BrandLeici
OriginShanghai, China
ModelJPB-607A
Instrument TypePortable
Measurement PrincipleElectrochemical (Clark-type Polarographic Probe)
Measurement RangeDissolved Oxygen: (0.0–20.0) mg/L
Temperature(0.0–40.0) °C
Accuracy±0.30 mg/L (DO), ±0.5 °C (Temperature)
Repeatability±0.1 mg/L
Stability±0.10 mg/L over 3 h
Response Time≤45 s to 90% of final value at 20 °C
Zero Oxygen Error≤0.10 mg/L
Power SupplyTwo AA alkaline batteries
Display3.5-digit LCD with optional blue backlight
CalibrationZero-oxygen and saturated-air (full-scale) calibration supported
Temperature CompensationAutomatic, integrated thermistor
Added to wishlistRemoved from wishlist 0
Add to compare
BrandLeici
OriginShanghai, China
ModelJPBJ-608
Instrument TypePortable
Measurement PrincipleElectrochemical (Polarographic Membrane-Type Probe)
Range(0.00–19.99) mg/L DO
Accuracy±0.30 mg/L
Repeatability±0.10 mg/L
Temperature Range(0.0–40.0) °C
Temp. Accuracy±0.5 °C
Saturation Range(0.0–199.9) %
Saturation Accuracy±10 %FS
Zero Error≤0.10 mg/L
Stability±0.10 mg/L / 3 h
Response Time (90%)≤45 s at 20 °C
Data Storage Capacity500 sets
Protection RatingIP65
InterfaceRS-232
Power Supply4 × AA alkaline batteries
Added to wishlistRemoved from wishlist 0
Add to compare
BrandLeici
OriginShanghai, China
Manufacturer TypeAuthorized Distributor
Product OriginDomestic (China)
ModelCOD-571
Instrument TypeBenchtop Laboratory Analyzer
Measurement PrincipleRapid Digestion Method (Closed-Tube, Dichromate Oxidation)
Measurement Range0–1500 mg/L COD
Accuracy±8% of reading ±1 mg/L (for 0–150 mg/L)
Repeatability≤3% RSD
Digestion Time0–120 min (programmable)
Batch Capacity21 samples per run
Digestion Temperature Range50–150 °C (control accuracy: ±5 °C below 150 °C)
Timing Accuracy±5 min at 120 min setting
Stability±2 mg/L drift over 15 min
Operating Environment5–40 °C, RH ≤85%, AC 220 V ±10%, 50 Hz ±1 Hz
Safety FeaturesOvertemperature alarm, auto-shutdown post-digestion
Data OutputIntegrated thermal printer (PP40 type), direct COD concentration readout
Added to wishlistRemoved from wishlist 0
Add to compare
BrandLeici
OriginShanghai, China
ModelPXSJ-216F
Instrument TypeBenchtop
Measurement RangepH (−2.000 to 20.000)
Detection Range0–19990 (units: µg/L, mg/L, g/L, mol/L, mmol/L)
Response Time10 s
Accuracy±0.002 pH/pX
Temperature Range−5.0 to 110.0 °C