Mercury Analyzers
Filter
Showing 31–60 of 61 results
| Brand | LUMEX |
|---|---|
| Origin | Canada |
| Model | Light 915 |
| Instrument Type | Portable |
| Measurement Principle | Zeeman Atomic Absorption Spectrometry with High-Frequency Modulation (ZAAS-HFM) |
| Detection Limit | 100 ng/m³ |
| Repeatability | ±1% RSD |
| Linearity Error | ±1% |
| Measurement Range | 0.1–3000 µg/m³ |
| Air Pump Flow Rate | 10 L/min |
| Analysis Time | Real-time |
| Dimensions | 290 × 210 × 110 mm |
| Weight | 3.3 kg |
| Power Supply | Internal 12 V DC rechargeable battery (12 h operation) / External 110–240 V AC, 50/60 Hz |
| Power Consumption | 20 W |
| Data Storage | Internal memory with USB and Bluetooth connectivity |
| Compliance | EPA Emergency Response Equipment List, ASTM D6784-22 (for mercury in air), ISO 6978-1:2003 (natural gas—mercury determination) |
| Brand | LUMEX |
|---|---|
| Origin | Canada |
| Model | LS-915 |
| Instrument Type | Benchtop Laboratory System |
| Measurement Principle | Cold Vapor Atomic Absorption Spectrometry (CVAAS) |
| Detection Limit | 1 pg Hg |
| Repeatability | ≤1% RSD |
| Linearity Error | ≤2% |
| Sample Capacity | 75 positions |
| Solid Sample Mass | ≥1000 mg |
| Liquid Sample Volume | ≥500 µL |
| Robotic Architecture | Cartesian XYZ Coordinate System |
| Crucible Material | High-Purity Quartz (operable up to >1000 °C) |
| Brand | LUMEX |
|---|---|
| Origin | Canada |
| Model | Pyro-915+ |
| Instrument Type | Portable |
| Principle | Cold Vapor Atomic Absorption Spectrometry (CVAAS) with Zeeman Background Correction |
| Detection Limit | 0.05 ng/mL (liquid) / 0.05 µg/kg (solid) |
| Repeatability | ±2% RSD |
| Linearity Error | ±1% |
| Sample Throughput | <2 min per sample |
| Operating Temperature Range | Programmable up to 950 °C |
| Air Monitoring LOD | 0.5 ng/m³ (10-s response) / 2 ng/m³ (1-s response) |
| Power Supply | 100–240 VAC, 50/60 Hz |
| Power Consumption | 700 W |
| Weight | 17.5 kg |
| Dimensions (Controller) | 400 × 280 × 135 mm |
| Dimensions (Furnace) | 430 × 340 × 135 mm |
| Continuous Operation | Up to 8 h |
| Carrier Gas Flow | 1.0–3.0 L/min |
| Sample Mass Range | 10–1500 mg |
| Upper Detection Limit | 200 mg/kg (based on 20 mg sample) |
| Brand | LUMEX |
|---|---|
| Origin | Canada |
| Model | RA-915AM |
| Instrument Type | Online Analyzer |
| Measurement Principle | Zeeman Atomic Absorption Spectrometry with High-Frequency Modulation (ZAAS-HFM) |
| Detection Limit | 0.5 ng/m³ |
| Repeatability | ±1% |
| Linearity Error | ±2% |
| Gas Flow Rate | 3–10 L/min |
| Response Time | 10–300 s |
| Range I | 0.5–2000 ng/m³ |
| Range II | 100–400,000 ng/m³ |
| Operating Temperature | −20 to +50 °C |
| Pressure Range | 84–106.7 kPa |
| Relative Humidity | <99% at 35 °C |
| Power Supply | 240 VAC, 50/60 Hz |
| Power Consumption | 120 W |
| Dimensions | 600 × 480 × 220 mm |
| Weight | 19 kg |
| Communication Interfaces | RS-232, Dual USB, LAN, Isolated 4–20 mA Output |
| Brand | LUMEX |
|---|---|
| Origin | Canada |
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | RA-915AMNG |
| Instrument Type | Online Analyzer |
| Principle | Zeeman Atomic Absorption Spectrometry with High-Frequency Modulation (ZAAS-HFM) |
| Detection Limit | 0.5 ng/m³ |
| Repeatability | ±1% |
| Linearity Error | ±1% |
| Measurement Range | 0–2000 ng/m³ (configurable multi-range options available) |
| Reporting Interval | ≥2 s (user-definable) |
| Benzene Correction | Fully Automatic |
| Zero Drift Correction | User-selectable (recommended every 60 min) |
| Span Drift Correction | User-selectable (recommended every 3 h) |
| Calibration | Fully Automated (calibration source ≈12,000 ng/m³ Hg) |
| Sample Flow Rate | 7–10 L/min |
| Communication Protocol | Standard Bavarian Protocol (customizable upon request) |
| Status Outputs | SERVICE (zero/calibration mode), FAILURE, MEASURE (Bavarian) |
| Operating Temperature | 14–40 °C |
| Atmospheric Pressure | 84–106.7 kPa |
| Relative Humidity | <99% at 35 °C |
| Power Supply | 220–240 VAC, 50 Hz |
| Dimensions | 600 × 480 × 220 mm |
| Weight | 19 kg |
| Brand | LUMEX |
|---|---|
| Origin | Canada |
| Model | RA-915F |
| Instrument Type | Portable |
| Measurement Principle | Cold Vapor Atomic Absorption Spectrometry (CVAAS) with Thermal Decomposition |
| Detection Limit | 0.005 ng Hg |
| Repeatability | ±1% RSD |
| Linearity Error | ±1% |
| Detection Range | 5 orders of magnitude |
| Sample Throughput | 1–5 min per analysis |
| Power Supply | 220 VAC, 50 Hz (with optional DC/vehicle power adapter) |
| Dimensions (Analyzer Unit) | 300 × 260 × 130 mm |
| Weight (Analyzer Unit) | 4.0 kg |
| Dimensions (Thermal Decomposer Module) | 350 × 260 × 120 mm |
| Weight (Thermal Decomposer Module) | 7.5 kg |
| Dimensions (Power Supply Unit) | 380 × 260 × 130 mm |
| Weight (Power Supply Unit) | 10.0 kg |
| Sample Mass Range | 10–500 mg |
| Liquid Detection Limit | 0.5 µg/L |
| Solid Upper Quantitation Limit | 200 mg/kg |
| Brand | LUMEX |
|---|---|
| Origin | Canada |
| Model | RA-915Lab |
| Instrument Type | Benchtop Laboratory Analyzer |
| Measurement Principle | Cold Vapor Atomic Absorption Spectrometry (CVAAS) |
| Repeatability | <1.5% RSD |
| Linearity Error | ±5% |
| Sample Throughput | Up to 45 samples unattended |
| Analysis Time | ≤3 minutes per sample |
| Linear Range | 0–2,000,000 µg/kg (ppb–ppm equivalent across matrices) |
| Detection Capability | Sub-ng absolute sensitivity (matrix-dependent) |
| Compliance | EPA Method 7473, EPA 1631, EPA 30B, EPA SW-846 |
| Brand | LUMEX |
|---|---|
| Origin | Canada |
| Model | RA-915M Series |
| Instrument Type | Portable |
| Principle | Zeeman Atomic Absorption Spectrometry with High-Frequency Modulation (ZAAS-HFM) |
| Detection Limit | 0.5 ng/m³ (10 s), 2 ng/m³ (1 s) |
| Repeatability | <2% RSD |
| Linearity Error | ±5% |
| Measurement Range | 0.5–200,000 ng/m³ |
| Zero Signal SD | <2 ng/m³ |
| 5-min Zero Drift | <2 ng/m³ |
| Warm-up Time | <20 min |
| Air Pump Flow Rate | >10 L/min |
| Power Supply | Internal 12 V DC battery (8 h continuous operation) or 110/220 V AC, 50/60 Hz |
| Dimensions | 460 × 110 × 220 mm |
| Weight | 7 kg |
| Power Consumption | 40 W |
| Compliance | EPA Method 1631, EPA SW-846 7473 & 30B, ASTM D7622-10, EN 1483 / 13806 / 15852, GB 7468-87 |
| Brand | LUMEX |
|---|---|
| Origin | Canada |
| Model | RA-915S |
| Instrument Type | Portable |
| Principle | Cold Vapor Atomic Absorption Spectrometry (CVAAS) with Zeeman-effect background correction |
| Detection Limit | 0.01 µg/kg |
| Repeatability | ≤1% RSD |
| Linearity Error | ≤1% |
| Sample Introduction | Direct thermal decomposition (10–1500 mg solid/liquid) |
| Analysis Time | 1–2 min per sample |
| Dynamic Range | 5 orders of magnitude |
| Wavelength | 254 nm (Hg resonance line) |
| Power Supply | 198–242 VAC, 50 ± 1 Hz |
| Dimensions (Detector Unit) | 470 × 210 × 110 mm |
| Integrated Air Pump | Yes |
| Onboard Calibration | Certified Hg standard traceable to NIST |
| Brand | LUMEX |
|---|---|
| Origin | Canada |
| Model | RA-915W |
| Instrument Type | Portable |
| Measurement Principle | Cold Vapor Atomic Absorption Spectrometry (CVAAS) with High-Frequency Zeeman Background Correction (ZAAS-HFM) |
| Detection Limit | 0.5 ng/L |
| Repeatability | ±1% RSD |
| Linearity Error | ±1% |
| Compliance | EPA Method 1631, EPA SW-846, EN 1483 & EN 13806, HJ 543–2009, HJ 597–2011, GB 3838–2002, GB 5009.17–2014, WS/T 25–1996, GB/T 15555.1–1995, GB/T 17136–1997 |
| Brand | LUMEX |
|---|---|
| Origin | Canada |
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | RP-91NG |
| Instrument Type | Portable |
| Measurement Principle | Cold Vapor Atomic Absorption Spectrometry (CVAAS) with High-Frequency Zeeman Modulation (ZAAS-HFM) |
| Detection Limit | 1 ng/m³ |
| Repeatability | ±1% RSD |
| Linearity Error | ±1% |
| Concentration Range | 0.5–20,000 ng/m³ (optional high- and ultra-high-concentration modes available) |
| Air Pump Flow Rate | 2–5 L/min |
| Operating Ambient Temperature | 1–40 °C |
| Atmospheric Pressure Range | 84.0–106.7 kPa |
| Relative Humidity | >95% at 35 °C |
| Power Supply | 100–240 VAC, 50/60 Hz |
| Power Consumption | 35 W |
| Internal Battery Runtime | 8 h |
| Dimensions | 320 × 290 × 160 mm |
| Weight | 6.5 kg |
| Brand | LUMEX |
|---|---|
| Origin | Canada |
| Model | RP-92 Liquid Mercury |
| Instrument Type | Portable |
| Measurement Principle | Cold Vapor Atomic Absorption Spectrometry (CVAAS) |
| Sample Volume | 0.5–20 mL |
| Air Pump Flow Rate | 0.8–4.2 L/min |
| Power Supply | 100–240 VAC, 50/60 Hz |
| Power Consumption | 40 W |
| Pump Unit Net Weight | 3.0 kg |
| Pump Unit Dimensions | 360 × 225 × 125 mm |
| Brand | LUMEX |
|---|---|
| Origin | Canada |
| Model | UMA |
| Instrument Type | Portable Mercury Analyzer |
| Principle | Cold Vapor Atomic Absorption Spectrometry (CVAAS) with Zeeman Background Correction |
| Detection Limit | 1 ppb |
| Repeatability | <1% RSD |
| Linearity Error | ±1% |
| Sample Throughput | Solid/Liquid analysis completed in ≤60 s |
| Gas Response Time | 1 s |
| Dynamic Range | 6 orders of magnitude |
| Power Supply | 100–240 VAC, 50/60 Hz or internal battery (8 h continuous operation) |
| Dimensions (Control Unit) | 400 × 280 × 135 mm |
| Dimensions (Thermal Decomposer Module) | 430 × 340 × 135 mm |
| Weight | 17.5 kg |
| Air Pump Flow Rate | 0.8–4.2 L/min |
| Liquid Sample Volume (CV Mode) | 0.5–20 mL |
| Solid Sample Mass (Thermal Decomposition) | 5–300 mg |
| Brand | LabTech |
|---|---|
| Model | MAX-S |
| Principle | Catalytic Thermal Decomposition–Gold Amalgamation–Cold Vapor Atomic Absorption Spectrometry (CV-AAS) |
| Detection Limit | 0.0005 ng Hg |
| Linear Range | 0–30,000 ng Hg (7 orders of magnitude) |
| Analysis Time | 2–5 min per sample |
| Sample Forms | Solid, liquid, and gas matrices without pretreatment |
| Repeatability | RSD < 3% (n = 10 at 10 ng level) |
| Calibration Stability | 3–6 months between standard curve verification |
| Compliance | Designed for GLP/GMP environments |
| Brand | Mercury Instruments |
|---|---|
| Origin | Germany |
| Model | AULA-254 Gold |
| Instrument Type | Benchtop Laboratory Analyzer |
| Measurement Principle | Cold Vapor Atomic Absorption Spectrometry (CVAAS) |
| Detection Limit | < 1 ng/L (with gold amalgamation trap) |
| Repeatability | ≤ 1.0% RSD |
| Linearity Error | ±1% |
| Optical Wavelength | 253.7 nm |
| UV Source | Electrodeless Low-Pressure Mercury Discharge Lamp |
| Sample Introduction | 53-position rotary autosampler |
| Sample Vial | 10 mL glass vials with optional aluminum foil caps |
| Carrier Gas | Argon (4–6 L/h), mass flow controlled |
| Digestion Heater Temperature | 98 °C |
| Optical Cell | Fused silica, 230 mm path length |
| Optical Cell Temperature | 50 °C |
| Measurement Cycle | 60–280 s |
| Software | AULA-WIN (Windows®-based) |
| Power Supply | 115/230 V~, 50–60 Hz, 100 W (120 W with ASD module) |
| Dimensions (W×H×D) | 37 × 38 × 44 cm |
| Footprint Requirement | 50 × 70 cm (W × D) |
| Weight | ~14 kg (without PC) |
| Compliance | EPA Methods 245.1, 245.7, 1631, 7470, 7471A |
| Optional Module | Automated Sample Digestion (ASD) |
| Brand | Mercury Instruments |
|---|---|
| Origin | Germany |
| Model | LabAnalyzer 254 |
| Instrument Type | Benchtop Laboratory Analyzer |
| Measurement Principle | Cold Vapor Atomic Absorption Spectrometry (CVAAS) |
| Wavelength | 254 nm |
| Detection Limit | ≤0.1 µg/L |
| Repeatability | ≤1.0% RSD |
| Linearity Error | ±1.0% |
| Sample Volume | 2–10 mL |
| UV Source | Electrodeless Discharge Lamp (EDL), temperature-stabilized |
| Detector | Temperature-controlled UV-enhanced silicon photodiode |
| Optical Cell | Fused quartz, 23 cm path length, heated to 70 °C |
| Flow Rate | 30 L/h (adjustable) |
| Pump | Long-life diaphragm pump |
| Sensitivity | 5 ng/L (0.05 ng absolute) |
| Linear Range | 0.01–10 µg/L (10 ppt–10 ppb) for 10 mL sample |
| Output Interfaces | 4–20 mA analog, USB/RS-232, parallel printer port |
| Power Supply | 230 VAC, 50–60 Hz (optional 115 VAC) |
| Power Consumption | 35 W |
| Dimensions (main unit) | 45 × 15 × 35 cm (W × H × D) |
| Optical Module | 24 × 48 × 27 cm (W × H × D) |
| Footprint Requirement | ~70 × 50 cm (W × D) |
| Weight | ~10 kg |
| Compliance | DIN 38406-12, EPA Method 7470A & 7471A |
| Brand | Mercury Instruments |
|---|---|
| Origin | Germany |
| Model | MC3000 |
| Instrument Type | Portable |
| Principle | Cold Vapor Atomic Absorption (CVAA) |
| Detection Limit | ≤1 ng/mL |
| Repeatability | ≤3% (CVAA method) |
| Linearity Error | 5% |
| Mercury Vapor Output Range | 15–500 µg/m³ Hg⁰ |
| Calibration Method | NIST-Recommended Mercury Saturated Vapor Pressure Technique |
| Stabilization Time | 30 min |
| Brand | Mercury Instruments |
|---|---|
| Origin | Germany |
| Model | MMS-NG |
| Instrument Type | Online Analyzer |
| Measurement Principle | Cold Vapor Atomic Absorption Spectroscopy (CVAAS) |
| Detection Limit | ≤0.1 ng/m³ |
| Repeatability | ≤1.0% (CVAAS) |
| Linearity Error | ±1% (CVAAS) |
| Measurement Range | 0.001–50 µg/m³ (extendable to 2000 µg/m³ with integrated dilution system) |
| Wavelength | 253.7 nm |
| Sample Inlet Pressure | up to 240 bar (3480 psi) |
| Sampling Points | 2–16 via multiplexed valve manifold |
| Cycle Time | ~3 minutes per point |
| Calibration Method | Saturated mercury vapor generation (NIST-recommended vapor pressure equation) |
| Certifications | ATEX 2G IIC T4 EEx p, IECEx, ISO/DIS 6978-3, ASTM D5954, VDI 2267 Part 8, ISO 6978-1 |
| Brand | Mercury Instruments |
|---|---|
| Origin | Germany |
| Model | 3000XS |
| Instrument Type | Portable |
| Measurement Principle | Cold Vapor Atomic Absorption Spectrometry (CVAAS) |
| Detection Limit | 0.1 µg/m³ |
| Repeatability | ≤3.0% |
| Linearity Error | ±10% |
| Measurement Ranges | 0–100 / 0–1000 / 0–2000 µg/m³ |
| Resolution | 0.1 µg/m³ |
| UV Source | Electrodeless Low-Pressure Mercury Lamp (253.7 nm) |
| Optical Cell | High-Purity Fused Silica |
| Pump Type | Maintenance-Free Diaphragm Pump |
| Battery Life | Up to 6 Hours |
| Data Logging | Integrated Recorder |
| Compliance | Designed for ISO 15202, ASTM D6784, and EPA Method 29–compliant field deployment |
| Brand | Mercury Instruments |
|---|---|
| Origin | Germany |
| Model | 3000XS |
| Instrument Type | Portable |
| Measurement Principle | Cold Vapor Atomic Absorption Spectrometry (CVAAS) |
| Detection Limit | ≤0.1 µg/m³ |
| Repeatability | ≤1.0% |
| Linearity Error | ±1% |
| Wavelength | 253.7 nm |
| Light Source | Electrodeless Discharge Lamp (EDL) |
| Response Time (99% signal) | 1 s |
| Sensitivity | 0.1 µg/m³ (0.01 ppb) |
| Measurement Ranges | 0–100, 0–1000, 0–2000 µg/m³ |
| Data Logger Capacity | 10,000 readings |
| Power Supply | Rechargeable NiMH battery (6 h operation), 12 V DC input, 100–240 V AC (50/60 Hz) |
| Weight | ~5 kg (with battery) |
| Display | Backlit graphical LCD |
| Output Interfaces | USB, RS232, 4–20 mA analog output |
| Housing | Corrosion-resistant ABS with IP65-rated connectors |
| Pump | Maintenance-free diaphragm pump |
| Filter | Integrated particulate filter at probe inlet |
| Brand | Mercury Instruments |
|---|---|
| Origin | Germany |
| Model | VM-3000 |
| Instrument Type | Portable |
| Measurement Principle | Cold Vapor Atomic Absorption (CVAAS) |
| Detection Limit | ≤ 0.1 µg/m³ |
| Repeatability | ≤ 1.0% (CVAAS) |
| Linearity Error | ±1% (CVAAS) |
| Wavelength | 253.7 nm |
| UV Source | Electrodeless Low-Pressure Mercury Lamp (EDL) |
| Measurement Cell | Heated Fused Silica |
| Response Time (99%) | < 1 s |
| Measuring Ranges | Range I: 0.1–100 µg/m³ (0.01–10 ppb) |
| Range II | 0–1000 µg/m³ |
| Range III | 0–2000 µg/m³ (0–200 ppb) |
| Sensitivity | 0.1 µg/m³ (0.01 ppb) |
| Alarm | 3-Level Programmable + Status Alarms (Cell Soiling, UV Lamp Exhaustion, Battery Level) |
| Display | Graphical LCD with Backlight |
| Outputs | 4–20 mA, RS-232, Optional Data Logger |
| Power | External 12 V DC or Integrated 12 V Rechargeable Battery (≈5 h runtime) |
| Origin | Germany |
|---|---|
| Type | Portable Mercury Analyzer |
| Principle | Cold Vapor Atomic Absorption Spectrometry (CVAAS) with Gold Amalgamation Preconcentration |
| Detection Limit | ≤0.1 ng/m³ (or 0.5 pg Hg absolute) |
| Repeatability | ≤1.0% RSD |
| Linearity Error | ±1% |
| Sample Volume Range | 0.1–100 L |
| Sampling Cycle | 10 s–15 min |
| Measurement Cycle | 10 s–16 min |
| Measurement Range (10 L sample) | 0.1–2000 ng/m³ |
| Measurement Range (1 L sample) | 1–10,000 ng/m³ |
| Flow Control | Electronic mass flow meter |
| Pump Type | Diaphragm pump with Viton membranes |
| Filter | Disposable 0.45 µm PTFE membrane |
| Light Source | Electrodeless low-pressure mercury lamp (253.7 nm) |
| Data Storage | 10,000 records (~100 days) |
| Interface | USB & RS232 |
| Power Supply | 110/230 V, 50/60 Hz, max. 125 W |
| Dimensions (W×H×D) | 45 × 15 × 35 cm |
| Weight | ~9 kg |
| Operating Temperature | 5–35 °C |
| Brand | MILESTONE |
|---|---|
| Origin | Italy |
| Model | DMA-1 |
| Instrument Type | Portable |
| Measurement Principle | Catalytic Thermal Decomposition – Gold Amalgamation – Cold Vapor Atomic Absorption Spectrometry (CVAAS) |
| Detection Limit | ≤0.1 ng |
| Linear Range | 0–1500 ng |
| Repeatability (RSD) | ≤1.0% |
| Linearity Error | ±10% |
| Sample Throughput | 2–5 min per analysis |
| Regulatory Compliance | EPA Method 7473, HJ 923–2017, GB 5009.17–2021 |
| Brand | Milestone |
|---|---|
| Origin | Italy |
| Model | DMA-80 evo |
| Principle | Catalytic Thermal Decomposition – Gold Amalgamation – Cold Vapor Atomic Absorption Spectrometry (CVAAS) |
| Detection Limit | 0.0002 ng |
| Linear Range | 0–30,000 ng |
| Repeatability (RSD) | ≤1.0% |
| Recovery Rate | >98.5% at 1000 ng |
| Compliance | EPA Method 7473, HJ 923–2017, GB 5009.17–2021 |
| Brand | Nippon Instruments Corporation (NIC) |
|---|---|
| Origin | Japan |
| Model | WA-5A |
| Instrument Type | Benchtop Laboratory Analyzer |
| Measurement Principle | Cold Vapor Atomic Absorption Spectrometry (CVAAS) with Dual Gold Amalgamation |
| Detection Limit | 0.001 ng (1 pg) |
| Linear Range | 0.001 – 1000 ng |
| Compliance | ASTM D5954-98(2014)e1, ISO 6978-2:2005, ISO 20552, JIS K 0312, HJ 910–2017, GB 16781.2–2010, HJ 597–2011, HJ 543–2009 |
| Sample Forms | Gaseous matrices (ambient air, natural gas, shale gas, LPG, LNG, hydrogen), aqueous samples (with S-MA/AQUA accessory) |
| Brand | Polytech |
|---|---|
| Origin | Anhui, China |
| Manufacturer | Yes |
| Instrument Type | Benchtop Laboratory Analyzer |
| Detection Principle | Cold Vapor Atomic Fluorescence Spectrometry (CVAFS) |
| Detection Limit | ≤0.002 ng/L (methylmercury), ≤0.002 ng/L (ethylmercury), ≤0.03 ng/L (total mercury) |
| Repeatability | ≤2% RSD |
| Linear Range | 0–40 ng/L (methylmercury), 0–40 ng/L (ethylmercury), 0–400 ng/L (total mercury) |
| Sample Capacity | 96 positions |
| Dimensions | 1200 × 400 × 400 mm |
| Weight | 43 kg |
| Power Supply | 220 VAC, 50 Hz, 5 A |
| Analysis Throughput | 9 samples/hour (methylmercury), 5 samples/hour (alkylmercury) |
| Sample Volume | 25 mL or 40 mL (aqueous) |
| Optional Total Mercury Module | Yes |
| Brand | -- |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Authorized Distributor |
| Regional Classification | Domestic (PRC) |
| Model | RP-92 |
| Pricing | Upon Request |
| Instrument Type | Online/Field-Deployable |
| Analytical Principle | Cold Vapor Atomic Fluorescence Spectrometry (CV-AFS) |
| Brand | Teledyne Leeman Labs |
|---|---|
| Origin | Imported |
| Model | Hydra II AA |
| Instrument Type | Benchtop Laboratory Analyzer |
| Measurement Principle | Cold Vapor Atomic Absorption Spectrometry (CVAAS) |
| Compliance | EPA Methods 245.1, 245.5, 245.6, 7470A, 7471B |
| Brand | Teledyne Leeman Labs |
|---|---|
| Origin | Imported |
| Model | Hydra II C |
| Instrument Type | Benchtop Laboratory Analyzer |
| Measurement Principle | Cold Vapor Atomic Absorption Spectrometry (CVAAS) |
| Compliance | EPA Method 7473, SN/T 3511–2013, HJ 910–2017, HJ 917–2017, HJ 923–2017, Cosmetics Safety Technical Specifications (2015 Edition) |
| Sample Introduction | Direct Solid/Liquid Injection Without Prior Digestion |
| Optical Design | Dual-Beam, Dual-Cell Configuration with Real-Time Baseline Monitoring |
| Moisture Management | Integrated Desiccant Drying Tube |
| Operational Readiness | Ambient-Temperature Operation, No Preheating Required |
| Brand | Titan Instruments |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Original Equipment Manufacturer (OEM) |
| Instrument Type | Benchtop Laboratory Analyzer |
| Model | DMA-500 |
| Pricing | Upon Request |
| Measurement Principle | Cold Vapor Atomic Absorption Spectrometry (CVAAS) |
| Sample Introduction | Direct Solid/Liquid/Gas Injection (No Acid Digestion Required) |
| Automation | 100-Position 4-Axis Robotic Autosampler |
| Thermal Treatment | Dual-Zone PID-Controlled Ashing & Catalytic Combustion Furnace |
| Trapping System | Constant-Power Heated Mercury Trap with Rapid Thermal Ramp |
| Detection Architecture | Tandem Dual-Absorption Cells + Dual-Beam Light Source Correction |