Empowering Scientific Discovery

TimePower TP140 Online Acid/Alkali Concentration Analyzer

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand TimePower
Model TP140
Origin Beijing, China
Manufacturer TimePower Instrument Co., Ltd.
Measurement Principle Conductivity-based concentration determination via temperature-compensated specific conductance
Measurable Media HCl (0–10.00%), NaOH (0–10.00%), H₂SO₄ (0–20.00%), NaCl (0–20.00%), HNO₃ (0–15.00%), Na₂CO₃ (0–10.00%)
Accuracy ±2% FS
Resolution 0.01%
Repeatability <1%
Temperature Compensation Pt1000 sensor, 0–60 °C
Sample Temp. Range 5–50 °C
Ambient Temp. 5–45 °C
Humidity ≤90% RH (non-condensing)
Display 128×64 dot-matrix LCD, Chinese interface
Output Signals Isolated 0–10 mA / 0–20 mA / 4–20 mA (user-selectable, load <300 Ω)
Relay Outputs 2 SPST, AC220V/3A or DC30V/3A
Enclosure Rating IP65
Power Supply AC 85–265 V, 45–65 Hz
Power Consumption ≤15 W
Dimensions (H×W×D) 145×120×150 mm
Panel Cutout 138×138 mm
Weight 0.64 kg
Data Storage Cyclic memory with auto-overflow overwrite
Data Retention (power loss) >10 years
Electrode Type Corrosion-resistant conductivity cell
Installation Options Flow-through, immersion, or flanged electrode mounting

Overview

The TimePower TP140 Online Acid/Alkali Concentration Analyzer is an industrial-grade conductivity-based process analyzer engineered for continuous, real-time monitoring of strong acid and alkali concentrations in aqueous solutions. It operates on the fundamental principle that the electrical conductivity of a solution exhibits a predictable, non-linear relationship with the concentration of dissociated electrolytes—such as HCl, NaOH, H₂SO₄, HNO₃, NaCl, and Na₂CO₃—within defined operational ranges. By measuring solution-specific conductance and applying temperature compensation (via integrated Pt1000 sensor), the instrument calculates mass percentage concentration using pre-calibrated, medium-specific conductivity-concentration curves. This approach eliminates the need for reagent addition or optical path alignment, offering robust, maintenance-light operation in demanding industrial environments including power plant demineralization systems, chemical synthesis lines, pharmaceutical purification loops, and food-grade cleaning-in-place (CIP) validation processes.

Key Features

  • High-stability corrosion-resistant conductivity electrodes designed for long-term exposure to aggressive acidic and alkaline media
  • Temperature-compensated measurement architecture with Pt1000 sensor (0–60 °C range), ensuring accuracy across variable process temperatures (5–50 °C sample range)
  • 24-bit high-resolution A/D conversion circuitry enabling precise digitization of low-level conductance signals
  • Triple-output analog configuration: user-selectable isolated current outputs (0–10 mA, 0–20 mA, or 4–20 mA) compliant with standard DCS/PLC input requirements (load <300 Ω)
  • Dual programmable relay outputs (AC220V/3A or DC30V/3A) for concentration threshold alarms, pump interlocks, or regeneration cycle initiation
  • IP65-rated NEMA-type enclosure with front-panel membrane keypad and 128×64 Chinese-language LCD display for intuitive local operation and status verification
  • Embedded cyclic data logging with automatic overwrite policy; >10-year non-volatile memory retention for calibration history and critical event timestamps
  • Low-power consumption (<15 W) and wide-input AC power supply (85–265 V, 45–65 Hz) suitable for global deployment without external transformers

Sample Compatibility & Compliance

The TP140 is validated for continuous online use with clarified, non-particulate aqueous streams containing common strong electrolytes. Its calibrated concentration ranges cover critical industrial chemistries: hydrochloric acid (0–10.00%), sodium hydroxide (0–10.00%), sulfuric acid (0–20.00%), nitric acid (0–15.00%), sodium chloride (0–20.00%), and sodium carbonate (0–10.00%). While not certified to ISO/IEC 17025 for metrological traceability, the analyzer supports GLP/GMP-aligned operational practices through configurable audit trails (via optional RS485 Modbus RTU interface), timestamped calibration logs, and password-protected parameter modification. Its design conforms to IEC 61000-6-2 (EMC immunity) and IEC 61000-6-4 (EMC emission) standards. For regulatory workflows requiring electronic record integrity, integration with third-party SCADA or MES platforms enables compliance with FDA 21 CFR Part 11 when paired with appropriate system-level controls.

Software & Data Management

The TP140 operates autonomously without external software dependencies. All configuration—including concentration range selection, temperature compensation coefficients, relay setpoints, and output scaling—is performed locally via the front-panel interface. Calibration routines follow a two-point method: zero point (deionized water or blank reference) and span point (certified standard solution at known concentration). Calibration data, including date/time stamps and operator ID (if enabled via external authentication), are stored in non-volatile memory. Optional RS485 serial communication (Modbus RTU protocol) permits remote readout of real-time concentration values, temperature, conductivity, alarm status, and diagnostic flags. When interfaced with historian systems, this enables trend analysis, batch reconciliation, and deviation detection aligned with ASTM D1125 (Standard Test Methods for Electrical Conductivity of Water) and USP (Water Conductivity).

Applications

  • Power Generation: Monitoring regenerant acid/alkali strength during ion exchange resin regeneration in condensate polishing and makeup water treatment systems
  • Chemical & Metallurgical Processing: In-line control of pickling bath concentration (HCl/H₂SO₄) and neutralization tank pH adjustment (NaOH) to maintain stoichiometric efficiency
  • Pharmaceutical Manufacturing: Verification of caustic (NaOH) and nitric acid (HNO₃) concentrations used in equipment passivation and clean-in-place (CIP) cycles per ISPE GAMP 5 guidelines
  • Food & Beverage Production: Quantitative assurance of sanitizer (e.g., NaOH, HNO₃) concentrations in reusable container washing lines to meet FDA Food Code §3-302.12 requirements
  • Wastewater Neutralization: Feedback control of acid/alkali dosing pumps in pH stabilization basins to comply with local discharge permit limits (e.g., EPA NPDES effluent criteria)

FAQ

What measurement principle does the TP140 employ?
It uses temperature-compensated electrical conductivity measurement, correlating specific conductance to mass percentage concentration via empirically derived, medium-specific calibration curves.
Can the TP140 be used for weak acids or buffered solutions?
No—it is optimized for strong, fully dissociated electrolytes. Weak acids (e.g., acetic acid) or buffered systems introduce non-linear conductance behavior outside its calibrated range and are not supported.
Is field calibration required, and how often?
Yes; initial two-point calibration is mandatory before commissioning. Routine verification is recommended every 30–90 days depending on process fouling risk, using NIST-traceable standard solutions.
Does the instrument support digital communication protocols?
Yes—optional RS485 interface with Modbus RTU protocol enables integration into distributed control systems for centralized monitoring and data archiving.
What electrode configurations are available for different installation scenarios?
Three mechanical options: flow-through cell (for pressurized pipelines), immersion probe (for tanks/vessels), and flanged insertion (for large-diameter ducts); all utilize the same corrosion-resistant electrode material and connector interface.

InstrumentHive
Logo
Compare items
  • Total (0)
Compare
0