Empowering Scientific Discovery

GreenPrima PM8202I Online Fluoride Ion Analyzer

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand GreenPrima
Origin UK
Manufacturer Type Manufacturer
Origin Category Imported
Model PM8202I
Price Upon Request
Measurement Range 0.00–20000 ppm
Resolution 0.0 / 0.1 / 1 ppm
Accuracy ±0.01 / ±0.1 / ±1 ppm
Temperature Compensation –10 to 130 °C (manual/automatic, NTC30K or PT1000)
Operating Temperature –10 to 70.0 °C
Storage Temperature –20 to 70.0 °C
Display Backlit large-dot-matrix LCD
Language Chinese/English
Data Storage 600,000 records
Power Supply 90–260 VAC, 50/60 Hz or optional 24 VDC
Output Two isolated 4–20 mA outputs (max. loop resistance 500 Ω, accuracy 0.1% F.S.), RS485 Modbus RTU
Enclosure Rating IP66/IP67 (fully sealed, waterproof & gas-tight)
Electrode Compatibility Bsens710 (F⁻), Bsens720 (Cl⁻), Bsens730 (NH₃-N)

Overview

The GreenPrima PM8202I Online Fluoride Ion Analyzer is a robust, multi-parameter electrochemical monitoring system engineered for continuous, real-time quantification of fluoride ion (F⁻) concentration in aqueous streams. It operates on the principle of potentiometric measurement using a solid-state, ion-selective electrode (ISE)—specifically the Bsens710 fluoride-selective sensor—paired with a reference electrode to generate a stable Nernstian potential response proportional to the logarithm of F⁻ activity. The analyzer supports interchangeable ISE modules, enabling flexible configuration for simultaneous or sequential monitoring of chloride (Cl⁻) and ammoniacal nitrogen (NH₃-N) via Bsens720 and Bsens730 electrodes respectively. Designed for deployment in demanding industrial water environments—including municipal drinking water treatment, wastewater effluent discharge points, semiconductor rinse water loops, and pharmaceutical process water systems—the PM8202I delivers high reproducibility under variable pH, temperature, and ionic strength conditions when operated within its specified calibration and maintenance protocols.

Key Features

  • IP66/IP67-rated fully sealed, gas-tight enclosure—suitable for outdoor mounting and humid, corrosive atmospheres
  • Intuitive dual-language (English/Chinese) graphical user interface with backlit large-dot-matrix LCD display showing real-time F⁻ concentration, measured temperature, system time, relay status, and diagnostic flags
  • Two independent, isolated 4–20 mA analog outputs configurable for primary measurement (e.g., F⁻ ppm) and secondary parameter (e.g., temperature), each compliant with 0.1% full-scale accuracy and 500 Ω maximum loop resistance
  • RS485 serial interface supporting Modbus RTU protocol for integration into SCADA, DCS, or PLC-based supervisory control architectures
  • Onboard data logging capacity of 600,000 timestamped records—including concentration, temperature, calibration timestamps, and alarm events—with non-volatile memory retention during power loss
  • User-accessible password protection (three-tier access: operator, engineer, administrator) to prevent unauthorized configuration changes or calibration overrides
  • Flexible temperature compensation using either NTC30K thermistor or PT1000 RTD input, supporting both manual and automatic modes across –10 to 130 °C

Sample Compatibility & Compliance

The PM8202I is validated for use with low- to medium-conductivity aqueous matrices typical of potable water (0.05–1.5 mS/cm), secondary-treated wastewater (1–15 mS/cm), and ultrapure rinse water (≤0.1 µS/cm). It requires no reagents or consumables beyond periodic electrode replacement and standard calibration solutions (e.g., 10⁻⁵ to 10⁻¹ M NaF). While not certified to ISO/IEC 17025 as a standalone testing laboratory, the instrument’s measurement traceability aligns with ISO 10523 (water quality — determination of pH) and ASTM D511 (standard test methods for chloride ion in water) frameworks when deployed with documented calibration procedures. Its electrical safety design complies with IEC 61010-1 (measurement, control and laboratory equipment), and electromagnetic compatibility meets EN 61326-1 for industrial environments. For regulated facilities operating under GMP or GLP requirements, the device supports audit-ready operation through timestamped calibration logs, change history tracking, and configurable alarm event recording.

Software & Data Management

Data retrieval is supported via front-panel USB port (for local export to USB flash drive in CSV format) or remotely via Modbus RTU polling. The embedded firmware implements linear and two-point calibration routines with slope validation, offset verification, and automatic stability detection prior to acceptance. All calibration actions—including date/time stamp, solution concentrations used, measured mV values, and calculated electrode slope (%Nernst)—are stored in the internal log. Optional PC-based configuration utility (GreenPrima ConfigTool v3.x) enables advanced setup of alarm thresholds (high/low, latching/non-latching), relay hysteresis, output scaling ranges, and Modbus slave ID assignment. No cloud connectivity or proprietary subscription services are required; all functionality resides onboard.

Applications

  • Real-time fluoride dosing control in municipal drinking water fluoridation plants (compliant with WHO and EPA guidelines)
  • Effluent compliance monitoring at industrial dischargers subject to fluoride limits under national environmental regulations (e.g., UK Environmental Permitting Regulations, US Clean Water Act NPDES permits)
  • Process water quality assurance in aluminum anodizing, glass etching, and uranium enrichment support facilities where HF or NaF usage necessitates strict F⁻ containment
  • Monitoring of fluoride leaching from landfill leachate collection systems and contaminated groundwater remediation sites
  • Validation of reverse osmosis and ion exchange resin performance in pharmaceutical water-for-injection (WFI) purification loops

FAQ

What is the recommended calibration frequency for the Bsens710 fluoride electrode?
Calibration is advised before initial commissioning, after electrode replacement, following extended idle periods (>72 h), and at least once per 24–72 h during continuous operation—depending on sample matrix fouling risk and regulatory reporting requirements.

Can the PM8202I measure fluoride in high-fluoride industrial waste streams (>5000 ppm)?
Yes—within its 0.00–20000 ppm range—but accurate measurement requires appropriate dilution and matrix-matched calibration standards. High-ionic-strength samples may necessitate ionic strength adjuster (ISA) addition during calibration to maintain activity coefficient consistency.

Does the analyzer support automatic temperature compensation with external PT1000 sensors?
Yes—both NTC30K and PT1000 inputs are natively supported, with auto-detection and configurable sensor mapping in the setup menu.

Is the 4–20 mA output intrinsically safe for hazardous area installation?
No—the PM8202I is rated for general-purpose industrial use (non-hazardous locations only). For Zone 1/2 or Class I Div 2 environments, an externally mounted IS barrier must be installed in series with the analog output circuit.

How is data integrity ensured during power interruption?
All logged data, configuration parameters, and calibration records are retained in non-volatile flash memory. The real-time clock maintains time/date continuity for up to 72 h on internal backup capacitance.