Empowering Scientific Discovery

Beijing Furbon Technology Development Co., Ltd.

Categories
  • All
  • Favorite
  • Popular
  • Most rated
Added to wishlistRemoved from wishlist 0
Add to compare
BrandKoehler
OriginUSA
ModelK94190
ComplianceASTM D5001
Test PrincipleBall-on-Cylinder Lubricity Evaluation (BOCLE)
Sample Volume50 ± 1.0 mL
Test Temperature25 ± 1 °C
Regulated Air RH10 ± 0.2 % at 25 °C
Dry/Air Flow Rates0.5 L/min (through sample), 3.8 L/min (over test surface)
Load1000 g (±1 g)
Cylinder Speed240 ± 0.5 rpm
Test Duration30 ± 0.1 min
Test RingSAE 8720 steel, Ø49.10–49.25 mm × 13.0–13.11 mm, hardness 58–62 Rc, Rq 0.51–0.71 µm
Test BallAISI 52100 steel, Ø12.7 mm, hardness 64–66 Rc
Power Supply230 V AC, single-phase, 5 A
Added to wishlistRemoved from wishlist 0
Add to compare
BrandKoehler
OriginUSA
ModelK33900
ComplianceASTM D381, IP 131, ISO 6246, DIN 51784, FTM 791-3302, NF M07004
Configuration6-position aluminum block bath with integrated steam superheater
Temperature RangeUp to 246 °C (475 °F)
Temperature Stability±0.5 °C
Heating Power0–3000 W
Steam SuperheaterStainless steel chamber, solid-state temperature control (0–290 °C), dry saturated steam delivery
Air/Steam Flow ControlDigital mass flow meter + pressure regulator with gauge
SafetyOver-temperature cut-off circuit, insulated aluminum block, dual LED display (°C/°F), PID microprocessor temperature control
Electrical220–240 V, 50/60 Hz, single-phase, 30 A
Dimensions (W×D×H)53.34 × 52.07 × 71.12 cm
Net Weight104.3 kg