50 Ω |
| Polarization | Linear (Vertical/Horizontal) |
| VSWR | ≤2.5:1 (typ. across band) |
| Gain | −2.5 to +1.5 dBi (frequency-dependent) |
| Connector | N-type (female) |
| Max Input Power | 1 kW (CW), 5 kW (peak, pulsed) |
| Operating Temperature | −40 °C to +70 °C |
| Brand | PMM |
|---|---|
| Origin | Italy |
| Manufacturer Type | Authorized Distributor |
| Import Status | Imported |
| Model | LP-02 |
| Pricing | Available Upon Request |
| Frequency Range | 200–3000 MHz |
| Brand | PMM |
|---|---|
| Origin | Italy |
| Manufacturer Type | Authorized Distributor |
| Import Status | Imported |
| Model | PMM 6000S/10 |
| Price | Upon Request |
| RF Signal Generator | PMM 3010 |
| Power Amplifier | PMM 6000N (10/15 W, optional 40/80/100 W) |
| Power Meter | PMM 6600 or 6600D, Boonton 4220/4230/9200 |
| Coupling Devices | M3-16 (220 V / 16 A CDN), M3-32, M4-32, M5-32, CDN-S, CDN-F, CDN-T series |
| Probes | F-120-9A Absorbing Probe, F-33 Current Probe, F-203I-23 Absorbing Clamp |
| Attenuators | ATT-25W (6 dB, 25 W), ATT-30W (6 dB, 30 W) |
| Software | SW-06 Immunity Control & Calibration Suite |
| Frequency Range | 100 Hz – 1000 MHz |
| Test Level Range | 1–100 V (0.01 V resolution) |
| Modulation | AM (400 Hz–1 kHz internal/external), Pulse (1 Hz / 200 Hz), Depth: 1–100% |
| Dwell Time | 100 ms – 10 s |
| Calibration Support | Automatic CDN/probe/clamp calibration with traceable correction factors |
| Compliance Standards | IEC/EN 61000-4-6, IEC/EN 61000-4-3, MIL-STD-461G RS103, DO-160 Section 20, ISO 11452-4/2013 |
| Brand | PMM |
|---|---|
| Origin | Italy |
| Model | RA-01, RA-01-HV, BC-01, LP-02, LP-03, AS-02, AS-03 |
| Frequency Range | 9 kHz – 6 GHz |
| Max Input Power | 1000 W (LP-series), 300 W (Horn) |
| Gain | up to 18.77 dBi |
| Compliance | CISPR 16-1-4, ANSI C63.4, IEC 61000-4-3 |
| Frequency Range | 9 kHz–3 GHz (6600/6600D: up to 1 GHz) |
|---|---|
| Power Range | 100 nW–1 W (-40 dBm to +30 dBm) |
| Linearity | ±0.2 dB (typ.) |
| Measurement Units | dBm, W, Vrms |
| Interface | USB 1.0/1.1/2.0 (6600/6600D: RS-486) |
| Probe Integration | Integrated sensor and meter architecture |
| RMS Detection | True RMS, >70 dB dynamic range |
| Bidirectional Measurement | Forward & reverse power |
| Calibration | Reference-source-free operation |
| Brand | PMM |
|---|---|
| Origin | Italy |
| Model | 9010 |
| Frequency Range | 10 Hz – 30 MHz (expandable to 2.7 GHz) |
| RF Input | 50 Ω, BNC female |
| Attenuator | 0–35 dB (5 dB steps) |
| Preamp Gain | 10 dB / 20 dB (post-preselector) |
| Max Input (non-destructive) | 137 dBµV (1 W sine), 97 dBµV/MHz (pulse spectral density) |
| Preselector | 7 filters (<9 kHz, 9–150 kHz, 150 kHz–500 kHz, 500 kHz–3 MHz, 3–10 MHz, 10–20 MHz, 20–30 MHz) |
| IF Bandwidths (6 dB) | 10 Hz, 100 Hz, 1 kHz, 10 kHz |
| Detector Modes | Peak, Quasi-Peak, RMS, Average (parallel) |
| Measurement Time | 1 ms – 30 s |
| Display Units | W, V, A, dBm, dBµV, dBµA, dBpW, dBµV/m, dBµA/m |
| Span/Division | 1–100 kHz |
| IF Resolution | 0.1–10 kHz |
| Amplitude Accuracy (S/N > 20 dB) | ±1.0 dB (1–150 kHz), ±0.8 dB (150 kHz–30 MHz) |
| Tracking Generator | 10 Hz–50 MHz, output 60–90 dBµV (0.1 dB steps), accuracy ±0.3 dB (9 kHz–30 MHz) |
| Demodulation | AM/FM |
| Interfaces | RS-232, USB, IEEE-488 (optional), wireless (optional) |
| Operating Temperature | 0–40 °C |
| Power | 12 V DC |
| Dimensions | 235 × 105 × 330 mm |
| Weight | 5 kg |
Show next