AUCMA DW-86L102 Ultra-Low Temperature Freezer
| Brand | AUCMA |
|---|---|
| Origin | Shanghai, China |
| Model | DW-86L102 |
| Capacity | 102 L |
| Internal Dimensions (mm) | 450 × 540 × 430 |
| External Dimensions (mm) | 1125 × 888 × 770 |
| Power Supply | 220 V / 50 Hz |
| Rated Power | 558 W |
| Temperature Range | −40 °C to −86 °C |
| Configuration | Vertical, Side-Opening |
| Door Type | Dual-layer insulated door with multi-point sealing |
| Insulation | High-density CFC-free polyurethane foam (≥120 mm thickness) |
| Compressor System | Cascade refrigeration with imported compressors and EBM EC fans |
| Safety Features | Dual-stage temperature alarm, voltage fluctuation compensation, condenser clog detection, sensor fault monitoring, battery-backed data retention, remote alarm interface |
| Compliance | Designed per IEC 61000-6-3 (EMC), GB/T 20154–2014 (Chinese ULT freezer standard), compatible with GLP/GMP environments requiring audit-trail-capable temperature monitoring |
Overview
The AUCMA DW-86L102 is a vertically configured ultra-low temperature (ULT) freezer engineered for long-term cryogenic storage of biologics, clinical specimens, and temperature-sensitive materials at stable operating temperatures ranging from −40 °C to −86 °C. Utilizing a cascade refrigeration architecture—comprising two independent refrigerant circuits (typically R404A/R23 or R508B-based)—the system achieves rapid cooldown, high thermal stability (< ±2 °C uniformity across working volume), and robust recovery after door openings. Its structural design prioritizes energy efficiency and operational resilience: vacuum-insulated panels are not employed, but the unit integrates ≥120 mm thick, zero-ozone-depletion-potential (ODP) polyurethane foam insulation, dual-pane tempered glass door with magnetic gasket seal, and pressure-equalization valve to minimize thermal shock and mechanical strain during access. The DW-86L102 meets national technical specifications outlined in GB/T 20154–2014 and aligns with foundational requirements for laboratory cold chain integrity under ISO/IEC 17025-accredited facilities.
Key Features
- Microprocessor-based temperature control with dual Pt100 sensors (one for control, one for redundancy and alarm triggering), enabling precise setpoint adjustment in 0.1 °C increments and real-time display via LED interface.
- Comprehensive safety architecture: audible/visual alarms for high/low temperature deviation, power failure, condenser fouling, ambient overheating (>35 °C), overvoltage (>242 V), and sensor failure; configurable alarm thresholds and optional dry-contact output for building management systems (BMS) or external alarm panels.
- Cascade refrigeration system with dual imported hermetic compressors and brushless EC fans (EBM Papst), delivering low-noise operation (<52 dB[A]) and optimized heat exchange via finned-tube evaporator and air-cooled condenser with auto-defrost cycle scheduling.
- Human-centered ergonomics: integrated leveling feet with locking screws, heavy-duty swivel casters (two with brakes), patented pull-out dust filter for condenser maintenance, calibrated access port (Φ25 mm) for external probe insertion, and programmable key lock with password protection to prevent unauthorized parameter modification.
- Energy-conscious construction: non-CFC/HCFC blowing agents in insulation foam; refrigerants compliant with current F-Gas Regulation (EU No 517/2014) equivalency standards; adaptive compressor staging to match thermal load profiles.
Sample Compatibility & Compliance
The DW-86L102 supports storage of primary human tissues (e.g., hematopoietic stem cells, corneal grafts), nucleic acid extracts, monoclonal antibodies, viral vectors, and small-molecule reference standards requiring sustained sub-−70 °C conditions. Its internal chamber accommodates standard cryoboxes (e.g., 10 × 10 format), racked vials (2 mL, 5 mL), and flat-bottomed specimen containers up to 540 mm depth. While not certified to UL 4750 or NSF/ANSI 456, the unit is routinely deployed in settings adhering to CLIA, CAP, and AABB guidelines where temperature mapping validation (per ISO 14644-3 or WHO TRS 961 Annex 9) is performed prior to commissioning. Optional accessories—including NIST-traceable temperature loggers with 21 CFR Part 11-compliant software, stainless-steel drawer dividers, and validated freeze-thaw cycle documentation templates—support compliance readiness for FDA inspections and internal quality audits.
Software & Data Management
The base unit operates without embedded network connectivity; however, it features analog and digital signal outputs (0–5 V, 4–20 mA, RS485 Modbus RTU) for integration into centralized environmental monitoring systems (EMS). When paired with AUCMA’s optional TMS-3000 temperature recording module, users gain continuous logging at user-defined intervals (1–60 min), local SD-card storage (≥1 year at 10-min intervals), and exportable CSV files compatible with LIMS platforms. Audit trails include operator ID (via PIN entry), timestamped configuration changes, alarm event logs with duration and acknowledgment status, and calibration history—all meeting minimum ALCOA+ principles (Attributable, Legible, Contemporaneous, Original, Accurate, Complete, Consistent, Enduring, Available).
Applications
This ULT freezer serves core functions across translational research and diagnostic service delivery: long-term archiving of peripheral blood mononuclear cells (PBMCs) for immunotherapy trials; stabilization of thermolabile enzymes (e.g., reverse transcriptase, Cas9 variants); preservation of marine-derived metabolites for natural product screening; cold-chain validation for vaccine distribution studies; and accelerated aging tests of electronic components per MIL-STD-810G Method 502.5. Its vertical footprint (770 mm depth) enables placement in space-constrained cleanrooms or biosafety cabinets corridors, while the side-hinged door facilitates ergonomic loading in high-throughput biorepositories managing >5,000 samples weekly.
FAQ
What is the typical cooldown time from ambient to −80 °C?
Under standard laboratory conditions (25 °C, 60% RH), the DW-86L102 reaches −80 °C in approximately 180–210 minutes when empty and properly leveled.
Does the unit support 21 CFR Part 11 compliance out-of-the-box?
No—the base model lacks electronic signature capability and role-based access control; full Part 11 compliance requires integration with validated third-party EMS software and documented IQ/OQ/PQ protocols.
Can this freezer operate reliably at ambient temperatures above 32 °C?
Continuous operation is rated for ambient temperatures ≤32 °C per GB/T 20154; performance degrades above this threshold, with extended cooldown times and elevated compressor duty cycles—site-specific HVAC assessment is recommended.
Is the interior chamber stainless steel?
Yes—the inner liner is electropolished 304 stainless steel, welded seamless at corners, and passivated to resist corrosion from repeated ethanol wipe-downs and cryogen exposure.
What maintenance schedule is recommended?
Condenser coil cleaning every 3 months (or more frequently in dusty environments), annual verification of door seal compression force and sensor calibration traceability, and biannual inspection of refrigerant charge integrity using manifold gauge set and leak detector.