Labthink C230H Tri-Chamber Isobaric Oxygen Transmission Rate (OTR) Analyzer
| Brand | Labthink |
|---|---|
| Origin | Shandong, China |
| Manufacturer Type | OEM Manufacturer |
| Country of Origin | China |
| Model | C230H |
| Price Range | USD 14,000 – 85,000 |
| Test Principle | Isobaric Method with Coulometric Oxygen Sensor |
| Standards Compliance | ASTM D3985, ASTM F1307, ASTM F1927, GB/T 19789, GB/T 31354, DIN 53380, JIS K7126-2-B, YBB 00082003-2015 |
| Test Range | 0.01–200 cm³/(m²·day) |
| Resolution | 0.001 cm³/(m²·day) |
| Repeatability | ±0.01 cm³/(m²·day) or ±2%, whichever is greater |
| Temperature Control | 10–55 °C (±0.2 °C) |
| Humidity Control | 0% RH, 5–90% RH (±1% RH), 100% RH |
| Test Area | 50 cm² |
| Sample Size | 108 mm × 108 mm |
| Max Sample Thickness | ≤3 mm |
| Chamber Count | 3 |
| Package Testing Capacity | Up to 3 L (optional) |
| Carrier Gas | ≥99.999% N₂, ≥0.28 MPa (40.6 psi) |
| Gas Interface | 1/8″ stainless steel tubing |
| Data Management | DataShield™ audit-trail-enabled software |
| Regulatory Support | Optional 21 CFR Part 11 compliance package |
| OS | Windows 10 on 12″ industrial touchscreen |
Overview
The Labthink C230H Tri-Chamber Isobaric Oxygen Transmission Rate (OTR) Analyzer is an engineered solution for precise, reproducible measurement of oxygen permeation through polymeric films, laminates, foils, coated papers, and rigid or flexible packaging structures. It operates on the isobaric (equal-pressure) principle in accordance with ASTM D3985, where oxygen diffuses across a conditioned test specimen under controlled temperature and humidity, while being swept by high-purity nitrogen on the downstream side to a coulometric oxygen sensor. This electrochemical detection method provides direct, real-time quantification of O₂ molecules traversing the material—enabling calculation of OTR in standardized units (cm³/m²·day). Unlike differential-pressure methods, the isobaric configuration eliminates barometric drift and minimizes edge leakage effects, delivering superior baseline stability and long-term measurement fidelity—particularly critical for low-permeability barrier materials used in pharmaceutical blister packs, retort pouches, and sterile medical device packaging.
Key Features
- Tri-chamber integrated test block: Patented monolithic aluminum manifold designed using computational fluid dynamics (CFD) and thermal modeling to minimize inter-chamber variance in temperature, humidity, and gas flow distribution—ensuring <±0.1 °C thermal uniformity and <±0.5% RH consistency across all three test stations.
- Independent environmental monitoring: Each chamber equipped with calibrated platinum RTD and capacitive RH sensors for closed-loop feedback control—supporting simultaneous testing of identical specimens (for statistical validation per ASTM E177) or three distinct materials under identical conditions.
- Automated climate conditioning: Integrated Peltier-based thermo-humidification system maintains setpoint accuracy without external chiller or humidifier—capable of stable operation at 10 °C/0% RH and 55 °C/90% RH within specified tolerances.
- Industrial-grade human-machine interface: 12-inch capacitive touchscreen running Windows 10 IoT Enterprise, enabling seamless integration with lab networks, USB peripherals, and third-party LIMS via standard TCP/IP or OPC UA protocols.
- DataShield™ software architecture: Role-based access control, electronic signatures, full audit trail (per ALCOA+ principles), and encrypted local database storage—certified for GxP environments when configured with optional 21 CFR Part 11 compliance module.
Sample Compatibility & Compliance
The C230H accommodates flat sheet materials up to 108 mm × 108 mm and ≤3 mm thick, with standardized test area of 50 cm². Optional fixtures enable OTR evaluation of formed packages—including trays, cups, and pouches up to 3 L volume—using internal nitrogen purge and external oxygen exposure (ASTM F1307 mode). All hardware and firmware comply with electromagnetic compatibility (EMC) requirements per IEC 61326-1 and safety standards per IEC 61010-1. Test methodology is validated against reference materials traceable to NIST SRM 1477 (polyethylene film) and certified reference standards from PTB (Physikalisch-Technische Bundesanstalt). Routine verification procedures align with ISO/IEC 17025 clause 7.7 and support GLP/GMP laboratory accreditation audits.
Software & Data Management
DataShield™ is a purpose-built, FDA-aligned data acquisition and management platform developed in-house by Labthink. It records all instrument parameters—including real-time sensor voltage, chamber T/RH, flow rates, and calibration history—with immutable timestamps and user attribution. Raw data are stored in SQLite format with SHA-256 hashing; export options include CSV, PDF reports (with digital signature), XML (for LIMS ingestion), and Excel-compatible .xlsx. The system supports scheduled automated backups to network drives or cloud storage (via SMB or SFTP). Audit trail logs capture every action—from method creation and result approval to user login/logout—and retain metadata for minimum 15 years as required by EU Annex 11 and USP . Optional CFR 21 Part 11 configuration includes biometric or smart-card authentication, dual-control approval workflows, and electronic record archiving with WORM (Write Once Read Many) media support.
Applications
- Pharmaceutical packaging: Quantifying OTR of Alu-Alu blisters, cold-form foil, and child-resistant HDPE bottles to ensure shelf-life stability of oxygen-sensitive APIs (e.g., biologics, vitamins, nitroglycerin).
- Food packaging R&D: Evaluating ethylene barrier performance of multilayer PE/EVOH/PA structures for fresh-cut produce and modified atmosphere packaging (MAP) systems.
- Medical device sterility assurance: Validating integrity of Tyvek® pouches, PETG trays, and peelable lidding films used in ISO 11607-compliant sterile barrier systems.
- Photovoltaic encapsulation: Measuring moisture and oxygen ingress rates through EVA, POE, and silicone-based encapsulants in solar module qualification testing (IEC 61215).
- Regulatory submission support: Generating GLP-compliant datasets for FDA 510(k), CE Mark technical files, and PMDA dossier submissions requiring ASTM/ISO-standardized barrier metrics.
FAQ
What gas purity is required for carrier nitrogen?
High-purity nitrogen ≥99.999% (5.0 grade) is mandatory; impurities >1 ppm O₂ or H₂O will compromise baseline stability and sensor longevity.
Can the C230H test metallized PET or SiOₓ-coated films?
Yes—its coulometric sensor detects sub-ppm O₂ concentrations, making it suitable for ultra-high-barrier substrates with OTR values as low as 0.01 cm³/m²·day.
Is calibration traceable to national standards?
All factory calibrations use NIST-traceable reference films; users receive calibration certificates with uncertainty budgets compliant with ISO/IEC 17025.
How does the system handle humidity conditioning at 100% RH?
It employs saturated salt solutions (e.g., KNO₃) in dedicated humidity generation chambers, verified by independent hygrometers meeting ISO 4677 Class II accuracy.
Does DataShield™ support electronic signatures per 21 CFR Part 11?
Yes—when the optional CFR Part 11 module is activated, it enforces identity verification, signature justification, and tamper-evident record sealing per FDA guidance.