Model PM8202I
Measurement Range 0.00–20,000 ppm F⁻
Resolution 0.0 / 0.1 / 1 ppm (user-selectable)
Accuracy ±0.01 / ±0.1 / ±1 ppm (range-dependent)
Temperature Compensation -10–130 °C (NTC30K or PT1000 sensor, manual/automatic)
Operating Temperature -10–70.0 °C
Storage Temperature -20–70.0 °C
Display Backlit large-dot-matrix LCD
Language English & Chinese
Data Storage 600,000 records
Power Supply 90–260 VAC, 50/60 Hz or optional 24 VDC
Output Two isolated 4–20 mA outputs (max. loop resistance 500 Ω, accuracy 0.1% FS), RS485 Modbus RTU
Enclosure Rating IP66/IP67 (fully sealed, waterproof and gas-tight)
Calibration Manual or scheduled two-point calibration using standard solutions
Security Password-protected configuration interface

Overview

The GreenPrima PM8202I Online Fluoride Ion Analyzer is a robust, multi-parameter electrochemical monitoring system engineered for continuous, real-time quantification of fluoride (F⁻), chloride (Cl⁻), and ammoniacal nitrogen (NH₃-N) in aqueous matrices. It operates on the principle of ion-selective potentiometry—utilizing solid-state or polymer-membrane ion-selective electrodes (ISEs) to generate a Nernstian voltage response proportional to the logarithm of target ion activity. The system integrates the Bsens710 fluoride electrode, Bsens720 chloride electrode, and Bsens730 ammonia electrode with a centralized controller, enabling flexible deployment across single-ion or tri-analyte configurations. Designed for unattended operation in demanding field environments, the PM8202I delivers stable baseline performance under variable pH, flow, and temperature conditions—critical for regulatory compliance in drinking water distribution, wastewater treatment effluent monitoring, and industrial process water management.

Key Features

  • IP66/IP67-rated fully sealed enclosure—resistant to dust ingress, water immersion, and corrosive ambient gases common in pump stations and chemical dosing rooms
  • Large backlit dot-matrix LCD with dual-language (English/Chinese) menu navigation and real-time display of ion concentration, temperature, timestamp, relay status, and calibration history
  • User-configurable resolution (0.0 / 0.1 / 1 ppm) and accuracy tiers aligned with application-specific reporting requirements (e.g., WHO drinking water guideline of 1.5 mg/L F⁻)
  • Integrated temperature compensation using either NTC30K thermistor or PT1000 platinum RTD—supporting both manual input and automatic in-situ correction per ISO 10523 and ASTM D1293
  • Two independent, galvanically isolated 4–20 mA analog outputs—each assignable to measured ion concentration or temperature, compliant with IEC 61000-6-2/4 EMC standards
  • RS485 Modbus RTU communication port for seamless integration into SCADA, DCS, or PLC-based supervisory systems
  • Password-protected parameter editing and calibration access—enabling role-based operational security per GLP and ISO/IEC 17025 documentation protocols

Sample Compatibility & Compliance

The PM8202I is validated for use with raw surface water, clarified potable water, secondary-treated wastewater, and recirculated industrial cooling water—provided samples meet standard ISE compatibility criteria: pH 4–10 (for fluoride), turbidity < 5 NTU, suspended solids < 50 mg/L, and absence of high concentrations of interfering ions (e.g., OH⁻, Al³⁺, Fe³⁺) without prior conditioning. Electrode selection follows ISO 10304-1 for ion chromatography cross-validation and supports method verification per US EPA Method 300.0 (for fluoride) and ISO 7888 (for chloride). The instrument’s data integrity architecture—including non-volatile memory logging of 600,000 timestamped records with audit trail—meets foundational expectations for FDA 21 CFR Part 11 electronic record retention where local regulatory frameworks apply.

Software & Data Management

Data acquisition and configuration are managed via embedded firmware—not external PC software—ensuring deterministic response time and eliminating dependency on third-party drivers. All measurements are time-stamped with internal RTC (real-time clock) and stored with metadata including electrode ID, calibration date, slope/mV offset, and temperature at measurement. Export is supported via USB mass storage mode (FAT32) for CSV-formatted batch retrieval. The system supports scheduled auto-calibration triggers and configurable alarm thresholds with dry-contact relay outputs for dose control or bypass activation—facilitating closed-loop feedback in fluoridation dosing systems per WHO Technical Report Series No. 929.

Applications

  • Continuous monitoring of fluoride addition in municipal drinking water plants to maintain 0.6–1.2 mg/L compliance per WHO and EU Directive 98/83/EC
  • Effluent surveillance at wastewater treatment plants discharging to sensitive aquatic ecosystems subject to fluoride discharge limits (e.g., EPA Effluent Guidelines, 40 CFR Part 423)
  • Process water quality assurance in semiconductor fabrication, aluminum anodizing, and phosphate fertilizer production—where residual fluoride impacts equipment corrosion and product purity
  • Environmental monitoring of groundwater near landfill leachate plumes or phosphate mining sites, where natural or anthropogenic fluoride mobilization poses ecological risk
  • Validation support for laboratory reference methods during method transfer studies under ISO/IEC 17025 accreditation

FAQ

What electrode types are compatible with the PM8202I controller?
The system is designed exclusively for GreenPrima Bsens-series ISEs: Bsens710 (fluoride), Bsens720 (chloride), and Bsens730 (ammonia). Cross-brand electrode integration is not supported due to proprietary signal conditioning and calibration algorithms.
Does the analyzer support automatic temperature compensation during fluoride measurement?
Yes—both NTC30K and PT1000 temperature sensors are supported, with automatic compensation applied in real time using the extended Debye-Hückel equation as implemented per ISO 5667-16 Annex B.
How frequently must the fluoride electrode be calibrated?
Calibration interval depends on matrix stability and required uncertainty; typical practice is daily for critical potable water applications, weekly for industrial process streams, and after any maintenance event—per ISO 17034 and manufacturer-recommended two-point (low/high) standardization.
Can the 4–20 mA outputs be assigned to different parameters simultaneously?
Yes—Output 1 and Output 2 may be independently assigned to fluoride concentration, chloride concentration, ammonia concentration, or temperature—enabling parallel integration with multiple control systems or data historians.
Is the data logger compliant with audit trail requirements for regulated laboratories?
The embedded logger records operator actions, calibration events, and measurement values with immutable timestamps; while not inherently 21 CFR Part 11-compliant out-of-the-box, it provides the foundational traceability required for internal GLP audits and serves as a validated data source when paired with qualified IT infrastructure.