CUBIC INSTRUMENTS Gasboard-3500 Exd Intrinsically Safe Multicomponent Gas Analyzer

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand CUBIC INSTRUMENTS
Origin Hubei, China
Model Gasboard-3500 (Explosion-Proof)
Measurement Principle Non-Dispersive Infrared (NDIR) & Electrochemical Detection (ECD)
Configuration Online Fixed-Mount Analyzer
Measured Gases CO, CO₂, CH₄, C₃H₈, O₂, H₂S
Range CO/CO₂/CH₄/C₃H₈: 0–100% vol
O₂ 0–25% vol
H₂S 0–9999 ppm
Response Time (T₉₀) <15 s (NDIR channels)
Accuracy ±1% FS (NDIR gases), ±2% FS (O₂, H₂S)
Repeatability ≤2%
Resolution 0.01% (low-range equivalent to 1 ppm)
Sample Flow Rate 0.7–1.2 L/min
Sample Pressure 2–50 kPa
Electrical Interface RS-485 / RS-232 / 4–20 mA / Ethernet
Power Supply AC 220 V, 50 Hz
Display LCD
Explosion Protection Ex d IIC T6 Gb / Ex tb IIIC T80°C Db
Certifications Compliant with GB 3836.1–2010, GB 3836.2–2010, GB 3836.8–2014

Overview

The CUBIC INSTRUMENTS Gasboard-3500 Exd is an online, fixed-mount multicomponent gas analyzer engineered for continuous, real-time monitoring in hazardous industrial environments. It integrates dual-sensing modalities—non-dispersive infrared (NDIR) spectroscopy for CO, CO₂, CH₄, and C₃H₈, and electrochemical detection (ECD) for O₂ and H₂S—within a single robust platform. The NDIR modules employ high-stability infrared sources, thermopile detectors, and optical filters tuned to the fundamental absorption bands of target gases (e.g., 4.26 µm for CO₂, 4.67 µm for CO), ensuring selectivity and immunity to spectral cross-interference. ECD cells for O₂ and H₂S are temperature-compensated and sealed with long-life electrolytes, delivering stable baseline performance over extended operational cycles. The instrument features integrated sample conditioning—including particulate filtration, condensate removal, and pressure regulation—to maintain measurement integrity under high-humidity, dusty, or variable-pressure process streams. Its Ex d IIC T6 Gb and Ex tb IIIC T80°C Db certifications confirm compliance with IEC 60079-1 and IEC 60079-31 for use in Zone 1 and Zone 21 explosive atmospheres containing Group IIC gases (e.g., hydrogen, acetylene) and combustible dusts.

Key Features

  • Explosion-proof enclosure conforming to EN/IEC 60079-1 (flameproof “d”) and IEC 60079-31 (dust ignition protection “tb”), rated for surface temperature class T6 (≤85°C) and T80°C for dust layers.
  • Dual-channel NDIR architecture per gas channel, enhancing signal stability and reducing drift-induced error—critical for unattended operation exceeding 6 months between calibration.
  • Integrated sample pre-treatment system with stainless-steel sintered filter (5 µm), PTFE-coated condensate separator, and precision mass-flow controlled bypass—eliminating need for external conditioning skids in most refinery, biogas, or landfill applications.
  • Air-based automatic zero calibration via built-in micro-diaphragm pump; supports scheduled or event-triggered zero checks without manual intervention or external gas cylinders.
  • Real-time sensor health diagnostics: monitors detector responsivity, reference cell integrity, electrochemical baseline offset, and thermal stability—logging fault codes per ISO 13849-1 diagnostic coverage requirements.
  • All wetted materials compliant with ASTM A269 TP316L stainless steel and PTFE (ASTM D4894), resistant to H₂S corrosion, acidic condensates, and oxidizing flue environments.

Sample Compatibility & Compliance

The Gasboard-3500 Exd accepts raw process gas streams meeting the specification: particulate load <1 mg/m³, dew point ≤5°C at sampling pressure, oil aerosol content <0.01 mg/m³. It is validated for continuous operation in ambient temperatures from –20°C to +50°C and relative humidity up to 95% non-condensing. Regulatory alignment includes conformity with GB 3836 series (China’s mandatory explosion-protection standard), CE marking under ATEX Directive 2014/34/EU (Category 2G/2D), and electromagnetic compatibility per IEC 61326-1:2020 (industrial environment). For regulated industries, audit-ready data logging supports 21 CFR Part 11-compliant electronic records when interfaced with validated SCADA or DCS platforms via Modbus RTU/TCP.

Software & Data Management

Field configuration and diagnostics are managed via embedded web interface (HTTP/HTTPS) or local RS-485 terminal. All measurement data—including raw detector voltages, temperature-compensated concentrations, diagnostic flags, and calibration timestamps—are timestamped and stored internally (32 MB non-volatile memory, ≥12 months at 1-min intervals). Standard protocols include Modbus RTU (RS-485), Modbus TCP (Ethernet), and 4–20 mA analog outputs with HART 7.0 support. Optional OPC UA server enables secure, structured data exchange with MES or IIoT platforms. Firmware updates are performed via signed binary packages to prevent unauthorized modification—meeting IEC 62443-3-3 SL2 security requirements.

Applications

  • Coal-fired power plants: Real-time CO monitoring in coal pulverizers and silos for fire prevention per NFPA 85 guidelines.
  • Biogas upgrading facilities: Simultaneous quantification of CH₄, CO₂, H₂S, and O₂ to optimize membrane separation and desulfurization efficiency.
  • Landfill gas collection systems: Tracking anaerobic digestion progress and detecting early-stage leachate-induced H₂S spikes.
  • Chemical manufacturing: Inerting atmosphere verification in reactors handling pyrophoric catalysts (e.g., Ni/Raney catalysts), where O₂ < 0.5% v/v is critical.
  • Research laboratories: Quantitative gas composition analysis during catalytic combustion studies, syngas reforming experiments, and microbial fermentation kinetics—supporting ASTM D1945, ISO 6974-1, and USP method validation frameworks.

FAQ

What certifications does the Gasboard-3500 Exd hold for hazardous area installation?

It carries dual certification: Ex d IIC T6 Gb per IEC 60079-1 for gaseous atmospheres (Zone 1), and Ex tb IIIC T80°C Db per IEC 60079-31 for combustible dust environments (Zone 21). Full test reports are available upon request.
Can the analyzer operate without external zero/span gas cylinders?

Yes—it performs automatic zero calibration using ambient air drawn by its internal diaphragm pump. Span calibration requires certified gas standards but can be scheduled remotely via Modbus command.
Is the instrument suitable for high-moisture biogas streams directly from digesters?

Yes, provided inlet gas is pre-filtered to remove >5 µm particulates and cooled to maintain dew point below 5°C prior to entry. The integrated condensate trap handles up to 2 mL/h liquid carryover.
How is measurement traceability maintained for regulatory reporting?

Each unit ships with factory calibration certificates traceable to NIM (National Institute of Metrology, China) standards. Internal audit logs record every calibration event, sensor fault, and firmware update with digital signatures.
What is the expected service interval for NDIR source and detector assemblies?

Rated lifetime exceeds 10 years under continuous operation at 25°C ambient, based on accelerated life testing per IEC 60721-3-3 Class 3K5. Electrochemical sensors require replacement every 24–36 months depending on H₂S exposure history.