Empowering Scientific Discovery

Chengmeng CM40 Diaphragm Vacuum Pump

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand Chengmeng
Origin Beijing, China
Manufacturer Type Authorized Distributor
Country of Origin China
Model CM40
Instrument Type Diaphragm Pump
Product Type Oil-Free Vacuum Pump
Ultimate Vacuum ≥0.085 MPa
Pumping Speed 40 L/min
Motor Power 250 W
Operating Voltage 220 V / 50 Hz
Motor Speed 1850–2100 rpm
Noise Level ≤50 dB(A)
Inlet Port Diameter Ø8 mm
Dimensions (L×W×H) 240 × 160 × 100 mm
Net Weight 4 kg

Overview

The Chengmeng CM40 Diaphragm Vacuum Pump is an oil-free, maintenance-light laboratory vacuum source engineered for precision and long-term reliability in demanding analytical, preparative, and process-support applications. Utilizing a corrosion-resistant PTFE-coated diaphragm and dual-head eccentric cam mechanism, the CM40 generates vacuum via positive displacement without lubricating oil—eliminating risk of hydrocarbon contamination, vapor backstreaming, or sample degradation. Its ultimate vacuum capability of ≥0.085 MPa (≈12.3 psi absolute, or ~170 mbar residual pressure) and stable pumping speed of 40 L/min at atmospheric inlet conditions make it suitable for rotary evaporation, vacuum filtration, degassing of solvents, vacuum drying ovens, and inert gas purging systems. Designed for continuous operation under moderate load, the pump features thermal overload protection and a robust aluminum alloy housing that dissipates heat efficiently while minimizing vibration transmission.

Key Features

  • Oil-free operation ensures clean, hydrocarbon-free vacuum—critical for GC sample introduction, mass spectrometry interfaces, and pharmaceutical solvent recovery.
  • PTFE-composite diaphragm and EPDM valve assemblies resist chemical attack from common organic solvents (e.g., acetone, ethanol, ethyl acetate) and mild acids/bases.
  • Integrated vacuum regulation via adjustable needle valve enables precise control over suction pressure—essential for gentle solvent removal from heat-sensitive compounds.
  • Low-noise design (<50 dB[A] at 1 m distance) meets ISO 3744 acoustic requirements for shared lab environments; no external muffler required.
  • Compact footprint (240 × 160 × 100 mm) and lightweight construction (4 kg) facilitate benchtop integration, mobile cart mounting, or placement inside fume hoods.
  • Thermally protected 250 W induction motor operates at 1850–2100 rpm with Class B insulation—rated for continuous duty per IEC 60034-1 standards.

Sample Compatibility & Compliance

The CM40 is compatible with standard laboratory vacuum accessories including Büchner funnels, Hirsch funnels, vacuum desiccators, and rotary evaporator condensers (with appropriate cold traps). Its chemically inert wetted path complies with ISO 8573-1:2010 Class 4 for solid particle content and Class 0 for oil contamination—making it appropriate for GLP- and GMP-aligned workflows where trace hydrocarbon exclusion is mandated. While not certified to UL/CSA or ATEX standards, the pump meets CE marking requirements for electromagnetic compatibility (2014/30/EU) and low-voltage safety (2014/35/EU). It supports IQ/OQ documentation protocols and is routinely deployed in QC labs conducting USP , ASTM D2879, and ISO 21620-1 compliant procedures.

Software & Data Management

The CM40 operates as a standalone analog device with manual vacuum regulation and does not incorporate digital controllers, RS-232/USB interfaces, or embedded firmware. However, its stable pressure output and repeatable flow profile enable seamless integration with third-party vacuum monitoring systems (e.g., capacitance manometers, Pirani gauges) and programmable logic controllers (PLCs) via 0–10 V or 4–20 mA analog signal inputs. Lab managers may log operational parameters—including runtime hours and pressure setpoints—using external data acquisition platforms compliant with FDA 21 CFR Part 11 when paired with audit-trail-enabled software (e.g., LabVIEW, DeltaV, or custom SCADA implementations).

Applications

  • Vacuum-assisted filtration of aqueous and organic suspensions in environmental testing (e.g., EPA Method 1664B).
  • Rotary evaporation of thermolabile natural product extracts under controlled sub-atmospheric conditions.
  • Pre-evacuation of gloveboxes and Schlenk lines prior to inert gas purging (N₂ or Ar).
  • Supporting vacuum oven drying cycles per ISO 293, ASTM D2240, and ICH Q5C stability protocols.
  • Gas sampling system priming for portable FTIR or electrochemical sensors requiring consistent negative pressure draw.

FAQ

Is the CM40 suitable for use with aggressive solvents such as chloroform or concentrated HCl?
The PTFE diaphragm and EPDM valves provide broad resistance, but prolonged exposure to halogenated solvents or strong oxidizers may accelerate elastomer fatigue. Consult the chemical compatibility chart before extended use.
Can the CM40 be operated continuously for 24/7 duty cycles?
Yes—its thermally protected motor and sealed bearing design support uninterrupted operation up to 8,000 hours between service intervals, provided ambient temperature remains below 35°C and ventilation is unobstructed.
Does the pump include a vacuum gauge or pressure sensor?
No. The CM40 is supplied without integrated pressure instrumentation; users must install external gauges compatible with the Ø8 mm inlet port.
What is the expected service life of the diaphragm under normal lab usage?
Under typical intermittent use (≤6 hrs/day), the diaphragm maintains performance for ≥12 months; replacement kits (part #CM-DK40) are available directly from Chengmeng’s authorized distribution network.
Is the CM40 compliant with ISO 13485 or FDA 21 CFR Part 820 for medical device manufacturing?
While the pump itself is not a medical device, its oil-free architecture and material certifications support compliance with ISO 13485 clause 7.5.10 (Contamination control) when validated as part of a larger production system.

InstrumentHive
Logo
Compare items
  • Total (0)
Compare
0