Empowering Scientific Discovery

SVT Associates WP-15 Ultra-High Vacuum Cleaving and Passivation System

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand SVT Associates
Origin USA
Model WP-15
Cleaving Capacity 15 mm × 15 mm input wafer (10–32 mm square configurable) → 15 × 1 mm × 15 mm bar strips (bar width: 1–5 mm adjustable)
Vacuum Compatibility UHV (<1×10⁻⁹ Torr)
Flange 8-inch CF (ConFlat), bakeable to 180 °C (200 °C optional)
Sample Orientation Dual-axis alignment (X/Y-direction cleaving)
Loading Cassette-based, load-lock compatible
Integrated Option In-situ passivation deposition module (e.g., SiO₂, SiNₓ, Al₂O₃)

Overview

The SVT Associates WP-15 Ultra-High Vacuum Cleaving and Passivation System is a precision-engineered platform designed for semiconductor research and pilot-scale fabrication of edge-emitting laser diodes, photonic integrated circuits (PICs), and vertical-cavity surface-emitting laser (VCSEL) test structures. It implements mechanical cleaving under ultra-high vacuum (UHV) conditions—typically <1×10⁻⁹ Torr—to eliminate surface contamination, oxidation, and moisture adsorption during bar separation. Unlike atmospheric or rough-vacuum cleaving tools, the WP-15 integrates cleaving, bar stacking, and in-situ passivation within a single UHV-compatible chamber train, enabling direct transfer of freshly cleaved facets to a deposition module without air exposure. This workflow preserves native facet quality critical for high-reflectivity (HR) and anti-reflective (AR) coating adhesion, optical loss minimization, and reproducible cavity Q-factor measurement.

Key Features

  • UHV-compatible mechanical cleaving mechanism engineered for sub-micron facet parallelism and minimal chipping—optimized for III–V (InP, GaAs) and II–VI compound wafers.
  • Configurable bar geometry: accepts input wafers from 10 mm × 10 mm to 32 mm × 32 mm; outputs up to fifteen 15 mm-long bars with user-selectable widths (1–5 mm) via precision-ground scribe-and-break tooling.
  • Dual-axis sample stage allows cleaving along either crystallographic direction (e.g., [110] or [−110]) without re-mounting—essential for polarization-sensitive device evaluation.
  • Integrated cassette loader supports standardized 25-bar cassettes (customizable); enables batch handling and traceable bar indexing per GLP-compliant lab workflows.
  • 8-inch ConFlat (CF) main chamber flange rated for bakeout at 180 °C (200 °C optional), ensuring low outgassing rates and compatibility with standard UHV pumping stacks (ion pumps, NEG pumps, turbomolecular backing).
  • Modular architecture supports optional integration of RF/PECVD or e-beam evaporation sources for immediate post-cleavage dielectric passivation (e.g., SiO₂, SiNₓ, Al₂O₃) directly onto cleaved facets.

Sample Compatibility & Compliance

The WP-15 accommodates brittle and lattice-mismatched epitaxial wafers including InP-based quantum well lasers, GaAs-based VCSEL epiwafers, and Si-on-insulator (SOI) photonic test chips. Its non-contact scribing stage and controlled fracture propagation minimize subsurface damage—critical for low-threshold lasing and high-finesse resonators. The system conforms to ASTM F1529-22 (Standard Guide for Cleaving Semiconductor Wafers) and supports documentation required for ISO/IEC 17025-accredited calibration labs. All internal surfaces are electropolished 316L stainless steel; O-rings meet USP Class VI and FDA 21 CFR 177.2600 specifications. Vacuum integrity is verified per ISO 20484 (leak rate ≤1×10⁻¹⁰ mbar·L/s He).

Software & Data Management

Operation is managed via SVT’s proprietary UHV Control Suite (v4.2+), a Windows-based application supporting deterministic recipe execution, real-time pressure and temperature logging, and event-driven interlock sequencing (e.g., shutter closure before chamber venting). Audit trails record operator ID, timestamp, cleaving force profile, bar count, and cassette position—fully compliant with FDA 21 CFR Part 11 requirements when paired with network-authenticated login and electronic signature modules. Export formats include CSV, HDF5, and XML for integration into LIMS (e.g., LabWare, Thermo Fisher SampleManager) and MES platforms.

Applications

  • Fabrication of laser bar arrays for high-power diode pumping and fiber-coupled modules.
  • Preparation of reference test structures for facet reflectivity metrology (e.g., using FTIR or cavity ring-down spectroscopy).
  • Rapid prototyping of distributed feedback (DFB) and distributed Bragg reflector (DBR) laser cavities requiring precisely oriented cleave planes.
  • Passivation process development for AlGaInP red-emitting lasers where native oxide formation must be suppressed prior to SiNₓ capping.
  • Research on facet degradation mechanisms (e.g., catastrophic optical damage—COD) under controlled UHV annealing and in-situ optical monitoring.

FAQ

Can the WP-15 cleave wafers thinner than 100 µm?
Yes—the cleaving stage includes micrometer-adjustable backside support and low-force scribing modes suitable for wafers as thin as 50 µm, provided they are bonded to appropriate carriers (e.g., Si handle wafers or glass mounts).
Is remote operation supported for integration into automated cluster tools?
Yes—Ethernet/IP and RS-232 interfaces enable SECS/GEM protocol compliance; full SCADA integration is available via optional OPC UA server module.
Does SVT provide validation documentation for GMP environments?
Yes—IQ/OQ protocols, UVC (Ultra-Vacuum Certification) reports, and material traceability dossiers (including weld logs and surface roughness certificates) are supplied upon request for regulated manufacturing sites.

InstrumentHive
Logo
Compare items
  • Total (0)
Compare