- All
- Favorite
- Popular
- Most rated
| Brand | Persee |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Instrument Type | Laboratory Gas Chromatograph |
| Model | G5 |
| Application Scope | General-Purpose |
| Oven Temperature Range | Ambient +7°C to 400°C |
| Max. Number of Simultaneous Inlet/Detector Mounts | 3 |
| Temperature Program Steps | 10-step |
| Inlet/Detector Configuration Flexibility | Modular |
| Carrier Gas Control | Digital Precision Flow/Pressure Regulation |
| Compliance | Designed to meet core requirements of ISO/IEC 17025, ASTM D3606, USP <621>, and FDA 21 CFR Part 11 for data integrity |
| Brand | Persee |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Instrument Type | Benchtop Laboratory Gas Chromatograph |
| Application Scope | General-Purpose Analytical Chemistry |
| Maximum Column Oven Ramp Rate | 0–120 °C/min |
| Simultaneous Inlet Configuration | Up to 3 Independent Inlets |
| Compatible Detectors | TCD, FID, ECD, FPD, NPD, PDD, MSD |
| Flow Control System | Precision Smart Flow Control (SFC) |
| Inlet Temperature Capability | Not Specified |
| Pressure & Flow Ranges | Not Specified |
| Cooling Rate | Not Specified |
| Cloud-Enabled Features | Method Optimization Engine, Remote Fault Diagnostics, Centralized Method Library |
| Brand | Persee |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Instrument Type | Benchtop Laboratory Gas Chromatograph |
| Model | G600 |
| Temperature Control Range | 7 °C above ambient to 400 °C |
| Oven Ramp Rate | 0.1–40 °C/min (adjustable in 0.1 °C increments) |
| Maximum Injector Temperature | 450 °C |
| Injector Configuration | Up to three independently configurable inlets |
| Column Oven Cooling | Forced-air active cooling |
| Detector Options | FID, TCD, ECD, PID (field-selectable) |
| Compliance | GLP/GMP-ready software audit trail, 21 CFR Part 11–compatible data handling |
| Brand | Purkinje |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Instrument Type | Laboratory Gas Chromatograph |
| Model | GC1100 |
| Temperature Control Range | Ambient +6°C to 399°C |
| Oven Ramp Rate | 0–30°C/min |
| Carrier Gas Flow Range | 0–1200 mL/min |
| Carrier Gas Pressure Range | 0–970 kPa |
| Injector Maximum Temperature | 200°C |
| Injector Pressure Setting Range | 0–970 kPa |
| Injector Total Flow Setting Range | 0–1200 mL/min |
| Column Oven Temperature Accuracy | ±0.1°C |
| Column Oven Temperature Gradient | ±1.5% |
| TCD Sensitivity | ≥4000 mV·mL/mg (n-hexadecane) |
| TCD Noise | ≤0.04 mV |
| TCD Baseline Drift | ≤0.2 mV/30 min |
| FID Detection Limit | ≤8×10⁻¹² g/s (n-hexadecane) |
| FID Noise | ≤0.02 mV |
| FID Baseline Drift | ≤0.2 mV/30 min |
| Dimensions (L×W×H) | 562×550×492.5 mm |
| Weight | 44 kg |
Show next