Empowering Scientific Discovery

PerkinElmer Clarus 680/600T Gas Chromatography-Mass Spectrometry (GC/MS) System

Add to wishlistAdded to wishlistRemoved from wishlist 0
Add to compare
Brand PerkinElmer
Origin USA
Model Clarus 680/600T
Mass Range 1–1200 u
Scan Speed Up to 60 scans/sec
Detection Limit <1 pg octafluoronaphthalene (OFN), S/N ≥ 180:1 RMS
Column Oven Rapid heating and cooling capability
Warranty 6 months
Operational Age 4 years
Instrument Type Quadrupole GC/MS
Compliance Designed for ASTM, ISO 17025, USP <621>, and FDA 21 CFR Part 11–ready software environments
Software TurboMass™ v8.x with GLP/GMP audit trail support
Included Components GC unit (Clarus 680), MS detector (Clarus 600T), autosampler, vacuum pump, dedicated workstation PC, GC/MS interface hardware, and consumables kit

Overview

The PerkinElmer Clarus 680/600T is a high-performance, quadrupole-based gas chromatography–mass spectrometry (GC/MS) system engineered for precision, reproducibility, and operational efficiency in regulated and research laboratory environments. Combining the Clarus 680 gas chromatograph—featuring one of the industry’s fastest column oven ramp rates (up to 120 °C/min) and rapid cooldown—and the Clarus 600T mass spectrometer, this integrated platform delivers robust separation power with real-time, high-fidelity mass spectral acquisition. The system operates on electron ionization (EI) and optional chemical ionization (CI) modes, supporting both quantitative and qualitative analysis across volatile and semi-volatile organic compounds. Its design adheres to core principles of analytical traceability: stable ion source temperature control, high-vacuum integrity (<5 × 10⁻⁶ Torr), and calibrated mass axis stability over extended run sequences.

Key Features

  • Rapid GC cycle times enabled by the Clarus 680’s advanced oven architecture—capable of heating from 35 °C to 350 °C in under 30 seconds and cooling from 350 °C to 50 °C in under 2.5 minutes.
  • Clarus 600T quadrupole mass spectrometer with mass range 1–1200 u, offering high mass accuracy and resolution across full-scan and selected-ion monitoring (SIM) acquisition modes.
  • Scan speed up to 60 full spectra per second—critical for coupling with fast GC methods and narrow peak elution profiles.
  • Detection sensitivity validated at ≤1 pg octafluoronaphthalene (OFN) with signal-to-noise ratio ≥180:1 (RMS, m/z 272), meeting or exceeding EPA Method 8270D and ASTM D7622 performance criteria.
  • TurboMass™ v8.x software with embedded method validation tools, electronic signature support, and 21 CFR Part 11–compliant audit trail functionality—including user access logs, parameter change history, and raw data immutability controls.
  • Pre-configured hardware integration: fully aligned GC-MS interface, turbomolecular vacuum pumping system (with backing pump), and factory-calibrated electron multiplier detector.

Sample Compatibility & Compliance

The Clarus 680/600T accommodates standard 0.25–0.53 mm ID capillary columns (e.g., DB-5ms, HP-5MS, Rxi-5Sil MS) and supports split/splitless, PTV, and cold-on-column injection techniques. It is routinely deployed for environmental analysis (PAHs, PCBs, pesticides), pharmaceutical residual solvent testing (ICH Q3C), food safety screening (mycotoxins, flavor volatiles), and forensic toxicology workflows. System validation documentation includes IQ/OQ protocols aligned with ISO/IEC 17025:2017 and GxP-compliant instrument qualification frameworks. All refurbished units undergo full functional verification per PerkinElmer service bulletins—including tune report generation, retention time reproducibility testing (<0.02 min RSD), and mass calibration stability assessment over 24-hour continuous operation.

Software & Data Management

TurboMass™ software provides unified control of GC parameters, MS acquisition, and post-run processing within a single interface. Key capabilities include automated spectral deconvolution (AMDIS), library search against NIST/EPA/ Wiley databases, quantitative batch processing with internal standard normalization, and customizable report templates compliant with LIMS export standards (e.g., .csv, .pdf, .xml). Audit trail records capture timestamped entries for all user actions—including method edits, data reprocessing, and report generation—with role-based permissions (administrator, analyst, reviewer). Raw data files (.qgd) are stored in a proprietary but exportable format compatible with third-party spectral interpretation platforms.

Applications

  • Environmental monitoring: Targeted and non-targeted screening of chlorinated hydrocarbons, phthalates, and emerging contaminants in water, soil, and air matrices.
  • Pharmaceutical quality control: Residual solvent quantitation per ICH Q3C guidelines; extractables/leachables profiling in packaging studies.
  • Food and agriculture: Pesticide multiresidue analysis (MRM-GC/MS/MS-ready), adulteration detection (e.g., olive oil authenticity), and aroma compound profiling.
  • Academic and industrial R&D: Reaction monitoring, metabolite identification, and structural elucidation of synthetic intermediates via EI spectral interpretation.
  • Forensics and clinical labs: Volatile organic compound (VOC) profiling in blood/headspace samples; drug metabolite confirmation using retention index matching and spectral match scoring.

FAQ

Is this a factory-refurbished unit or dealer-refurbished?
This Clarus 680/600T system has been comprehensively refurbished by an ISO 9001-certified third-party service provider specializing in PerkinElmer instrumentation, including replacement of critical wear components (ion source, filament, detector, GC injectors), full vacuum integrity testing, and performance verification against original factory specifications.
What documentation accompanies the system?
Each unit ships with a complete refurbishment dossier: pre- and post-service tune reports, IQ/OQ checklists, calibration certificates traceable to NIST standards, TurboMass™ software license transfer documentation, and a 6-month limited warranty covering parts and labor.
Can the system be configured for compliance with FDA 21 CFR Part 11?
Yes—the installed TurboMass™ v8.x software includes configurable electronic signature modules, audit trail logging, and data integrity safeguards required for regulated environments. Validation support packages (including URS, FRS, and test scripts) are available upon request.
Are consumables and spare parts included?
A starter consumables kit is provided, containing one GC column (30 m × 0.25 mm × 0.25 µm), five septa, ten liner O-rings, one ion source cleaning kit, and two filaments. Extended spares inventory is maintained for immediate dispatch.
What level of technical support is offered post-purchase?
Remote diagnostics and application support are provided for the duration of the warranty period. On-site service contracts—including preventive maintenance visits and emergency response—are available globally through authorized PerkinElmer service partners.

InstrumentHive