BrandAUTOMESS
OriginGermany
Model6150AD-5
Detection PrincipleIonization Chamber / GM Tube (Internal) + External Probe Compatibility
Energy Range45 keV – 3 MeV (γ/X-rays)
Dose Rate Range (Digital)0.1 µSv/h – 999 mSv/h
Dose Rate Range (Analog)1 µSv/h – 1000 mSv/h
Power SupplyIEC 6LR61 9 V battery
CertificationPTB-certified (Physikalisch-Technische Bundesanstalt, Germany)
InterfaceRS232 serial port
Max. External Cable Length100 m (probe-dependent)
Alarm FunctionsProgrammable dose & dose rate thresholds
Data StorageNon-volatile memory for accumulated dose
Auto-probe RecognitionYes (e.g., “ext 17” display upon connection)
Added to wishlistRemoved from wishlist 0
Add to compare
BrandAutomess
OriginGermany
Model6150AD-6
TypePortable Dose Rate Meter
Measurement PrincipleIonization Chamber / GM Tube Based Photon Detection (γ/X-ray)
Energy Range60 keV – 1.3 MeV
Dose Rate Range (Digital)0.01 µSv/h – 9.99 mSv/h
Dose Rate Range (Analog)0.1 µSv/h – 10 mSv/h
Power SupplyIEC 6LR61 (9 V) Battery
InterfaceRS232 Serial Port
External Probe Cable Length Options1.25 m, 3 m, 10 m, 20 m, 75 m, 100 m
CompliancePTB Certified (Physikalisch-Technische Bundesanstalt, Germany)
Added to wishlistRemoved from wishlist 0
Add to compare
BrandATOMTEX
OriginBelarus
ModelAT1121
Detection PrincipleScintillation (Plastic Scintillator with Heavy Metal Mixture, 30 × 15 mm)
Measured QuantityAmbient Dose Equivalent H*(10)
Continuous Dose Rate Range50 nSv/h – 10 Sv/h
Short-Term Dose Rate Range5 µSv/h – 10 Sv/h
Pulse Dose Rate CapabilityDown to 10 ns (for AT1123/AT1123A variants only)
Energy Range15 keV – 10 MeV
β Detection CapabilityYes
Operating Temperature−30 °C to +40 °C
IP RatingIP54
Weight0.9 kg
Dimensions233 × 85 × 67 mm
Power SupplyRechargeable 6 V battery (≥12 h), AC 220 V, or DC 12 V
Data InterfaceRS-232, 300–19200 baud
Memory Capacity100 measurement records
DisplayBacklit LCD
AlarmAudible and visual threshold alerts
Added to wishlistRemoved from wishlist 0
Add to compare
BrandCapintec
OriginUSA
ModelCRC55tR
Instrument TypePortable Dose Calibrator
Measurement PrincipleIonization Chamber (Thin-Walled, High-Pressure Argon-Filled Well Chamber)
Max Activity Range250 GBq (6 Ci)
Minimum Detectable Activity0.001 MBq (0.01 µCi)
Display8" VGA Color Touchscreen
Accuracy±2%
Linearity±2%
Response Time≤2 s (standard), 4–16 s (low-activity samples)
Chamber Fill GasUltra-High-Purity Argon at 12 atm
Well DimensionsØ6.1 cm × Depth 25.4 cm
Console Dimensions42 × 23 × 27 cm
Console Weight3.4 kg
Chamber Weight13.6 kg
Power100–240 VAC, 50/60 Hz, 90 mA
InterfacesRS-232, USB
ComplianceMeets NIST-traceable calibration requirements for nuclear medicine
Added to wishlistRemoved from wishlist 0
Add to compare
BrandHoladay
OriginUSA
ModelHI3604
Frequency Range30–2000 Hz
Electric Field Range1 V/m – 200 kV/m
Magnetic Field Range0.2 mG – 20 G
Accuracy±1% (typical, referenced to NIST-traceable calibration)
DisplayLCD with unit selection (mG, G, V/m, kV/m) and bar-graph visualization
Data Storage127 internal readings
PowerTwo 9V alkaline batteries (NEDA 1604A)
Operating Temperature10–40 °C
Relative Humidity5–95% RH (non-condensing)
RMS DetectionTrue RMS for non-sinusoidal waveforms
Polarization ResponseSingle-axis (electric and magnetic sensors optimized for linear polarization alignment)
OutputAnalog preamp output (1 mV per unit full scale) via 3.5 mm audio jack
Optional Digital InterfaceFiber-optic link compatible with HI-3616 remote control unit
Added to wishlistRemoved from wishlist 0
Add to compare
BrandSARAD
OriginGermany
ModelRTM2200
Sampling MethodActive
Measurement Range0–10,000,000 Bq/m³
Minimum Detectable Concentration1 Bq/m³
Response Time12 min (Po-214 excluded) / 120 min (Po-214 included)
Detector TypeSilicon semiconductor alpha spectrometer with deep ion implantation
Sensitivity3.2 cpm/(kBq/m³) (Po-214 excluded) / 7 cpm/(kBq/m³) (Po-214 included)
Chamber Volume~135 mL
Built-in Pump Flow Rate300–800 mL/min
Operating Interface240 × 128 pixel graphical touchscreen
Data StorageSD card with GPS tagging (optional)
Communication InterfacesUSB, RS232 (supports GSM modem, ZigBee, or Wi-Fi adapter upgrade)
Dimensions235 × 140 × 255 mm
Environmental Sensors (standard)Air temperature, relative humidity, flow rate, barometric pressure, battery status
Added to wishlistRemoved from wishlist 0
Add to compare
BrandIBA
OriginGermany
ModelKerma X-plus IDP
ComplianceIEC 60580, IEC 60601-1, IEC 60601-2-43, FDA 21 CFR 1020, EU MDD 93/42/EEC
Transmission>75%
kV Range40–150 kV
DAP Range0.1–99,999,999.99 µGy·m²
DAP Resolution0.01 µGy·m²
DAP Linearity±5% (<10 µGy·m²), ±2% (≥10 µGy·m²)
DAP Rate Range0.01–3,000.00 µGy·m²/s
DAP Rate Resolution0.01 µGy·m²/s
DAP Rate Linearity±5% (<1 µGy·m²/s), ±2% (≥1 µGy·m²/s)
Exposure Time Resolution0.5 ms
Minimum Exposure Time0.01 s
Repeatability<2%
Energy Dependence (DAP)±8% (40–150 kV)