ERWEKA
Filter
Showing all 6 results
| Brand | ERWEKA |
|---|---|
| Origin | Germany |
| Model | DT950 |
| Instrument Type | Offline Dissolution Tester |
| Dissolution Vessels | Compliant with USP, JP, and EP Pharmacopoeias |
| Temperature Range | +5 °C to 40 °C |
| Temperature Accuracy | Compliant with USP/JP/EP |
| Rotational Speed Range | Compliant with USP/JP/EP |
| Paddle Wobble Amplitude | Compliant with USP/JP/EP |
| Basket Wobble Amplitude | Compliant with USP/JP/EP |
| Water Bath Temperature Control Range | +5 °C to 40 °C |
| Water Bath Temperature Resolution | ±1 °C |
| Water Bath Volume | Compliant with USP/JP/EP |
| Brand | ERWEKA |
|---|---|
| Origin | Germany |
| Model | EASYCHECK |
| Hardness Range | 3–500 N (extendable to 1000 N, ±1 N) |
| Diameter/Length Range | 2–28 mm (±0.05 mm) |
| Thickness Range | 2.0–10.0 mm (±0.05 mm) |
| Weight Range | 20.0 mg–64 g (±0.1 mg) |
| Test Speed | 0.1–3 mm/s (constant velocity) |
| Force Ramp Rate | 10–200 N/s (constant pressure) |
| Dimensions (W×D×H) | 350 × 375 × 240 mm |
| Weight | 15 kg |
| Power Supply | 100–240 V, 50–60 Hz |
| Interfaces | SD card, Ethernet (LAN), RS-232C*, 1× USB-A, 1× USB-B (for printer or AutoCal 2.0) |
| Calibration | Statistical weight calibration included |
| Data Storage | Up to 100 methods onboard |
| Compliance | Fully compliant with USP <1217>, Ph. Eur. 2.9.8, ASTM D7848, and FDA 21 CFR Part 11 (electronic records & signatures) |
| Brand | ERWEKA |
|---|---|
| Origin | Germany |
| Model Series | GTL / GT / GTB |
| Instrument Type | Multi-Method Powder Flowability Tester |
| Measurable Parameters | Angle of Repose, Difference Angle, Dispersibility |
| Maximum Sample Vessel Capacity | 480 mL |
| Minimum Sample Volume | 100 mL |
| Brand | ERWEKA |
|---|---|
| Origin | Germany |
| Model | SVM II |
| Instrument Type | Tap Density Tester |
| Application | Powder Bulk & Tapped Density Measurement |
| Compliance | USP <616>, EP 2.9.34, USP <1174>, ASTM B527-15, DIN EN ISO 787-11:1995 |
| Temperature Range | +10 °C to +30 °C |
| Dimensions (H×W×D) | 196.5 × 300 × 326 mm |
| Weight | 17 kg |
| Power Supply | 115/230 V, 50/60 Hz |
| Protection Class | IP21 / IEC 60529, Class I / EN 61140 |
| Test Stations | 1 or 2 (configurable via SwitchPlate system) |
| Tap Frequency & Height | USP Method 1: 300 ±15 taps/min, 14 ±2 mm |
| USP Method 2 | 250 ±15 taps/min, 3 ±0.2 mm |
| Data Storage | Last 10 test reports (local) |
| Interfaces | USB-A, USB-B, RS232, Ethernet |
| Report Export | XML/PDF via Export Manager software |
| Brand | ERWEKA |
|---|---|
| Origin | Germany |
| Model | TAR II |
| Compliance | USP <1216>, EP 2.9.7, JP 6.05 |
| Display | 7-inch high-resolution touchscreen |
| Test Stations | Configurable for 1 or 2 stations (field-upgradable) |
| Test Modes | Rotation count, time-based, and fully automated USP-compliant mode |
| Data Integration | Direct balance interface with real-time weight acquisition |
| Software | TestAssist™ embedded control & reporting suite |
| Qualification Support | Built-in IQ/OQ verification routines |
| Reporting | Auto-generated, printable test reports (last 10 stored) |
| User Assistance | Context-sensitive on-screen help & service diagnostics |
| Brand | ERWEKA |
|---|---|
| Origin | Germany |
| Model | ZT730 |
| Test Stations | 1 (ZT731) or 2 (ZT732) |
| Touchscreen | 7-inch capacitive LCD |
| Memory Capacity | 200 product/method profiles + ~4 GB for ≥1,000,000 test results |
| Stroke Speed | 30 ± 1 mm/min |
| Stroke Height | 55 ± 2 mm |
| Heating | Integrated flow-through heater with PT100 temperature sensor |
| Temperature Control Range | Compliant with USP <701> and Ph. Eur. 2.9.1 (typically 37.0 ± 0.5 °C) |
| AirBasket Types | Type A (6 tubes) or Type B (3 tubes + disc holder) |
| Power Supply | 230 VAC ±10%, 50–60 Hz / 115 VAC ±10%, 50–60 Hz |
| Max. Power Consumption | 1750 W |
| Interfaces | USB-A, USB-B, Ethernet (LAN) |
| Printer Support | Oki/HP PCL-compatible laser printers |
| Dimensions (W×D×H) | 397 × 365 × 667 mm |
| Weight | 30–35 kg |