[BrandPersee
OriginBeijing, China
Manufacturer TypeDirect Manufacturer
Instrument ArchitectureDual-Beam
DetectorPhotomultiplier Tube (PMT)
Wavelength Range185 nm – 900 nm
Wavelength Accuracy±0.1 nm (at D₂ line, 656.1 nm)
Wavelength Repeatability≤0.05 nm (at D₂ line, 656.1 nm)
Spectral BandwidthContinuously Adjustable from 0.1 nm to 5.0 nm
Stray Light≤0.0001% T (NaI, 220 nm)
Photometric Range–6 Abs to +6 Abs
Baseline Flatness±0.0008 Abs
Absorbance Accuracy±0.002 Abs (0–0.5 Abs), ±0.004 Abs (0.5–1.0 Abs), ±0.3% T (>1.0 Abs)
Absorbance Repeatability≤0.001 Abs (0–0.5 Abs), ≤0.002 Abs (0.5–1.0 Abs), ≤0.1% T (>1.0 Abs)
Drift≤0.1% T/h
Optical SystemHybrid Czerny-Turner Dual-Monochromator with High-Reflectivity Coated Optics
Light SourcesIntegrated Deuterium Lamp, Tungsten-Halogen Lamp, and Mercury Calibration Lamp (Auto-Switching)]
Added to wishlistRemoved from wishlist 0
Add to compare
BrandPurkinje
OriginBeijing, China
Manufacturer TypeDirect Manufacturer
ModelTU-1901 / TU-1900
Optical SystemDual-Beam
DetectorPhotomultiplier Tube (PMT)
Wavelength Range190 nm – 900 nm
Wavelength Accuracy±0.3 nm (auto-calibrated at startup)
Wavelength Repeatability≤0.15 nm
Spectral Bandwidth0.1 nm, 0.2 nm, 0.5 nm, 1.0 nm, 2.0 nm, 5.0 nm
Stray Light≤0.01%T at 220 nm (NaI), ≤0.05%T at 360 nm (NaNO₂)
Photometric Accuracy±0.002 Abs (0–0.5 Abs), ±0.004 Abs (0.5–1.0 Abs), ±0.3%T (0–100%T)
Photometric Range–4.0 to +4.0 Abs
Baseline Flatness±0.001 Abs
Baseline Drift0.0004 Abs/h (at 500 nm, 0 Abs, after 2 h warm-up)
Photometric Noise±0.0004 Abs
Added to wishlistRemoved from wishlist 0
Add to compare
BrandPersee
OriginBeijing, China
Manufacturer TypeOriginal Equipment Manufacturer (OEM)
Instrument ArchitectureSingle-Beam
DetectorPhotomultiplier Tube (PMT)
Wavelength Range190–1100 nm
Wavelength Scanning ModeMotorized Automatic Wavelength Drive
Spectral Bandwidth0.5 / 1.0 / 2.0 / 5.0 nm (TU-1810S models, variable slit)
Wavelength Accuracy±0.3 nm (with automatic on-boot wavelength calibration)
Wavelength Repeatability≤0.2 nm
Stray Light≤0.02% T at 220 nm (using NaI solution), ≤0.05% T at 360 nm (using NaNO₂ solution)
Added to wishlistRemoved from wishlist 0
Add to compare
BrandPersee
OriginBeijing, China
Manufacturer TypeDirect Manufacturer
Product CategoryDomestic
ModelSolid Sample Testing Solution
Price RangeUSD 14,000 – 21,000 (FOB)
Optical ArchitectureDual-Wavelength Configuration
DetectorPhotomultiplier Tube (PMT)
Wavelength Scanning ModeAutomatic
Stray Light< 0.0004% (4 × 10⁻⁴ %)
Compatible Accessories150 mm Integrating Sphere (Standard), 60 mm Integrating Sphere (Standard), Eyewear-Specific Measurement Fixture
Sample TypesOpaque/Translucent Solids, Powders, Turbid Liquids
Measurement ModesDiffuse Reflectance, Specular Reflectance, Total Transmittance, Diffuse Transmittance
Added to wishlistRemoved from wishlist 0
Add to compare
BrandPurkinje
OriginBeijing, China
Manufacturer TypeDirect Manufacturer
Instrument ArchitectureDual-Beam
Detector TypePhotomultiplier Tube (PMT)
Wavelength Range185–900 nm
Wavelength Accuracy±0.1 nm (at D₂ line, 656.1 nm)
Wavelength Repeatability≤0.05 nm (at D₂ line, 656.1 nm)
Spectral BandwidthContinuously Adjustable from 0.1 to 5.0 nm
Stray Light≤0.00004%T (NaI, 220 nm)
Stray Light≤0.00002%T (NaNO₂, 360 nm)
Photometric Range–8 to +8 Abs
Baseline Flatness±0.0005 Abs
Photometric Accuracy±0.002 Abs (0–0.5 Abs), ±0.004 Abs (0.5–1.0 Abs), ±0.3%T (>1.0 Abs)
Photometric Repeatability≤0.001 Abs (0–0.5 Abs), ≤0.002 Abs (0.5–1.0 Abs), ≤0.1%T (>1.0 Abs)
Drift≤0.1%T/h
Added to wishlistRemoved from wishlist 0
Add to compare
BrandPersee
OriginBeijing, China
Manufacturer TypeDirect Manufacturer
Product OriginDomestic (China)
ModelT700/T600
Optical DesignDual-Beam
Detector TypePhotocell
Wavelength Range190–1100 nm
Wavelength Accuracy±0.3 nm
Wavelength Repeatability≤0.1 nm
Spectral Bandwidth5 nm / 2 nm / 1 nm / 0.5 nm / 0.2 nm (selectable)
Stray Light≤0.01% T at 220 nm
Absorbance Range–4 to +4 Abs
Scan SpeedUp to 30,000 nm/min
InterfaceEthernet port
Display9.7-inch capacitive touchscreen
ComplianceMeets JJG 178–2015 (Chinese National Verification Regulation for UV-Vis Spectrophotometers, Class I accuracy)
Added to wishlistRemoved from wishlist 0
Add to compare
BrandPersee
OriginBeijing, China
Manufacturer TypeOriginal Equipment Manufacturer (OEM)
Product TypeDomestic
ModelT6 Series
Price RangeUSD 1,400–2,800
Optical DesignPseudo-Double-Beam
DetectorPhotomultiplier Tube (PMT)
Wavelength Range190–1100 nm
Wavelength ScanningMotorized Auto-Positioning
Spectral Bandwidth2.0 nm
Wavelength Accuracy±1 nm
Wavelength Repeatability≤0.2 nm
Stray Light0.05% T at 220 nm
Added to wishlistRemoved from wishlist 0
Add to compare
BrandPersee
OriginBeijing, China
Manufacturer TypeDirect Manufacturer
Product CategoryDomestic
ModelUV-Vis Automated Liquid Sampling Solution
Price RangeUSD 14,000 – 21,000
Optical DesignDouble-beam
DetectorPhotomultiplier Tube (PMT)
Wavelength Range190–1100 nm
Spectral Bandwidth1.8 nm (typical)
Wavelength Accuracy±0.3 nm
Stray Light<0.05% T at 220 nm and 360 nm
Automation LevelFully Automated Wavelength Scanning & Sample Handling
Sample CapacityThree Independent Sample Carriers Supporting Configurable Racks (50 mL × 10 wells, 25 mL × 18 wells, or 20 mL × 18 wells)
Positioning SystemProgrammable 3-Axis Robotic Arm with Sub-millimeter Repeatability
Cleaning MechanismPeristaltic Pump–Driven Dynamic Immersion Rinse with Solvent Exchange Protocol
Added to wishlistRemoved from wishlist 0
Add to compare
BrandPurkinje
OriginBeijing, China
Manufacturer TypeOEM Manufacturer
Instrument ArchitectureDual-Beam
Detector TypePhotomultiplier Tube (PMT)
Wavelength Range190–900 nm
Wavelength Accuracy±0.3 nm (auto-calibrated at startup)
Wavelength Repeatability≤0.15 nm
Spectral BandwidthContinuously Variable from 0.1 to 5.0 nm (0.1 nm increments)
Stray Light≤0.01% T at 220 nm (NaI), ≤0.05% T at 360 nm (NaNO₂)
Baseline Flatness±0.001 Abs
Drift≤0.1%/h
Dimensions610 × 570 × 300 mm
Weight39 kg
Added to wishlistRemoved from wishlist 0
Add to compare
BrandPersee
OriginBeijing, China
Manufacturer TypeDirect Manufacturer
Product OriginDomestic (China)
ModelT400 / T500
Price RangeUSD 1,400 – 4,200
Optical DesignSingle-Beam
Detector TypeSilicon Photocell
Wavelength Range325–1100 nm (T400) / 190–1100 nm (T500)
Wavelength Accuracy±0.2 nm (T400) / ±0.1 nm (T500)
Wavelength Repeatability≤0.4 nm (T400) / ≤0.2 nm (T500)
Spectral Bandwidth2 nm
Stray Light≤0.05%T at 360 nm (T400) / ≤0.05%T at 220 nm (T500)
Wavelength Scanning Speed2500 nm/min
Wavelength Setting Speed7000 nm/min
Dimensions (W×D×H)470 × 400 × 210 mm
Weight11 kg
Display7-inch Color Touchscreen
Data ExportUSB Flash Drive Support
Sample CompartmentOpen-Design with Integrated Cuvette Storage Rack
Software InterfacePC-Connectable Workstation with GLP Compliance Features
Firmware UpgradabilityOver-the-Air (OTA) for Application Module Expansion (e.g., Protein Quantitation, Pesticide Residue Analysis, Water Quality Testing)
Show next