- All
- Favorite
- Popular
- Most rated
| Brand | Bruker |
|---|---|
| Origin | Germany |
| Model | G8 GALILEO |
| Detection Principle | Inert Gas Fusion (IGF) coupled with Thermal Conductivity Detection (TCD) and Mass Spectrometry (MS) |
| Mass Range | 0–100 amu/mz |
| Detection Limits | <1 ppb (ng/g) for H, O, N (1 g sample) |
| Measurement Method | Direct gas-phase analysis via Smart Molecule Sequence™ |
| Furnace Options | Water-cooled graphite electrode furnace with FusionControl™ real-time non-contact temperature monitoring |
| Optional External IR Furnace | For diffusion hydrogen analysis per ISO 360 and AWS A4.3 |
| Compliance | Designed for GLP/GMP environments |
| Automation | Fully automated sample/crucible handling, electrode cleaning, and 10-step gas dosing calibration |
| MS Configuration | Industrial-grade quadrupole mass spectrometer optimized for H₂, N₂, O₂, Ar, and H₂O detection with reference channel stabilization |
| Particle Filtration | Integrated high-capacity dust collector + inline particulate filter |
| Brand | Bruker |
|---|---|
| Origin | Germany |
| Model | G4 PHOENIX DH |
| Detection Principle | Thermal Desorption Mass Spectrometry (TDMS) & Thermal Conductivity Detection (TCD) |
| Detection Limit | <1 ng/g (with optional quadrupole MS) / <50 ng/g (standard TCD) |
| Furnace Max Temp | 900 °C (IR-heated, low thermal mass) |
| Sample Chamber Diameter | 30 mm |
| Calibration | 10-volume automated gas standard system |
| Compliance | Fully compatible with ISO 3690, AWS A4.3, ASTM E1442, EN ISO 17025 |
| Brand | Bruker |
|---|---|
| Origin | Germany |
| Model | G6 LEONARDO |
| Technique | Inert Gas Fusion (IGF) with Thermal Conductivity Detection (TCD) for H/N |
| Detection Limits | < 0.1 ppm (ng/g) for H and N (1 g sample) |
| Maximum Furnace Temperature | > 3000 °C (water-cooled electrode furnace) |
| Carrier Gas | Argon (He-free operation) |
| Compressed Air Consumption | 0 L/min |
| Compliance | ASTM E1019, ISO 14284, ISO 15350, USP <233>, GLP/GMP-ready software architecture |
| Brand | Bruker |
|---|---|
| Origin | Germany |
| Model | G4 ICARUS Series 2 |
| Detection Principle | High-frequency induction combustion coupled with UV absorption (SO₂) and infrared absorption (CO₂) |
| Analysis Time | < 60 s per cycle including auto-cleaning |
| Filter Pore Size | 3 µm |
| Signal-to-Noise Ratio (SO₂) | 10× improvement vs. conventional IR |
| Dynamic Range (SO₂) | 50-year linear calibration stability |
| Compliance | ASTM E1019, ISO 15350, DIN 51018, USP <231>, GLP/GMP-ready data integrity architecture |
Show next