PMM 6000S/10 Automated Immunity Test System

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand PMM
Origin Italy
Manufacturer Type Authorized Distributor
Import Status Imported
Model PMM 6000S/10
Price Upon Request
RF Signal Generator PMM 3010
Power Amplifier PMM 6000N (10/15 W, optional 40/80/100 W)
Power Meter PMM 6600 or 6600D, Boonton 4220/4230/9200
Coupling Devices M3-16 (220 V / 16 A CDN), M3-32, M4-32, M5-32, CDN-S, CDN-F, CDN-T series
Probes F-120-9A Absorbing Probe, F-33 Current Probe, F-203I-23 Absorbing Clamp
Attenuators ATT-25W (6 dB, 25 W), ATT-30W (6 dB, 30 W)
Software SW-06 Immunity Control & Calibration Suite
Frequency Range 100 Hz – 1000 MHz
Test Level Range 1–100 V (0.01 V resolution)
Modulation AM (400 Hz–1 kHz internal/external), Pulse (1 Hz / 200 Hz), Depth: 1–100%
Dwell Time 100 ms – 10 s
Calibration Support Automatic CDN/probe/clamp calibration with traceable correction factors
Compliance Standards IEC/EN 61000-4-6, IEC/EN 61000-4-3, MIL-STD-461G RS103, DO-160 Section 20, ISO 11452-4/2013

Overview

The PMM 6000S/10 Automated Immunity Test System is a fully integrated, standards-compliant platform engineered for conducted and radiated immunity testing in electromagnetic compatibility (EMC) laboratories. Based on the Coupling/Decoupling Network (CDN)-based injection method per IEC/EN 61000-4-6, the system delivers repeatable, traceable, and auditable test sequences across the 100 Hz to 1 GHz frequency band. It combines a precision RF signal generator (PMM 3010), broadband power amplifier (PMM 6000N, 10/15 W base configuration), calibrated coupling hardware (M3-16 CDN, F-series probes, attenuators), and the SW-06 control software into a single synchronized architecture. Designed for compliance verification against IEC/EN 61000-4-6 (conducted immunity), IEC/EN 61000-4-3 (radiated immunity via absorbing clamp or probe methods), MIL-STD-461G RS103, DO-160 Section 20 (avionics), and ISO 11452-4 (automotive component immunity), the 6000S/10 provides deterministic test execution without manual intervention—ensuring consistency across operator shifts and audit cycles.

Key Features

  • Full automation of immunity test sequences per IEC/EN 61000-4-6, including frequency sweep, level stepping, dwell time control, and modulation application (AM, pulse)
  • SW-06 software with built-in calibration factor management: supports user-defined 1 V, 3 V, and 10 V reference curves; stores correction data for CDNs, absorbing clamps (F-203I-23), and probes (F-120-9A, F-33)
  • Multi-instrument interoperability: native support for PMM 3010/3030, R&S SMX/SMY/SMT02/SMG, HP 8647/8656/8648 generators; PMM 6600/6600D, Boonton 4220/4230/9200 power meters
  • Scalable coupling infrastructure: M3-16 (220 V/16 A) included as standard; optional M3/M4/M5 CDNs (up to 380 V/32 A), CDN-S (shielded lines), CDN-F (floating/unstable lines), CDN-T (telecom lines)
  • Five programmable breakpoints with configurable warning messages and logging; up to eight independent scan profiles for multi-condition testing
  • GLP/GMP-aligned data integrity: full audit trail, timestamped parameter logs, report generation with embedded calibration metadata and instrument IDs
  • Local calibration mode: correction factors downloadable to PMM 3000-series internal memory for offline operation without PC dependency

Sample Compatibility & Compliance

The PMM 6000S/10 accommodates a wide range of EUT (Equipment Under Test) configurations—including AC/DC mains ports (single- and three-phase), shielded/unshielded signal lines, telecom interfaces, and automotive harnesses—via its modular CDN and probe ecosystem. All coupling devices are manufactured to IEC 61000-4-6 Annex A specifications and supplied with NIST-traceable calibration certificates. The system meets the measurement uncertainty requirements of ISO/IEC 17025 for accredited EMC labs and supports FDA 21 CFR Part 11 compliance when deployed with validated SW-06 installation and electronic signature protocols. Test reports include full traceability to calibration dates, instrument serial numbers, environmental conditions (optional sensor integration), and pass/fail status per test point—enabling direct submission to Notified Bodies.

Software & Data Management

SW-06 is a Windows-based control and reporting suite designed for regulatory-grade test documentation. It features dual-mode operation: automated sequence execution with real-time power monitoring and manual frequency/level override for diagnostic troubleshooting. Calibration tables are editable in CSV format and support interpolation between discrete points. All test parameters—including dwell time, modulation depth, step size, and breakpoint logic—are stored in XML-based project files with SHA-256 checksums for version integrity. Reports export to PDF with embedded digital signatures, watermarking, and hierarchical metadata (test plan ID, EUT ID, lab accreditation number). Software updates are delivered via secure HTTPS download with cryptographic signature verification. Minimum system requirements: Windows 10 (64-bit), 4 GB RAM, USB 2.0 or RS-232 interface, and NI GPIB-USB-HS adapter for legacy instrument control.

Applications

  • Pre-compliance and certification testing for industrial electronics, medical devices (IEC 60601-1-2), and IoT gateways
  • Automotive component validation per ISO 11452-4 (bulk current injection) and ISO 11452-8 (direct drive)
  • Aerospace avionics qualification per DO-160 Section 20 (conducted RF susceptibility)
  • Military equipment testing per MIL-STD-461G RS103 (radiated susceptibility, 10 kHz–40 GHz)
  • Power supply and motor drive immunity assessment under variable load conditions using M4/M5 CDNs
  • Research-grade immunity mapping of PCB-level interfaces using F-33 current probes and F-120-9A absorbing probes

FAQ

What standards does the PMM 6000S/10 support out-of-the-box?
IEC/EN 61000-4-6 (conducted immunity), IEC/EN 61000-4-3 (radiated immunity via clamp/probe), ISO 11452-4, DO-160 Section 20, and MIL-STD-461G RS103.
Can the system be upgraded to higher power levels?
Yes—the base 6000N amplifier (10/15 W) can be replaced with 40 W, 80 W, or 100 W variants; CDN modules scale from M1 (10 A) to M5 (32 A) for high-current applications.
Is SW-06 compliant with FDA 21 CFR Part 11?
When installed and validated per lab SOP—including electronic signatures, audit trail activation, and role-based access control—it meets Part 11 requirements for regulated industries.
Does the system support unattended overnight testing?
Yes—SW-06 includes job queuing, automatic error recovery, and email/SNMP alerts upon completion or failure.
How is calibration traceability maintained across instruments?
All CDNs, probes, and attenuators ship with ISO/IEC 17025-accredited calibration certificates; SW-06 enforces mandatory calibration date checks before test initiation.