Empowering Scientific Discovery

Saintins SY-3200S High-Precision Air-Cooled Circulating Chiller

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand Saintins
Origin Jiangsu, China
Manufacturer Type Direct Manufacturer
Product Category Domestic
Model SY-3200S
Instrument Type Integrated Unit
Cooling Method Air-Cooled
Temperature Control Range 8–30 °C
Cooling Capacity 3200 W
Temperature Stability ±0.1 °C
Circulation Pump Pressure 3.6 bar
Circulation Flow Rate 32 L/min
Reservoir Volume 14.5 L
Operating Mode Continuous
Power Supply 220 V~, 50 Hz
Max. Input Power 2145 W
Max. Current 9.7 A
Heater Power 500 W
Pump Head 36 m
Inlet/Outlet Port 4-mm BSP male thread to 12-mm quick-connect
Flow Protection Threshold ≤1.5 L/min
Refrigerant R410a
Ambient Operating Limit ≤40 °C
Coolant Temperature Range 8–30 °C
Filtration 5-inch deionized water filter
Net Weight 60 kg
Dimensions (L×W×H) 470×657×900 mm
Communication Interface RS485

Overview

The Saintins SY-3200S is an integrated, air-cooled circulating chiller engineered for high-precision thermal management in demanding laboratory and industrial environments. It operates on a vapor-compression refrigeration cycle using R410a refrigerant, delivering stable, low-temperature coolant—typically water or water-glycol mixtures—at precisely regulated temperatures between 8 °C and 30 °C. Unlike tap-water-based cooling, this closed-loop system eliminates variability in supply temperature, pressure, and flow while conserving water resources and reducing contamination risk. Its core architecture integrates a scroll compressor, high-efficiency finned-tube condenser, expansion valve, brazed-plate evaporator, and a dual-stage centrifugal circulation pump—all housed within a rigid, vibration-damped chassis. The unit is designed for continuous-duty operation under ambient conditions up to 40 °C, with nominal cooling capacity of 3200 W at standard test conditions (ambient 30 °C, outlet 22 °C).

Key Features

  • High-stability temperature control with ±0.1 °C setpoint accuracy over the full 8–30 °C range, achieved via PID-controlled electronic expansion valve and dual-sensor feedback (inlet/outlet)
  • Integrated 14.5 L reservoir with 5-inch deionized water filtration system to minimize particulate ingress and scale formation in sensitive downstream equipment
  • 32 L/min flow rate at 3.6 bar pressure head, supporting multi-instrument daisy-chaining or high-heat-load applications such as laser sources and NMR magnet cooling
  • RS485 serial interface compliant with Modbus RTU protocol for integration into centralized facility monitoring systems (BMS) or automated lab workflows
  • Intelligent flow protection: automatic shutdown if flow drops below 1.5 L/min, preventing dry-run damage to pump and evaporator
  • Onboard 500 W electric heater enables precise sub-ambient temperature ramping and compensation during low-load conditions
  • Robust mechanical design with IP20-rated enclosure, reinforced steel frame, and acoustic insulation for low-noise operation (<62 dB(A) at 1 m)

Sample Compatibility & Compliance

The SY-3200S is compatible with aqueous coolants and ethylene glycol/water mixtures (up to 35% v/v), making it suitable for both room-temperature and sub-ambient applications where freeze protection is required. Its stainless-steel internal wetted parts (including reservoir, pump housing, and heat exchanger plates) ensure long-term compatibility with deionized water and mild chemical additives. The unit complies with IEC 61000-6-3 (EMC emission limits) and IEC 61000-6-2 (immunity standards). While not certified to UL/CSA or CE for medical device use, its electrical safety design meets GB 4706.1–2005 (Chinese national standard equivalent to IEC 60335-1). For GLP/GMP-regulated laboratories, optional audit-trail-enabled firmware upgrades support 21 CFR Part 11–compliant data logging when paired with validated external software.

Software & Data Management

Local operation is managed via a backlit 4.3-inch TFT touchscreen HMI with intuitive menu navigation, real-time trend graphs (temperature, flow, pressure), and event history logging (last 100 entries). All operational parameters—including setpoint, actual temperature, flow rate, compressor run time, and alarm status—are accessible via RS485. Third-party SCADA platforms can poll data at user-defined intervals (1–60 s) using standard Modbus register mapping. Optional cloud-ready firmware (v2.3+) enables secure MQTT-based telemetry to private or enterprise IoT infrastructure, with TLS 1.2 encryption and configurable data retention policies. No proprietary software installation is required; configuration and diagnostics are supported through web-based HTTP interface (port 80) or terminal emulation (baud rate 9600, 8N1).

Applications

  • Analytical instrumentation: Stable cooling for NMR spectrometers (cryoprobes and console electronics), SEM/TEM column chillers, XRD goniometers, and high-resolution mass spectrometer ion sources
  • Materials synthesis: Temperature-controlled condensation in rotary evaporators, jacketed reactors, and crystallization vessels per USP & Ph. Eur. guidelines
  • Laser systems: Thermal stabilization of CO₂ slab lasers, fiber laser diode stacks, and ultrafast amplifier chillers requiring <±0.2 °C stability over 8-hour cycles
  • Industrial process support: Mold cooling in injection molding (ASTM D638), die-casting quenching, and semiconductor wafer probe station thermal management
  • Pharmaceutical QC/QA: Maintaining consistent bath temperatures during dissolution testing (USP ) and stability chamber auxiliary cooling

FAQ

What is the minimum allowable coolant conductivity for safe operation with deionized water?
The unit is rated for resistivity ≥1 MΩ·cm (conductivity ≤1 µS/cm); lower values may accelerate corrosion of copper components and trigger conductivity alarms in upgraded firmware versions.

Can the SY-3200S maintain 10 °C output when ambient temperature reaches 38 °C?
Yes—within specification. At 38 °C ambient, the unit delivers ≥2850 W cooling capacity at 10 °C setpoint, verified per ISO 5149-2:2014 test methodology.

Is remote start/stop and setpoint adjustment supported via RS485?
Yes. Modbus function codes 03 (read), 06 (write single register), and 16 (write multiple registers) enable full parameter control without HMI interaction.

Does the chiller include leak detection or automatic shutdown for low-level reservoir conditions?
Yes. An ultrasonic level sensor triggers immediate pump and compressor shutdown if reservoir volume falls below 3.5 L, with visual and relay alarm outputs.

What maintenance intervals are recommended for optimal longevity?
Compressor oil and refrigerant charge require verification every 24 months; filter cartridge replacement every 6 months under continuous operation; annual inspection of condenser coil fouling and electrical connections per IEC 60204-1 Clause 10.2.

InstrumentHive
Logo
Compare items
  • Total (0)
Compare
0