BrandADOS
OriginGermany
ModelMWS 903
Sensor Inputs8 (2-wire or 3-wire, e.g., LCTR 903)
Power Supply24 V DC / 200 mA
Operating Temperature−10 to +40 °C
Digital Input1 (alarm mute)
Digital Outputsup to 12 alarm relays + 1 fault relay + 1 alarm label relay (all floating changeover contacts, 250 V AC / 4 A max)
Analog Output4–20 mA
Communication InterfaceRS232
Main Power Options230 V AC / 50 Hz
Enclosure Dimensions240 × 160 × 90 mm
Weight~2 kg
Ingress ProtectionIP54 per DIN 40050
Display4-line LCD with real-time value, 30-min average, and error messages
Keypad6-key menu-driven interface
Adjustable Alarm Thresholds2 independent setpoints per channel (5–100% of full scale)
Added to wishlistRemoved from wishlist 0
Add to compare
BrandADOS
OriginGermany
ModelMWS906
Sensor Inputs16 (2-wire or 3-wire, e.g., GTR 196)
Power Supply24 V DC / 2 A
Operating Temperature-10 to +45 °C
Enclosure RatingIP54 per DIN 40050
Dimensions390 × 270 × 160 mm
Weight~5 kg
Input/Output1 digital input (alarm reset), up to 48 digital outputs (floating changeover contacts, max. 250 V / 4 A), 1 alarm relay, 1 fault relay, 1 service relay, 1 alarm message relay
Analog Output4–20 mA
CommunicationRS232 (for printer or PC connection)
Main Power Options230 V AC, 50 Hz (60 VA max)
Gas Measurement RangesO₂: 0–25 % vol
CO₂0–4 % vol or 0–2000 ppm
CO0–300 ppm
NO₂0–30 ppm
H₂S0–100 ppm or 0–10,000 ppm
NH₃0–1000 ppm
H₂0–100 % LEL or 0–1000 ppm
HCl0–10 ppm or 0–100 ppm
N₂O0–1 % vol
He0–100 % vol
SF₆0–1000 ppm
O₃0–5 ppm
ClO₂0–50 ppm
CH₄0–100 % LEL
C₂H₂0–100 % LEL
C₂H₈O (ethanol)0–1000 ppm or 0–100 % LEL
C₃H₈0–100 % LEL
C₄H₁₀0–100 % LEL
C₄H₁₀O (diethyl ether)0–100 % LEL
C₇H₈ (toluene)0–100 % LEL or 0–2000 ppm
C₈H₁₀ (xylene)0–100 % LEL or 0–2000 ppm
LPG0–100 % LEL
R134a0–2000 ppm