- All
- Favorite
- Popular
- Most rated
| Brand | Huayun Instruments |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Product Type | Online Fixed-Mount Analyzer |
| Principle | Non-Dispersive Infrared (NDIR) Absorption |
| Response Time | T₀–T₉₀ ≤ 15 s (from sample gas entry) |
| Repeatability | ≤ ±1% FS for CO, CO₂, CH₄ |
| Linearity Error | ≤ ±2% FS |
| Zero Drift | ≤ ±2% FS / 72 h |
| Span Drift | ≤ ±2% FS / 72 h |
| Measurement Ranges | CO (0–500) ppm, CO₂ (0–500) ppm, CH₄ (0–1000) ppm |
| Output Signal | 4–20 mA analog |
| Power Supply | 220 VAC ±10%, 50 Hz ±10% |
| Dimensions | 436 × 426 × 133 mm (L×W×H, excluding protrusions) |
| Weight | ≤10 kg |
| Operating Temperature | 5–45 °C |
| Relative Humidity | ≤80% RH (non-condensing) |
| Environmental Requirements | Non-corrosive atmosphere, airflow ≤0.5 m/s, vibration-free horizontal mounting |
| Brand | Huayun Instruments |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Product Type | Online Analyzer |
| Measurement Principle | Non-Dispersive Infrared (NDIR) Absorption |
| Response Time | T₀–T₉₀ ≤ 15 s (from sample gas entry) |
| Repeatability | ≤ ±1% FS for CO, CO₂, CH₄, NO, SO₂, NH₃ |
| Linearity Error | ≤ ±2% FS |
| Zero Drift | ≤ ±2% FS / 72 h |
| Span Drift | ≤ ±2% FS / 72 h |
| Measurement Ranges | CO (0–500 ppm), CO₂ (0–500 ppm), CH₄ (0–1000 ppm), configurable up to 0–100.0% vol |
| Output Signal | 4–20 mA analog |
| Power Supply | 220 VAC ±10%, 50 Hz ±10% |
| Dimensions | 436 × 426 × 133 mm (L×W×H, excluding protrusions) |
| Weight | ≤10 kg |
| Operating Temperature | 5–45 °C |
| Relative Humidity | ≤80% RH (non-condensing) |
| Environmental Requirements | Non-corrosive atmosphere, airflow ≤0.5 m/s, horizontal mounting, vibration-free installation |
Show next