Empowering Scientific Discovery

Langbo LB-90 Nitrogen-Purged Precision Drying Oven

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand Langbo Instruments
Origin Jiangsu, China
Model LB-90
Instrument Type Precision Nitrogen-Purged Drying Oven
Temperature Range RT+20°C to 250°C
Temperature Uniformity ±1°C
Temperature Fluctuation ±0.5°C
Control Resolution 0.1°C
Internal Chamber Dimensions (W×D×H) 300 mm × 250 mm × 80 mm
External Dimensions (W×D×H) 450 mm × 400 mm × 350 mm
Heating Method Forced-Air Electric Heating
Inner Chamber Material Stainless Steel (AISI 304)
Power Supply 220 V ±10%, 50/60 Hz
Rated Power 500 W
Net Weight ~20 kg
Nitrogen Inlet Pressure Range 0.01–0.1 MPa
Flowmeter Range 0.3–3.0 L/min
Max. Air Exchange Rate 15 air changes per hour
Compliance Standards GB/T 211–2007, GB/T 212–2008

Overview

The Langbo LB-90 Nitrogen-Purged Precision Drying Oven is an engineered solution for moisture determination in coal and other oxygen-sensitive materials under inert atmospheric conditions. Designed specifically to meet the stringent requirements of national standard methods—namely GB/T 211–2007 “Determination of Total Moisture in Coal” and GB/T 212–2008 “Methods of Industrial Analysis of Coal”—this oven operates on the principle of controlled nitrogen purging to eliminate ambient oxygen and water vapor interference during thermal drying. Unlike conventional hot-air ovens, the LB-90 integrates a calibrated gas flow control system, precision PID temperature regulation, and a fully sealed stainless-steel chamber to ensure reproducible, bias-free moisture loss measurement. Its compact internal volume (300 × 250 × 80 mm) minimizes nitrogen consumption while enabling rapid, uniform thermal equilibration—critical for achieving high inter-laboratory repeatability in arbitration-grade coal analysis.

Key Features

  • Stainless-steel inner chamber (AISI 304) with seamless welding and high-temperature silicone gasket sealing for leak-tight inert atmosphere maintenance
  • Forced-air convection heating via embedded stainless-steel heating elements, delivering ±0.5°C temperature fluctuation and ±1°C spatial uniformity across the working zone
  • Dual-stage nitrogen management: integrated pressure-reducing valve (0.01–0.1 MPa input) and precision laminar-flow meter (0.3–3.0 L/min range) with visual float indicator
  • Intelligent digital controller with 0.1°C resolution, programmable ramp-hold profiles, and real-time display of chamber temperature and setpoint
  • Automated venting mechanism synchronized with nitrogen purge cycles to prevent over-pressurization and ensure safe exhaust of displaced gases
  • Compact footprint optimized for benchtop use; accommodates up to twelve standardized moisture pans (Φ70 mm × 35 mm height) simultaneously without compromising airflow dynamics

Sample Compatibility & Compliance

The LB-90 is validated for use with all coal ranks—including anthracite, bituminous, sub-bituminous, and lignite—as specified in GB/T 211–2007 and GB/T 212–2008. Its nitrogen-purged operation eliminates oxidative degradation of volatile matter and prevents reabsorption of ambient moisture during cooling—two primary sources of systematic error in air-drying protocols. The design conforms to ISO/IEC 17025 requirements for method validation infrastructure and supports GLP-compliant documentation when paired with external data loggers. While not certified to IEC 61000 or UL standards out-of-the-box, its electrical architecture complies with GB 4793.1–2007 (Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use).

Software & Data Management

The LB-90 operates as a standalone instrument with no embedded software stack; however, its analog temperature output (0–5 V or 4–20 mA optional upon request) enables integration with third-party SCADA systems, LIMS platforms, or validated data acquisition hardware. For laboratories requiring audit-trail capability under 21 CFR Part 11, external recording devices with electronic signature support can be interfaced via RS-485 (Modbus RTU protocol, available as add-on module). All temperature setpoints, dwell times, and nitrogen flow settings are manually configured and retained in non-volatile memory during power interruption.

Applications

  • Arbitration-level total moisture determination in commercial coal transactions per GB/T 211–2007
  • Air-dried moisture (Mad) and inherent moisture (Minh) quantification in coal quality control laboratories
  • Thermal stability testing of pharmaceutical excipients where oxidation must be suppressed during desiccation
  • Moisture loss profiling of hygroscopic polymers, catalysts, and battery electrode materials under inert atmosphere
  • Pre-conditioning of reference standards prior to gravimetric calibration or elemental analysis

FAQ

Is the LB-90 suitable for compliance with ASTM D3173 or ISO 589?
Yes—the nitrogen-purge functionality satisfies the inert-atmosphere requirement in ASTM D3173–22 (Standard Test Method for Moisture in the Analysis Sample of Coal and Coke) and aligns with the drying conditions stipulated in ISO 589:2022 (Coal — Determination of Moisture Content), provided nitrogen purity ≥99.995% is maintained.
Can the oven be used for ash content determination?
No—the LB-90 is not rated for muffle furnace temperatures (>500°C) and lacks the ceramic insulation required for ashing. It is strictly intended for low-temperature moisture removal under inert gas.
What nitrogen supply specifications are recommended?
A regulated source delivering 99.995% pure nitrogen at 0.05 MPa inlet pressure and minimum 5 L/min flow capacity is advised to sustain continuous 3 L/min purge during extended runs.
Does the unit include calibration certification?
Factory calibration certificates for temperature uniformity and controller accuracy (traceable to NIM, China) are available upon request; on-site verification using accredited PT100 probes is recommended annually per ISO/IEC 17025 clause 6.5.
Is remote monitoring supported natively?
Not natively—but optional RS-485 Modbus interface allows integration with building management systems or centralized lab monitoring dashboards when configured with appropriate gateway firmware.