Aolong AL-BP-9010A Wavelength Dispersive X-Ray Fluorescence Spectrometer
| Brand | Aolong |
|---|---|
| Origin | Liaoning, China |
| Manufacturer Type | Direct Manufacturer |
| Instrument Type | Simultaneous WDXRF Spectrometer |
| X-ray Tube Power | 400 W (50 kV / 8 mA) |
| Analyzed Elements | Na (11) to U (92), pre-configured for Ca, Fe, Si, Al, Na, Mg, S, K, Cl, P |
| Vacuum Level | < 7 Pa |
| Sample Diameter | 40 mm |
| Analysis Area | up to 35 mm |
| Sample Rotation Speed | 30 rpm |
| Stability (24 h) | ≤ 2% |
| Energy Resolution (Detector) | ≤ 40% |
| Counting Linearity | ≤ 1% |
| Measurement Time per Sample (incl. vacuum & exchange) | 2–5 min |
| Detector Types | Flow-proportional counter, sealed proportional counter, scintillation counter |
| Pulse Height Analyzer | 12-channel, 2048-channel MCA |
| Flow Gas Control Precision | ±0.01 kPa |
| Constant-Temperature Chamber Accuracy | ±0.1 °C |
| Power Supply | 220 V AC, 2 kVA stabilized |
Overview
The Aolong AL-BP-9010A is a high-performance simultaneous wavelength dispersive X-ray fluorescence (WDXRF) spectrometer engineered for precise, reproducible elemental quantification in industrial and research laboratories. Based on Bragg diffraction principles, the instrument employs crystal monochromators to separate characteristic X-ray emission lines by wavelength, enabling high spectral resolution and minimal peak overlap—critical for accurate multi-element analysis in complex matrices. Unlike energy dispersive (EDXRF) systems, WDXRF delivers superior detection limits, better inter-element interference correction, and higher precision for major, minor, and trace elements across diverse sample types. Designed for continuous operation in quality control (QC) and process analytical technology (PAT) environments, the AL-BP-9010A meets the rigorous demands of ISO/IEC 17025-accredited labs and supports compliance with ASTM C114, ASTM E1621, ISO 29581-2, and GB/T 223.79–2007 for metallurgical and cementitious materials.
Key Features
- Simultaneous multi-channel configuration supporting ≥10 fixed analysis channels for real-time quantification of key elements—including Ca, Fe, Si, Al, Na, Mg, S, K, Cl, and P—without sequential scanning delays.
- High-stability 400 W X-ray tube (50 kV / 8 mA) with thin beryllium window and closed-loop thermoelectric cooling; no external water supply required.
- Dual-vacuum chamber architecture: independent pump station with rapid evacuation (< 90 s) and sustained vacuum < 7 Pa in the measurement chamber.
- Precision flow-gas system using imported mass-flow controllers with pressure stability ±0.01 kPa, ensuring consistent detector gain and long-term counting stability.
- Automated sample rotation at 30 rpm and programmable sample positioning to minimize heterogeneity effects and enhance measurement reproducibility.
- Integrated constant-temperature chamber (±0.1 °C) for optical path and crystal alignment stability—essential for maintaining calibration integrity over extended runs.
- 12-channel, 2048-channel multichannel analyzers (MCA) with real-time pulse height discrimination, enabling robust background subtraction and peak deconvolution.
Sample Compatibility & Compliance
The AL-BP-9010A accommodates solid pressed pellets (40 mm diameter × 10 mm height), fused beads, and polished metal blocks. It is routinely deployed for routine analysis of cement raw meals, clinkers, and finished cements (per ASTM C114 and GB/T 176); ferrous and non-ferrous alloys (per ASTM E1086, GB/T 223 series); petroleum products (S, P quantification per ASTM D4294); and geological samples (rocks, ores, soils). All hardware and software components comply with IEC 61000-6-3 (EMC) and IEC 61010-1 (safety). The system supports audit-ready data handling under GLP/GMP frameworks, including electronic signatures, full audit trail logging, and 21 CFR Part 11–compliant user access controls when integrated with validated LIMS or ELN platforms.
Software & Data Management
The instrument operates with Aolong’s proprietary WDXRF control and analysis suite, compatible with Windows 10/11. The software provides fully automated qualitative identification (peak search, library matching, background smoothing), quantitative analysis via fundamental parameter (FP) and empirical calibration models, and matrix correction algorithms for absorption/enhancement effects. Cement-specific modules include real-time LSF/SM/IA ratio calculation, raw mix optimization feedback, and statistical process control (SPC) charting. Data export supports CSV, XML, and OPC-UA protocols; native DCS integration (Modbus TCP, OPC DA) enables seamless linkage with plant-wide automation systems. Remote diagnostics, self-calibration routines, and predictive maintenance alerts are embedded—reducing downtime and technician dependency.
Applications
Primary deployment domains include: cement manufacturing (raw material blending, kiln feed QC, clinker optimization), iron and steel production (slag composition, alloy grade verification), petrochemical refining (sulfur content monitoring in fuels), glass and ceramics R&D (batch formulation validation), mining and exploration (ore grade estimation), and academic materials science (thin-film thickness, coating composition). In cement plants, the AL-BP-9010A reduces offline lab turnaround from hours to minutes—enabling sub-hour feedback loops for raw mix adjustment. Its 2–5 minute cycle time (including vacuum and sample exchange) supports high-throughput batch analysis without compromising accuracy (RSD 1 wt%).
FAQ
What elements can the AL-BP-9010A quantify, and what is the detection limit range?
The system is configured for simultaneous analysis of Na (Z=11) through U (Z=92); typical detection limits range from 1–10 ppm for trace elements in homogeneous solids, depending on matrix, counting time, and line selection. Pre-installed calibrations cover CaO, SiO₂, Al₂O₃, Fe₂O₃, MgO, SO₃, K₂O, Na₂O, Cl⁻, and P₂O₅ in cementitious materials.
Does the instrument require liquid nitrogen or external cooling water?
No. The X-ray tube uses thermoelectric (Peltier) cooling; the vacuum pumps and electronics are air-cooled. No cryogens or tap-water connections are needed.
Can the AL-BP-9010A integrate with existing plant DCS or MES systems?
Yes. Native support for Modbus TCP and OPC UA ensures bidirectional data exchange with distributed control systems, SCADA platforms, and enterprise manufacturing execution systems.
Is method validation and regulatory documentation available for pharmaceutical or food-grade applications?
While primarily designed for industrial QC, the platform supports IQ/OQ/PQ protocols and can be qualified for regulated environments upon request; full validation packages—including uncertainty budgets and traceable CRM-based calibration reports—are available as optional services.
What maintenance intervals and service support options are offered?
Preventive maintenance is recommended every 12 months. Aolong provides remote firmware updates, on-site technical support, and spare-part logistics through its global distributor network. Extended warranty and service-level agreements (SLAs) are available.