- All
- Favorite
- Popular
- Most rated
| Brand | Xufeng |
|---|---|
| Origin | Shanghai, China |
| Model | XF-3241 |
| Temperature Control Range | 20–380 °C |
| Temperature Stability | ±1 °C (20–380 °C) |
| Fuel Flow Rate | 3.0 ± 0.3 mL/min |
| Sample Volume | 600 mL |
| Test Duration | 150 ± 2 min |
| Nitrogen Pressure Regulation Range | 0–3.45 MPa |
| Fuel System Pressure | 3.45 ± 0.345 MPa |
| Differential Pressure Measurement Range | 0–340 mmHg |
| Differential Pressure Accuracy | Class 0.3 |
| Differential Pressure Resolution | 0.1 mmHg |
| Cooling Water Flow Rate | 38 ± 8 L/h |
| Cooling Water Pressure | 0.1–0.4 MPa |
| Power Supply | AC 220 V ± 10% |
| Dimensions (Bath) | 500 × 190 × 424 mm (L × W × D) |
| Overall Dimensions | 550 × 450 × 600 mm (L × W × H) |
| Compliance | ASTM D3241, GB/T 9169 |
| Brand | XuFeng |
|---|---|
| Origin | Shanghai, China |
| Model | XF-2272A |
| Temperature Control Range | Ambient to 150 °C (±0.1 °C) |
| Rotation Speed | 100 ±5 rpm |
| Oxygen Bomb Angle | 30° from Horizontal |
| Pressure Measurement Accuracy | ±0.1% of 620 kPa |
| Data Storage | 100 Test Records |
| Power Consumption | 2.6 kW |
| Dimensions | 450 W × 500 D × 650 H mm |
| Weight | ~40 kg |
| Brand | Xufeng |
|---|---|
| Origin | Shanghai, China |
| Model | OS-F6 |
| Compliance | SH/T 0175, ASTM D2274, ISO 12205, GB/T 12581, ASTM D943 |
| Bath Dimensions | 500 × 190 × 424 mm (L×W×D) |
| Overall Dimensions | 870 × 300 × 1300 mm (L×W×H) |
| Oxidation Tube Ports | 6 |
| Heating Power | 3.0 kW |
| Pressure Regulation Range | 0–588 kPa (0–6 kgf/cm²) |
| Flow Control Range | Domestic Flowmeter: 1.0–4.0 L/h (16.7–66 mL/min) |
| Imported Flowmeter | 5–50 mL/min |
| Flow Stability | ±1.6 mL/min at 50 mL/min (ambient temperature ≤20°C) |
| Temperature Control Range | (Ambient +10)°C to 100°C |
| Temperature Fluctuation & Uniformity | ≤±0.1°C |
| Construction | Full 316 stainless steel enclosure |
| Diagnostic Alarms | Sensor Open/Short, Heater Open, Gas/Water Under-Pressure |
| Brand | Xufeng |
|---|---|
| Origin | Shanghai, China |
| Model | XF-2272 |
| Standards Compliance | SH/T 0193, ASTM D2112, ASTM D2272, GB/T 8018 |
| Temperature Control Range | 0–150 °C (±0.1 °C accuracy) |
| Pressure Measurement Accuracy | ±0.1% |
| Data Storage Capacity | 100 test records |
| Power Consumption | 2.6 kW |
| Dimensions (W×D×H) | 620 × 780 × 1150 mm |
| Weight | ~85 kg |
| Interface | RS-232C |
| Display | Large-screen LCD with Windows-based GUI |
| Brand | Xufeng |
|---|---|
| Origin | Shanghai, China |
| Model | OS-T12 |
| Standard Compliance | SH/T 0206, IEC 61125:1992, ASTM D2440 |
| Bath Dimensions | 550 W × 220 D × 250 H mm |
| Overall Dimensions | 650 W × 320 D × 900 H mm |
| Oxidation Tube Capacity | 12 positions |
| Heating Power | 2.4 kW |
| Heating Medium | Silicone oil or mineral oil (open-cup flash point > 200 °C) |
| Pressure Regulation Range | 0–588 kPa (0–6 kgf/cm²) |
| Air Flow Rate Adjustment | 5–50 mL/min |
| Temperature Control Range | (Ambient + 10) °C to 120 °C |
| Temperature Stability | ≤ ±0.1 °C |
| Temperature Uniformity | ≤ ±0.1 °C |
| Control System | Microprocessor-based with real-time diagnostics, auto-timing, and fault logging |
| Optional Configuration | OS-T12+ with imported precision mass flow meter |
Show next