Shimadzu GC-2010 Plus Used Gas Chromatograph
| Brand | Shimadzu |
|---|---|
| Origin | Japan |
| Model | GC-2010 Plus |
| Category | Gas Chromatograph |
| Instrument Age | 5–7 years |
| Warranty | 6 months |
| Temperature Range | Ambient +4°C to 450°C (down to –50°C with optional liquid CO₂ cooling) |
| Oven Cooling Rate | ~3.4 min (450°C → 50°C) |
| Max Programmed Ramp Rate | 250°C/min |
| Column Oven Volume | 13.7 L |
| Temperature Accuracy | ±1% of setpoint (calibratable to 0.01°C) |
| Temperature Uniformity | <2°C deviation across 200 mm diameter circle at 30 mm from chamber wall |
| Ambient Drift Compensation | <0.01°C/°C |
| Max Independent Inlet Units | 3 |
| Max Independent Detector Units | 4 |
| FID Detection Limit | 1.5 pgC/s (n-dodecane) |
| Dynamic Range (FID) | 10⁷ |
| Advanced Flow Control (AFC) | Pressure range 0–970 kPa |
| Dimensions (Oven) | 280 × 175 × 280 mm |
| Dimensions (Main Unit) | 515 × 530 × 440 mm |
| Weight (FID configuration) | 30 kg |
| Power | AC 230 V ±10%, 2600 VA, 50/60 Hz |
Overview
The Shimadzu GC-2010 Plus is a high-performance, modular gas chromatograph engineered for precision, reproducibility, and regulatory-compliant operation in QC, environmental, food safety, and pharmaceutical laboratories. Based on Shimadzu’s proven dual-microprocessor architecture, the system employs advanced flow control (AFC) technology to deliver precise, real-time regulation of carrier gas pressure and flow—enabling constant average linear velocity (ALV) throughout temperature-programmed runs. Its oven design features optimized air circulation and an independent high-efficiency cooling fan, achieving rapid thermal stabilization and a documented cool-down time of ≤3.4 minutes from 450°C to 50°C. The instrument supports up to three independently temperature-controlled inlet modules—including split/splitless (SPL-2010 Plus), on-column (OCI/PTV-2010), and direct injection (WBI-2010 Plus)—and accommodates up to four detectors with individual thermal management. This configuration flexibility ensures method robustness across diverse sample matrices and analyte volatility profiles.
Key Features
- Advanced Flow Controller (AFC) with dual-mode operation: pressure- or flow-based control, programmable in up to 7 segments with ramp rates from –400 to +400 kPa/min or mL/min, supporting ALV maintenance during temperature ramps
- Rapid-cool oven with enhanced airflow design and low-noise fan—validated for ≤3.4 min cooldown (450°C → 50°C)
- Integrated backflush capability to terminate analysis early upon target elution, reducing cycle time and column contamination
- Detector分流 system enabling parallel detection of eluting compounds across multiple detectors (e.g., FID + ECD), improving compound identification confidence
- Comprehensive self-diagnostic suite monitoring temperature sensors, gas pressures, valve actuation, and heater status—generating event logs for preventive maintenance planning
- Carrier gas conservation mode: automatically reduces split flow from 50 mL/min to 5 mL/min post-injection, lowering operational cost without compromising method integrity
- Safety-engineered hydrogen fittings with left-hand (reverse) thread design and directional labeling to prevent misconnection
Sample Compatibility & Compliance
The GC-2010 Plus is routinely deployed in laboratories performing regulated analyses under ISO/IEC 17025, USP , EPA Methods 502.2, 624, 8021B, and Chinese standard NY/T 761–2008 for multiresidue pesticide screening. Its validated temperature accuracy (±1% of setpoint, calibratable to 0.01°C) and uniformity (<2°C across defined zone) meet ASTM E1510 and ICH Q2(R2) requirements for system suitability. The FID detector delivers a documented detection limit of 1.5 pgC/s (n-dodecane) and a dynamic range of 10⁷—sufficient for trace-level hydrocarbon quantitation per ASTM D3606 and EN 15407. All hardware configurations support GLP/GMP data integrity when operated with compliant chromatography data systems (CDS) meeting FDA 21 CFR Part 11 audit trail and electronic signature requirements.
Software & Data Management
The system operates with Shimadzu’s GCsolution software (v2.8 or later), which provides full instrument control, sequence management, peak integration, calibration curve generation, and report export in PDF, CSV, and XML formats. GCsolution includes built-in system suitability testing (SST) templates aligned with pharmacopeial standards, automated baseline noise/drift calculation, and customizable pass/fail criteria. Audit trail functionality records all parameter changes, user actions, and data modifications with timestamps and operator IDs—fully compliant with 21 CFR Part 11 when deployed on validated Windows OS environments. Raw data files (.gcd) are stored in a structured, non-proprietary binary format with embedded metadata (method, sequence, instrument ID, calibration history), ensuring long-term archival integrity and third-party software interoperability.
Applications
- Residual solvent analysis in pharmaceutical APIs and excipients (ICH Q3C)
- Volatile organic compound (VOC) profiling in drinking water (EPA 524.4, 624)
- Pesticide multiresidue screening in fruits, vegetables, and grains (NY/T 761–2008, EU SANTE/11312/2021)
- Fatty acid methyl ester (FAME) analysis in biodiesel (ASTM D6584)
- Hydrocarbon fingerprinting in petroleum products (ASTM D3606, D5186)
- Environmental forensics of chlorinated solvents and PAHs in soil extracts (EPA 8270)
FAQ
Is this unit certified for use in regulated environments?
Yes—the GC-2010 Plus meets mechanical and thermal specifications required for GLP and GMP applications when operated with validated software and documented IQ/OQ/PQ protocols.
What documentation accompanies the instrument?
Each unit ships with a full refurbishment report, post-calibration verification certificate (temperature, flow, detector response), electrical safety test record, and complete log of functional tests performed at SpectraLab Scientific.
Can the system be upgraded with additional detectors or inlets?
Yes—hardware modularity allows field installation of TCD, ECD, FPD, or NPD detectors, as well as OCI/PTV or WBI inlets, subject to physical bay availability and firmware compatibility.
Is remote diagnostics supported?
GCsolution supports secure remote access via VPN-configured network connections, enabling technical support teams to review system logs, verify configurations, and assist with troubleshooting without onsite visits.
What warranty and service options are available beyond the standard 6-month coverage?
Extended service agreements—including annual PM contracts, priority response SLAs, and loaner instrument provisions—are available upon request and can be tailored to laboratory throughput requirements.