GM-0.5A Dual-Mode Diaphragm Vacuum Pump
| Origin | Tianjin, China |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Origin Category | Domestic (China) |
| Model | GM-0.5A |
| Pricing | Available Upon Request |
| Pumping Speed | 30 L/min |
| Operating Temperature Range | 7–40 °C |
| Diaphragm Housing Surface Temperature | <60 °C |
| Ultimate Vacuum Pressure | 200 mbar (abs) |
| Maximum Positive Pressure | 0.08 MPa (g) / 40 psi (g) |
| Noise Level | <55 dB(A) |
| Inlet/Outlet Port Diameter | φ6 mm |
| Dimensions (L×W×H) | 210 × 160 × 235 mm |
| Motor Power | 160 W |
| Net Weight | 8 kg |
| Dual-Mode Operation | Simultaneous Vacuum & Positive Pressure Capability |
Overview
The GM-0.5A Dual-Mode Diaphragm Vacuum Pump is an engineered solution for laboratories requiring reliable, oil-free, and chemically resistant pressure control across both vacuum and positive-pressure regimes. Based on a robust single-stage diaphragm architecture, it operates without lubricating oil—eliminating risk of sample contamination, vapor backstreaming, or maintenance-intensive oil changes. Its core principle relies on the reciprocating motion of an elastomeric diaphragm driven by an eccentric cam, alternately expanding and compressing the pump chamber to generate suction (vacuum mode) or displacement (pressure mode). Designed for continuous-duty operation under ambient conditions (7–40 °C), the GM-0.5A delivers a stable pumping speed of 30 L/min at atmospheric intake, with thermal management ensuring diaphragm housing surface temperature remains below 60 °C during extended use—critical for preserving seal integrity and long-term repeatability.
Key Features
- Oil-free, corrosion-resistant diaphragm construction using EPDM or FKM-compatible elastomers—suitable for mild organic solvents, aqueous solutions, and inert gases.
- Dual-mode functionality: seamlessly switches between vacuum generation (down to 200 mbar absolute) and positive pressure delivery (up to 0.08 MPa gauge / 40 psi gauge) without hardware modification.
- Low-noise operation (<55 dB[A]) achieved via optimized motor damping, balanced crank mechanism, and integrated acoustic enclosure—enabling placement in shared lab spaces or near sensitive instrumentation.
- Compact footprint (210 × 160 × 235 mm) and lightweight design (8 kg) facilitate benchtop integration, mobile cart mounting, or OEM integration into analytical systems such as solid-phase extraction (SPE) manifolds or vacuum filtration stations.
- Thermally stable 160 W induction motor with built-in thermal cutoff protection ensures consistent torque delivery and prevents overheating under variable load conditions.
- Standardized φ6 mm push-fit or barbed ports simplify tubing connection using common laboratory-grade PVC, silicone, or fluoropolymer tubing (e.g., Tygon® or PTFE-lined).
Sample Compatibility & Compliance
The GM-0.5A is compatible with non-particulate, non-explosive, and non-condensable gas streams typical in routine laboratory workflows—including solvent vapors from rotary evaporation, air aspiration in vacuum desiccators, inert gas purging, and membrane degassing. It is not rated for Class I, Division 1 hazardous locations per NEC/ATEX standards nor intended for handling abrasive particulates, high-humidity saturated vapors, or aggressive halogenated compounds without upstream filtration or scrubbing. While not certified to ISO 8573-1 for compressed air purity, its oil-free operation meets general GLP requirements for traceability and contamination control in QC/QA environments. Users implementing this pump in regulated workflows (e.g., USP , ISO 14644 cleanroom support) must validate system-level performance—including downstream filtration, leak integrity, and pressure stability—per internal SOPs.
Software & Data Management
The GM-0.5A operates as a standalone analog device with no embedded firmware, digital interface, or data logging capability. It features a manual on/off rocker switch and requires external control for automation (e.g., relay-driven start/stop via PLC or programmable timer). For integration into networked lab infrastructures, users may add third-party pressure transducers (e.g., 4–20 mA or 0–10 V output) paired with SCADA or LabVIEW-compatible DAQ systems to monitor real-time vacuum/pressure status and trigger interlocks. Audit trails, electronic signatures, and 21 CFR Part 11 compliance are achievable only at the system level—not through the pump itself—and depend on the configuration of upstream controllers and documentation practices.
Applications
- Vacuum-assisted filtration (e.g., Buchner funnels, HPLC sample preparation)
- Routine solvent removal in rotary evaporators (with cold trap compatibility)
- Gas transfer and blanketing in gloveboxes or reaction vessels
- Positive-pressure actuation of pneumatic valves or fluidic manifolds
- Desiccator evacuation and vacuum drying ovens (≤100 °C chamber temp)
- Supporting vacuum-based sample introduction in elemental analyzers (e.g., CHNS/O instruments)
- OEM integration into automated liquid handlers, environmental monitoring stations, and portable field analyzers
FAQ
Can the GM-0.5A be used continuously for 24/7 operation?
Yes—when operated within its specified ambient temperature range (7–40 °C) and with adequate ventilation, the GM-0.5A supports uninterrupted duty cycles. However, periodic inspection of diaphragm elasticity and valve seat integrity is recommended every 500 operating hours.
Is this pump suitable for vacuum distillation of ethyl acetate?
It is appropriate for initial rough vacuum stages; however, for high-purity fractional distillation requiring <10 mbar, a secondary high-vacuum pump (e.g., turbomolecular or diffusion pump) is necessary.
Does the pump include an integrated vacuum gauge?
No—vacuum or pressure monitoring requires an external analog or digital gauge connected downstream of the inlet or outlet port.
What maintenance intervals are recommended?
Diaphragm and valve kit replacement is advised every 1,000–1,500 hours of cumulative runtime, depending on chemical exposure and duty cycle. Motor bearings are sealed for life and non-serviceable.
Can I connect multiple devices to a single GM-0.5A unit?
Yes—via manifold distribution with individual shut-off valves; however, total flow demand must remain ≤30 L/min to maintain specified ultimate pressure and avoid thermal overload.