Empowering Scientific Discovery

AZ7721/AZ77231 Wall-Mounted CO₂ and Temperature Monitor

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Origin Zhejiang, China
Manufacturer Type Distributor
Origin Category Domestic (China)
Model AZ7721/AZ77231 Wall-Mounted CO₂ and Temperature Monitor
Price Range USD 7,000–14,000
Instrument Type Fixed Installation (Wall-Mounted)
Target Analyte Carbon Dioxide (CO₂) and Ambient Temperature
Measurement Principle Non-Dispersive Infrared (NDIR) Spectroscopy
CO₂ Range (AZ7721) 0–2,000 ppm (±50 ppm ±5% accuracy)
CO₂ Range (AZ77231) 0–5,000 ppm (±30 ppm ±5% accuracy)
CO₂ Resolution 1 ppm
Temperature Range −10 °C to +60 °C (14 °F to 140 °F), Accuracy: ±0.6 °C / ±0.9 °F
Warm-up Time 30 seconds
Operating Temp. 0–50 °C, RH: 0–95% non-condensing
Storage Temp. −20–60 °C, RH: 0–99% non-condensing
Dimensions 130 × 85 × 60 mm (L×W×T)
Output Interface RS232 serial port (2-second data interval)
Display Backlit LCD
Alarms Visual and audible threshold alerts
Units °C/°F toggle
Optional Accessories RS232 communication software, data cable

Overview

The AZ7721/AZ77231 Wall-Mounted CO₂ and Temperature Monitor is a fixed-position environmental sensing instrument engineered for continuous, real-time monitoring of carbon dioxide concentration and ambient temperature in indoor air quality (IAQ) applications. It employs non-dispersive infrared (NDIR) spectroscopy—a well-established, interference-resistant optical detection method—where CO₂ molecules absorb specific wavelengths of infrared radiation (centered at 4.26 µm). The instrument’s dual-channel NDIR sensor measures absorbance intensity against a reference wavelength to derive CO₂ partial pressure, while a high-stability thermistor provides concurrent temperature readings. Designed for permanent wall installation in offices, classrooms, laboratories, cleanrooms, HVAC ducts, and controlled-environment facilities, the device delivers reliable baseline IAQ data without requiring manual intervention or field calibration under normal operating conditions. Its compact form factor (130 × 85 × 60 mm), low power consumption, and 30-second thermal stabilization enable rapid deployment and integration into building management systems (BMS) or standalone logging networks.

Key Features

  • True wall-mounted architecture with integrated mounting bracket—no external enclosure or enclosure modification required.
  • Dual-range CO₂ measurement capability: AZ7721 optimized for standard occupancy monitoring (0–2,000 ppm); AZ77231 extends dynamic range for higher-concentration environments such as fermentation rooms or industrial process areas (0–5,000 ppm).
  • 1 ppm resolution across full scale, supporting detection of subtle ventilation inefficiencies or occupancy-driven CO₂ accumulation trends.
  • Backlit LCD display with real-time dual-parameter readout (CO₂ ppm + temperature in °C/°F), user-selectable unit toggle, and configurable alarm thresholds.
  • Visual (LED flash) and audible (buzzer) alert outputs triggered when CO₂ exceeds user-defined limits—critical for demand-controlled ventilation (DCV) compliance and occupant health safeguards.
  • RS232 serial interface with fixed 2-second data output interval—compatible with SCADA systems, PLCs, and custom data acquisition platforms using ASCII protocol.
  • Low-drift NDIR optical path design with temperature-compensated detector, minimizing zero and span drift over extended operational cycles within specified environmental envelopes.

Sample Compatibility & Compliance

The AZ7721/AZ77231 is validated for gaseous sampling of ambient indoor air only. It is not suitable for corrosive, condensing, particulate-laden, or high-humidity (>95% RH non-condensing) airstreams without pre-conditioning. The instrument complies with general functional safety expectations for Class II indoor electrical equipment per IEC 61000-6-3 (EMC emission) and IEC 61000-6-1 (immunity). While not certified to ISO 16000-8 or EN 13779 for formal IAQ reporting, its NDIR methodology aligns with the physical principles referenced in ASTM D6196-22 (Standard Practice for Selecting Environmental Monitoring Methods for Air Quality) and supports documentation requirements under LEED v4.1 EQ Credit: Indoor Air Quality Assessment. Data integrity meets basic GLP traceability needs when paired with timestamped RS232 logging and operator-maintained calibration records.

Software & Data Management

The optional RS232 communication software enables local PC-based data capture, time-series visualization, and CSV export. Each transmitted frame includes CO₂ value (ppm), temperature (°C or °F), and UTC timestamp—enabling correlation with HVAC schedules or occupancy logs. The firmware implements no internal memory; all historical data must be captured externally. Audit trail functionality is limited to host-system logging; the device itself does not support FDA 21 CFR Part 11-compliant electronic signatures or user-access controls. For regulated environments, integration with validated third-party data acquisition software (e.g., LabVIEW with NI-VISA drivers or Python-based PySerial implementations) is recommended to satisfy ALCOA+ (Attributable, Legible, Contemporaneous, Original, Accurate) data governance expectations.

Applications

  • Occupancy-based ventilation control in smart buildings and educational institutions.
  • Baseline IAQ verification during ASHRAE 62.1-compliant commissioning and re-commissioning.
  • Environmental monitoring in pharmaceutical stability chambers and incubator anterooms where CO₂ buildup may affect sample integrity.
  • Process air monitoring in food packaging lines utilizing modified atmosphere (MAP) environments.
  • Field validation of HVAC system performance before and after filter replacement or duct cleaning.
  • Research-grade CO₂ trend logging in greenhouse climate studies (when installed in representative interior zones).

FAQ

Is this instrument suitable for outdoor or duct-mounted installation?
No—only indoor, non-condensing environments within the specified operating range (0–50 °C, 0–95% RH) are supported. Duct mounting requires an inline probe adapter and airflow conditioning to prevent laminar flow artifacts and moisture ingress.
Does the device include automatic baseline correction (ABC) or field calibration capability?
No—calibration is factory-performed using certified gas standards. Field recalibration requires NIST-traceable CO₂ reference gas and proprietary service tools; end users should follow documented procedures in the operator manual.
Can the RS232 output be interfaced directly with Modbus RTU or BACnet MS/TP networks?
Not natively—the unit outputs ASCII-formatted serial data only. Protocol translation requires an external gateway or middleware layer (e.g., a Modbus-to-serial converter with ASCII parsing logic).
What is the expected service life of the NDIR sensor under continuous operation?
Based on accelerated lifetime testing per IEC 60721-3-3, the optical module maintains metrological stability for ≥5 years under nominal conditions (25 °C, 50% RH, 24/7 operation), assuming no exposure to silicone vapors, solvents, or mechanical shock.
Is firmware upgrade supported via RS232?
No—firmware is固化 (hard-coded) and not field-upgradable. Functional updates require hardware revision and manufacturer reprogramming prior to shipment.