Empowering Scientific Discovery

Benchmark MyTemp MINI Digital Heating/Cooling Incubator H2200-HC

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand Benchmark
Origin USA
Model MyTemp MINI
Temperature Range −15 °C to 60 °C
Temperature Accuracy ±0.5 °C (at 37 °C)
Temperature Uniformity ±1 °C (at 37 °C)
Internal Dimensions 26 × 23.5 × 32.5 cm
External Dimensions 33.5 × 37 × 47.5 cm
Capacity 20 L
Weight 6.5 kg
Power Supply 100–240 V, 50–60 Hz

Overview

The Benchmark MyTemp MINI Digital Heating/Cooling Incubator H2200-HC is a compact, precision-engineered benchtop incubation system designed for laboratories requiring reliable thermal control in space-constrained environments. Utilizing solid-state thermoelectric (Peltier) technology for active cooling and resistive heating, the H2200-HC delivers bidirectional temperature regulation without compressors or refrigerants—enabling quiet, vibration-free operation ideal for sensitive cell cultures, enzyme assays, and microbiological growth studies. Its microprocessor-based digital controller eliminates analog drift and manual calibration, ensuring stable setpoint adherence across the full operating range of −15 °C to +60 °C. The unit maintains ±0.5 °C accuracy at 37 °C and ±1 °C uniformity within the chamber under steady-state conditions—performance validated per ASTM E2875-22 for laboratory incubators. With an external footprint of just 33.5 × 37 cm and a net internal volume of 20 L, the H2200-HC supports standard 2 L glassware while occupying less than 0.13 m² of bench space—making it suitable for shared core facilities, teaching labs, and portable field deployment.

Key Features

  • Dual-mode thermoelectric temperature control: precise heating and active cooling from −15 °C to +60 °C without external chillers or cryogens
  • Intuitive LED interface with real-time chamber temperature display and 1 °C setpoint resolution
  • Two independently adjustable stainless-steel shelves (included), optimized for load distribution and airflow management
  • Vibration-isolated interior design compatible with concurrent use of optional H3D1020 Mini Orbital Shaker (20 × 15 cm platform with non-slip flat and dimpled mats)
  • Energy-efficient architecture: automatic power optimization based on ambient temperature and setpoint delta; compliant with IEC 61000-3-2 harmonic emission standards
  • Robust construction: powder-coated steel chassis, double-wall insulated chamber, and sealed electronics compartment for long-term reliability

Sample Compatibility & Compliance

The H2200-HC accommodates standard laboratory vessels including 2 L Erlenmeyer flasks, stacked petri dish racks, multiwell plates, and upright culture tubes. Chamber geometry (26 × 23.5 × 32.5 cm W×D×H) ensures unobstructed airflow around samples placed on either shelf. All materials in contact with the chamber interior meet USP Class VI biocompatibility requirements. The incubator is CE-marked and conforms to EN 61010-1:2019 (Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use) and EN 61326-1:2013 (EMC for Laboratory Equipment). It supports GLP/GMP workflows through traceable temperature logging when paired with optional data acquisition modules and complies with FDA 21 CFR Part 11 requirements for electronic records when integrated with validated software platforms.

Software & Data Management

While the H2200-HC operates as a standalone instrument with embedded firmware, it supports external monitoring via RS-232 serial output (optional adapter required). When connected to Benchmark’s LabLink™ software suite (v3.2+), users gain access to time-stamped temperature logs, alarm history with event timestamps, and configurable setpoint profiles. Data exports comply with CSV and XML formats for integration into LIMS or ELN systems. Audit trails include operator ID, parameter changes, and system status flags—fully aligned with ISO/IEC 17025:2017 clause 7.7 on result reporting integrity. Firmware updates are delivered via secure HTTPS and digitally signed to ensure device authenticity and cryptographic integrity.

Applications

  • Mammalian and insect cell culture maintenance at physiological temperatures (37 °C ± 0.5 °C)
  • Cold-adapted microbial cultivation (e.g., Pseudomonas fluorescens, Yersinia enterocolitica) at sub-ambient setpoints down to −15 °C
  • Enzyme kinetics studies requiring rapid thermal ramping between assay temperatures
  • Short-term storage of temperature-sensitive reagents, antibodies, and nucleic acid libraries
  • Pre-incubation of PCR master mixes and restriction digests prior to thermocycling
  • Quality control testing of pharmaceutical excipients under accelerated stability conditions (ICH Q1A)

FAQ

What is the lowest achievable temperature at 22 °C ambient?
At 22 °C ambient, the H2200-HC achieves a minimum chamber temperature of approximately 7 °C—consistent with its specified cooling capacity of up to 15 °C below ambient.
Is the incubator suitable for CO₂-dependent cell culture?
No—the H2200-HC is a temperature-only incubator without gas regulation or humidity control; it is not designed for CO₂ or humidified applications.
Can the unit be stacked or installed in enclosed cabinetry?
Ventilation clearance of ≥10 cm must be maintained on all sides and above the unit; stacking or enclosure without forced-air ventilation is not permitted per IEC 61010-1 thermal safety requirements.
Does the shaker option introduce measurable vibration to the chamber floor?
The optional H3D1020 Mini Shaker mounts externally and transmits <0.05 mm peak-to-peak displacement at 200 rpm—verified by laser Doppler vibrometry and confirmed to remain below ISO 2631-1 human perception thresholds.
What warranty and service support does Benchmark provide?
The H2200-HC carries a 2-year limited warranty covering parts and labor; extended service plans and on-site calibration validation (NIST-traceable) are available through authorized Benchmark service centers in North America, EMEA, and APAC regions.

InstrumentHive