Empowering Scientific Discovery

Chengdu Instrument Factory

Categories
  • All
  • Favorite
  • Popular
  • Most rated
Added to wishlistRemoved from wishlist 0
Add to compare
BrandSCChengyi
OriginSichuan, China
ModelQFS601
Measurement Range-80 to +20 °C dew point (≈ 0.5–23,632 µL/L H₂O)
Accuracy≤ ±1 °C
Response Time<30 s (63%), <60 s (90%) for rise at DP < -40 °C
Output Signals4–20 mA, RS232 or USB (user-specified)
Power SupplyAC 220 ± 22 V, 50–60 Hz / Internal rechargeable battery (DC)
Operating ConditionsGas Temp < +60 °C, Pressure < 2 MPa, Recommended Flow Rate: 200 mL/min
Dimensions260 × 170 × 300 mm (W × H × D)
Added to wishlistRemoved from wishlist 0
Add to compare
OriginSichuan, China
Manufacturer TypeDistributor
Origin CategoryDomestic
ModelKLY-2
PricingUpon Request
Product TypePortable Dew Point Meter
Measurement Range–80 to +20 °C dew point (approx. 0.5–23,632 µL/L H₂O)
Accuracy≤±0.5 °C (at factory calibration)
Response Time<30 s (63%), <60 s (90%) for rise at DP < –40 °C
Output Signal4–20 mA analog
Operating Media ConditionsGas temperature < +60 °C, pressure < 2 MPa, flow rate 200 mL/min
Dimensions260 mm × 170 mm × 300 mm (W×H×D)
Added to wishlistRemoved from wishlist 0
Add to compare
Brandscchengyi
OriginSichuan, China
Manufacturer TypeAuthorized Distributor
Product CategoryDomestic
ModelUSI-1AB
Product TypeOnline Dew Point Moisture Analyzer
Measurement PrinciplePhosphorus Pentoxide (P₂O₅) Electrolytic Method
Operating Temperature5–40 °C (ambient)
Supply Gas Pressure<100 kPa
Accuracy±5% of reading
Repeatability<0.08 µL/L at sample moisture ≤260 µL/L
Range0–1000 µL/L (ppmv)
Response Time (t₆₃)≤3 min
Output Signals0–10 mV and 4–20 mA
Power SupplyAC 220 V ±10%, 50–60 Hz
Sample Flow Rate100 mL/min (measurement), 1 L/min (bypass)
Dimensions380 × 300 × 170 mm
Weight6.5 kg
Added to wishlistRemoved from wishlist 0
Add to compare
BrandChengyi
OriginSichuan, China
Manufacturer TypeAuthorized Distributor
Product CategoryDomestic Lab-Scale Dew Point Analyzer
ModelUSI-3
PricingAvailable Upon Request
Instrument PrincipleElectrolytic (Phosphorus Pentoxide Sensor)
Operating Temperature5–35 °C
Accuracy±3% of total water mass
Repeatability≤1% RSD
Measurement Range0.0000–99.999 mg H₂O
Power SupplyAC 220 ± 22 V, 50–60 Hz or DC 10–45 V (for integrated dryer unit)
Carrier GasUltra-Dry Nitrogen (<10 ppmv H₂O), Flow Rate: 70 ± 20 mL/min