Empowering Scientific Discovery

BOXUN GZX-9076MBE Forced-Air Hot Air Oven

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand BOXUN
Origin Shanghai, China
Model GZX-9076MBE
Temperature Range Ambient +5°C to 300°C
Temperature Uniformity ±2.5%
Temperature Fluctuation ±1°C (at 105°C)
Temperature Resolution 0.1°C
Interior Dimensions (L×W×H) 410×350×450 mm
Exterior Dimensions (L×W×H) 555×566×722 mm
Heating Method Forced-air convection with preheated air chamber
Interior Material Mirror-finish stainless steel
Power Input 1550 W
Timer Range 0–9999 min/h (switchable)
Standard Shelving 2 adjustable shelves (max. 5)
Shelf Load Capacity 15 kg per shelf
Operating Ambient Temperature 5–40°C
Compliance GB/T 30435–2013

Overview

The BOXUN GZX-9076MBE is a microprocessor-controlled forced-air hot air oven engineered for precision thermal processing in laboratory, quality control, and industrial R&D environments. It operates on the principle of uniform convective heat transfer—air is preheated in an isolated chamber, then actively circulated via a high-efficiency centrifugal blower and optimized ducting system to ensure rapid thermal equilibration and minimal spatial temperature gradients across the working chamber. Designed for applications including moisture content determination, drying of granular or powdered samples, thermal stabilization of materials, wax melting, sterilization of glassware and instruments, and pre-conditioning of test specimens prior to mechanical or chemical analysis, the GZX-9076MBE meets fundamental requirements for reproducible, traceable thermal treatment under ambient atmospheric conditions.

Key Features

  • Microprocessor-based PMMA II control system with intuitive touch-key interface and high-contrast color digital display for real-time monitoring of setpoint, actual temperature, elapsed time, and alarm status.
  • Preheated air chamber design ensures consistent airflow temperature at inlet, reducing thermal lag and improving ramp rate (>5°C/min at 150°C) and steady-state stability.
  • Forced convection architecture employs a shielded, low-noise AC motor coupled with aerodynamically profiled impeller blades to maintain laminar yet effective internal air renewal—critical for homogeneous drying and residue-free thermal processing.
  • Double-layer insulated chamber with high-density mineral wool insulation minimizes external surface temperature rise and improves energy efficiency without compromising thermal integrity.
  • Rotary dual-point door locking mechanism ensures uniform compression of the silicone gasket against the stainless-steel frame, achieving superior sealing performance and minimizing heat loss during extended operation.
  • Interior features radius-corner construction and removable, corrosion-resistant stainless-steel shelves (standard: 2; expandable to 5), facilitating complete cavity access for cleaning, calibration verification, or validation protocols.
  • Power-loss recovery function automatically restores previous setpoints, timer values, and operational mode upon resumption of mains supply—supporting continuity in GLP/GMP-aligned workflows where process interruption must be documented and mitigated.

Sample Compatibility & Compliance

The GZX-9076MBE accommodates a broad range of non-volatile, non-explosive, and non-corrosive solid and semi-solid samples—including filter papers, polymer pellets, ceramic powders, pharmaceutical excipients, soil aggregates, and metal oxide precursors. Its stainless-steel interior and absence of catalytic surfaces make it suitable for ASTM D2231 (moisture in wood), ISO 712 (cereal moisture), and USP (thermal sterilization of dry heat-stable items) applications when operated within validated parameters. The unit conforms to GB/T 30435–2013, the Chinese national standard for electric hot-air ovens, specifying requirements for temperature accuracy, uniformity, safety interlocks, and electromagnetic compatibility. While not intrinsically rated for hazardous area use, its design supports integration into ISO/IEC 17025-accredited laboratories where equipment qualification (IQ/OQ/PQ) is performed per internal SOPs.

Software & Data Management

The GZX-9076MBE operates as a standalone instrument with embedded firmware—no external PC or proprietary software is required for basic operation. All critical parameters (set temperature, dwell time, alarm thresholds) are stored in non-volatile memory and retained during power cycling. The display logs real-time deviations and triggers visual/audible alerts for over-temperature events (>305°C) or sensor failure. For regulated environments, users may integrate the oven into centralized monitoring systems via optional RS485 Modbus RTU output (available on select configurations), enabling timestamped data logging, remote status interrogation, and audit-trail generation compatible with FDA 21 CFR Part 11-compliant platforms when paired with validated third-party SCADA or LIMS interfaces.

Applications

  • Determination of loss on drying (LOD) in pharmaceutical raw materials and finished dosage forms per USP .
  • Thermal conditioning of tensile test specimens (e.g., ASTM D638) to stabilize hygroscopic polymers prior to mechanical evaluation.
  • Residue ash preparation for elemental analysis (e.g., AOAC 990.19) by controlled charring of organic matrices.
  • Pre-drying of filter media and collection substrates used in environmental particulate matter sampling (EPA Method 201A).
  • Accelerated aging studies of packaging materials under elevated temperature stress (ISO 11607-1 Annex B).
  • Heat treatment of calibration standards, reference materials, and laboratory glassware prior to gravimetric or volumetric procedures.

FAQ

Is the GZX-9076MBE suitable for sterilization of surgical instruments?
No—it is designed for dry-heat applications such as depyrogenation of glassware and metal tools at ≥160°C for ≥2 hours, but does not meet ISO 17665 or AAMI ST55 requirements for medical device sterilization validation.
Can the oven be calibrated externally using NIST-traceable probes?
Yes—the unit includes a dedicated PT100 sensor port (optional accessory) for insertion of external reference thermometers during temperature mapping and IQ/OQ verification.
Does the timer support both minute and hour modes?
Yes—the digital timer is configurable for either minutes (0–9999 min) or hours (0–9999 h), selectable via front-panel menu.
What is the maximum load weight the chamber can accommodate?
With five fully installed shelves (15 kg each), the total static load capacity is 75 kg—however, optimal airflow uniformity is maintained at ≤60% volume occupancy.
Is the stainless-steel interior electropolished or passivated?
The mirror-finish interior uses AISI 304 stainless steel with standard mill finish; electropolishing is available as a factory customization option upon request.

InstrumentHive
Logo
Compare items
  • Total (0)
Compare
0