Thermo Fisher
Filter
Showing 121–150 of 995 results
| Brand | Thermo Fisher |
|---|---|
| Origin | USA |
| Model | ECtestr 11+ |
| Conductivity Range | 0–200.0 µS/cm, 0–2000 µS/cm, 0–20.00 mS/cm |
| Resolution | 0.1 µS/cm, 1 µS/cm, 0.01 mS/cm |
| Accuracy | ±1% of full scale |
| Calibration Points | 3-point (auto/manual) |
| Temperature Range | 0–50.0 °C |
| Temperature Resolution | 0.1 °C |
| Temperature Accuracy | ±0.5 °C |
| TDS Conversion Factor | Adjustable (0.40–1.00) |
| Sensor Material | SS316 stainless steel with Valox protective housing |
| Ingress Protection | IP67 |
| Power | Replaceable batteries |
| Battery Status Retention | Calibration settings retained after battery replacement |
| Brand | Thermo Fisher |
|---|---|
| Origin | USA |
| Manufacturer | Thermo Fisher Scientific |
| Product Type | Imported |
| Model | EPD MK2 |
| Instrument Category | Dose/Dose Rate Meter |
| Form Factor | Portable |
| Measured Radiation Types | X-ray, γ-ray, and β-particle radiation |
| Detector Type | Multi-stage solid-state detector |
| Measurement Quantities | Ambient dose equivalent H*(10), personal dose equivalent Hp(10) (deep/body), and Hp(0.07) (shallow/skin) |
| Compliance Standards | IEC 61283, IEC 61526, ANSI N13.11, ANSI N13.27, ANSI N42.20 |
| Data Retention | 10 years without battery power |
| Power Supply | Two AA 1.5 V alkaline batteries (8 weeks continuous operation) or one 3.6 V lithium battery (5 months continuous operation) |
| Communication Interface | High-speed infrared (IR) port |
| Alarm Capability | Configurable dose and dose rate thresholds with audible alert |
| Memory Security | Password-protected dose log storage |
| Certification Status | Only electronic personal dosimeter to pass all IAEA 2007 performance tests |
| Brand | Thermo Fisher |
|---|---|
| Origin | USA |
| Manufacturer | Thermo Fisher Scientific |
| Product Type | Imported |
| Model | EPD-N2 |
| Instrument Category | Dose Rate Meter / Personal Dosimeter |
| Form Factor | Portable |
| Radiation Types Measured | Gamma and Neutron |
| Detector Technology | Multi-stage Solid-State Detector |
| Energy Range | 25 keV – 10 MeV (photons), Thermal neutrons – 15 MeV (neutrons) |
| Dose Quantity Displayed | Ambient and Personal Dose Equivalent Hp(10) |
| Compliance | IEC 61526 Ed. 4.0 (2018), ANSI N42.20, ISO 15265:2022 |
| Brand | Thermo Fisher |
|---|---|
| Origin | Singapore |
| Manufacturer | Thermo Fisher Scientific |
| Product Type | Imported Instrument |
| Model | C401 |
| Price | Upon Request |
| Chlorine Range | 0–1.99 ppm / 2.0–6.0 ppm |
| Chlorine Resolution | 0.01 ppm / 0.1 ppm |
| Chlorine Accuracy | ±0.02 ppm / ±0.2 ppm |
| pH Range | 5.9–8.2 |
| pH Resolution | 0.1 |
| pH Accuracy | ±0.1 |
| Cyanuric Acid Range | 5–90 ppm |
| Cyanuric Acid Resolution | 1 ppm |
| Cyanuric Acid Accuracy | ±4 ppm |
| Light Source | LED (525 nm) |
| Calibration | Single-point user-selectable per test |
| Operating Temperature | 0–50 °C |
| Power Supply | 4 × 1.5 V AAA batteries |
| Battery Life | >3000 measurements |
| IP Rating | IP67 |
| Dimensions (Instrument) | 15.5 × 6.8 × 4.6 cm |
| Weight (Instrument) | 200 g |
| Packaging Dimensions | 35.5 × 16.5 × 10.5 cm |
| Package Weight | 1170 g |
| Brand | Thermo Fisher |
|---|---|
| Origin | Singapore |
| Manufacturer | Thermo Fisher Scientific |
| Product Type | Imported Benchtop Conductivity Meter |
| Model | Eutech CON700 |
| Measurement Parameters | Conductivity, TDS, Temperature |
| Conductivity Range | 0 μS/cm to 200.0 mS/cm |
| TDS Range | 0–100 ppt (TDS factor 0.5) |
| Temperature Range | 0.0–100.0 °C |
| Accuracy | ±1% of full scale |
| Resolution | 0.01 / 0.1 / 1 μS/cm |
| Temperature Accuracy | ±0.5 °C |
| Temperature Resolution | 0.1 °C |
| Calibration Points | Up to 5-point conductivity/TDS calibration |
| Data Storage | 100 measurement records |
| Electrode Compatibility | 2-cell or 4-cell conductivity cells (K = 0.1, 1.0, 10.0 cm⁻¹) |
| Auto-Endpoint Detection | Yes |
| Temperature Compensation | Automatic or manual (linear, 0.00–10.00%/°C) |
| Reference Temperature | 15.0–30.0 °C |
| Display | Dual-line LCD (5.6 × 7.5 cm), large-font numeric readout |
| Power Supply | 9 V DC regulated adapter (100–240 VAC, SMPS) |
| Dimensions (H×W×D) | 17.5 × 15.5 × 6.9 cm |
| Weight | 650 g (instrument only) |
| Brand | Thermo Fisher |
|---|---|
| Origin | Singapore |
| Manufacturer | Thermo Fisher Scientific |
| Product Type | Imported Instrument |
| Model | Ion 6+ |
| Price | Upon Request |
| Brand | Thermo Fisher |
|---|---|
| Origin | Singapore |
| Manufacturer | Eutech Instruments (a Thermo Fisher Scientific brand) |
| Model | pH 6+ |
| Type | Portable Electrochemical Meter |
| Measurement Modes | pH, ORP (mV), Relative mV (RmV), Temperature |
| pH Range | 0.00–14.00 |
| pH Resolution | 0.01 |
| pH Accuracy | ±0.01 |
| Calibration Points | Up to 3-point auto-buffer recognition (USA, NIST, Pure Water) |
| ORP/RmV Range | ±1000.0 mV |
| ORP/RmV Resolution | 0.1 mV (±199.9 mV), 1 mV (>±199.9 mV) |
| ORP Accuracy | ±0.2 mV (±199.9 mV), ±2 mV (>±199.9 mV) ±1 LSD |
| Temperature Range | 0.0–100.0 °C |
| Temperature Resolution | 0.1 °C |
| Temperature Accuracy | ±0.5 °C |
| Temperature Compensation | Automatic or Manual (0.0–100.0 °C) |
| Display | Dual-line LCD (4.5 × 2.3 cm) |
| Power | 4 × 1.5 V AAA alkaline batteries |
| Battery Life | >500 hours |
| Auto-Off | 20 min inactivity |
| Dimensions (Instrument) | 15.7 × 8.5 × 4.2 cm |
| Weight (Instrument) | 255 g |
| Input Connectors | BNC + 2.5 mm phono sockets |
| Operating Environment | 0–50 °C, ≤85% RH non-condensing |
| Enclosure Rating | IP67-rated rubber boot included |
| Brand | Thermo Fisher |
|---|---|
| Origin | USA |
| Model | Salt6+ |
| Measurement Principle | Conductivity-based salinity determination with automatic temperature compensation (ATC) |
| Salinity Range | 1.0–50.0 ppt / 0.1–5.00% |
| Salinity Resolution | 0.1 ppt / 0.01% |
| Salinity Accuracy | ±1% of full scale |
| Temperature Range | −10.0 to 110.0 °C |
| Temperature Resolution | 0.1 °C |
| Temperature Accuracy | ±0.5 °C |
| Temperature Compensation | ATC/MTC (0–50 °C) |
| Adjustable Temperature Coefficient | 0.0–3.0 %/°C |
| Reference Temperatures | 20.0 °C or 25.0 °C (selectable) |
| Cell Constant | 1.0 cm⁻¹ |
| Data Hold Function | Yes |
| Auto Power-Off | After 20 min of inactivity |
| Power Supply | 4 × 1.5 V AAA batteries |
| Dimensions (Instrument) | 14 × 17 × 3.5 cm |
| Weight (Instrument) | 200 g |
| Dimensions (Packaging) | 24 × 17 × 8 cm |
| Weight (Packaging) | 550 g |
| Compliance | Designed for GLP-compliant field and lab use |
| Brand | Thermo Fisher |
|---|---|
| Origin | Taiwan |
| Manufacturer Type | OEM Manufacturer |
| Country of Origin | China (Domestically Assembled) |
| Model | EV-1000B |
| Detection Principle | Gas Chromatography–Flame Ionization Detection (GC-FID) |
| Instrument Type | Online Continuous Analyzer |
| Configuration | Multi-range Composite VOC Analyzer |
| Measurement Ranges | 0–50 ppm / 0–500 ppm / 0–5000 ppm / 0–5% / 0–100% (auto-ranging) |
| Resolution | As specified in technical manual |
| Accuracy | ±1% F.S. or ±0.1 ppm (whichever is tighter) |
| Response Time | Application-specific (45 s to 60 min per cycle) |
| Brand | Thermo Fisher |
|---|---|
| Origin | USA |
| Manufacturer Type | Original Equipment Manufacturer (OEM) |
| Product Origin | Imported |
| Model | Evolution Pro |
| Instrument Configuration | Double-beam |
| Detector Type | Photodiode Array (PDA) |
| Wavelength Range | 190–1100 nm |
| Wavelength Scanning | Motorized Auto-Scanning |
| Spectral Bandwidth | Selectable 0.5, 1.0, 1.5, 2.0, or 4.0 nm |
| Wavelength Accuracy | ±0.20 nm at 546.07 nm (Hg emission line) |
| Wavelength Repeatability | Standard deviation <0.05 nm over 10 consecutive measurements |
| Stray Light | ≤0.4% T at 198 nm (KCl), ≤0.032% T at 220 nm (NaI), <0.01% T at 340 nm (NaNO₂) |
| Brand | Thermo Fisher |
|---|---|
| Origin | Shanghai, China |
| Manufacturer Type | Authorized Distributor |
| Origin Category | Domestically Manufactured |
| Model | Evolution™ Pro UV-Vis Spectrophotometer (Domestically Manufactured) |
| Pricing | Available Upon Request |
| Brand | Thermo Fisher |
|---|---|
| Origin | Shanghai, China |
| Manufacturer Type | Authorized Distributor |
| Product Category | Domestic (China-manufactured) |
| Model | EVOS M3000 |
| Pricing | Upon Request |
| Camera | 3.2 MP CMOS (2064 × 1536), 8-bit output |
| Display | 10.1" LCD touchscreen (1920 × 1080) |
| Dimensions (W × D × H) | 11.7" × 12.7" × 19.1" |
| Weight | 8.3 kg |
| Objective Turret | 4-position |
| Magnification Range | 1.25× to 60× |
| Light Source | LED-based |
| Imaging Modes | Brightfield, Phase Contrast, Color, Dual-Channel Fluorescence |
| Onboard Storage | 128 GB SSD |
| RAM | 8 GB |
| Connectivity | Wi-Fi (USB adapter included), USB 3.0, Thermo Fisher Connect & OneDrive™ integration |
| Compliance | For Research Use Only (RUO), ISO 13485-manufactured, compatible with GLP/GMP lab environments |
| Brand | Thermo Fisher |
|---|---|
| Origin | USA |
| Model | EVOS M5000 |
| Light Source | Adjustable LED Cube (Lifespan > 50,000 hrs per cube) |
| Imaging Modes | Fluorescence (4-channel simultaneous acquisition), Color Brightfield, Z-Stack, Time-Lapse, Maximum Intensity Projection (MIP), 3D Reconstruction |
| Focus Control | Motorized Auto-Focus with Software-Driven Z-Axis Control |
| Environmental Control | Optional Live-Cell Chamber (Temperature, Humidity, CO₂/O₂/N₂ Regulation) |
| Compliance | Designed for GLP/GMP-aligned workflows |
| Brand | Thermo Fisher |
|---|---|
| Origin | USA |
| Model | AMF7000 |
| Optical System | Infinity-corrected, RMS-threaded objectives (45 mm parfocal distance) |
| Illumination | 4-position LED fluorescence cube turret + brightfield channel |
| Objectives | 5-position motorized turret |
| Cameras | Dual 3.2 MP CMOS (monochrome & color, 2048 × 1536, 3.45 µm pixel size) |
| Stage | Motorized X/Y scanning stage (120 mm × 80 mm travel, sub-micron resolution) |
| Focus | Automated autofocus with sub-micron precision |
| Display | 27-inch 4K monitor (3840 × 2160) |
| Software Platform | EVOS Acquisition & Analysis Suite (standard) |
| Compliance | Designed for GLP/GMP-aligned workflows |
| Brand | Thermo Fisher |
|---|---|
| Origin | Shanghai, China |
| Manufacturer Type | Authorized Distributor |
| Product Category | Domestic (China-manufactured) |
| Model | EVOS M7000 |
| Pricing | Upon Request |
| Brand | Thermo Fisher |
|---|---|
| Origin | USA |
| Manufacturer Authorization | Authorized Distributor |
| Import Status | Imported |
| Model | EVOS XL / EVOS XL Core |
| Pricing | Upon Request |
| Brand | Thermo Fisher |
|---|---|
| Origin | USA |
| Model | AMF3000 |
| Dimensions | 30.5 × 48.3 × 33.0 cm (W × D × H) |
| Display | 10.3-inch capacitive LCD touchscreen |
| Camera | 2064 × 1536-pixel (3.2 MP) color CMOS sensor, 8-bit depth |
| Image formats | TIFF, PNG, JPG, MP4 |
| Storage | 128 GB onboard SSD |
| Connectivity | Wi-Fi 5 (802.11ac), USB 3.0, Ethernet |
| Objective turret | 4-position, compatible with 1.25×–60× objectives (achromat/fluorite/apochromat) |
| Fluorescence | Dual-channel via interchangeable LED light cubes |
| Contrast modes | Brightfield, phase contrast, color, two-color fluorescence |
| Confluency analysis | Real-time, ML-based, pre-trained on multiple adherent cell lines |
| Cleanroom compatibility | Yes (UV/H₂O₂ disinfection compliant) |
| Optional accessories | Mechanical stage (Cat. No. AMEP5011) |
| Brand | Thermo Fisher |
|---|---|
| Origin | Germany |
| Manufacturer Type | Original Equipment Manufacturer (OEM) |
| Origin Category | Imported |
| Model | Exactive™ GC Orbitrap™ GC-MS |
| Instrument Type | Orbitrap Mass Spectrometer |
| Chromatography Type | Gas Chromatography |
| Mass Range | Not Specified |
| Resolution | 60,000 (FWHM) at m/z 200 |
| Sensitivity | <6 fg OFN (Octafluoronaphthalene) |
| Scan Speed | Not Specified |
| Ionization Modes | Electron Ionization (EI) / Chemical Ionization (CI) |
| Ion Source | Thermo Scientific™ ExtractaBrite™ with Vacuum Lock |
| Brand | Thermo Fisher |
|---|---|
| Origin | Jiangsu, China |
| Manufacturer Type | OEM Manufacturer |
| Region of Origin | Domestic (China) |
| Model | EXTREVA™ ASE™ Accelerated Solvent Extractor |
| Instrument Type | Solid-Liquid Extraction (not Liquid-Liquid) |
| Temperature Range | 40–200 °C |
| Pressure Range | 200 bar (corrected from erroneous "200 Pa") |
| Extraction Cell Volumes | 1, 5, 10, 22, 34, 66, and 100 mL |
| Extraction Time | 2–300 min |
| Dimensions (W×D×H) | 100 × 65.2 × 86.6 cm |
| Sample Capacity | Dual trays, 16 positions total |
| Brand | Thermo Fisher |
|---|---|
| Origin | USA |
| Model | F6000-60 / F30400-60 |
| Temperature Range | 100–975 °C |
| Chamber Volume | 15 L or 45 L |
| Air Flow Control | 0–80 L/min (adjustable via front-mounted flowmeter) |
| Temperature Stability at 750 °C | ±3 °C |
| Heating Rate | 8 °C/min (≤500 °C), 6 °C/min (500–750 °C) |
| Compliance | ASTM D3174 |
| Gas Inlet Ports | 1 × 0.95 cm thermocouple port, 1 × 0.635 cm or 0.876 cm soft-hose barb for inert gas supply |
| Control Options | B1 (single-setpoint), C1 (8-segment single program), D1 (16-segment × 4 programs, RS232 interface) |
| Brand | Thermo Fisher |
|---|---|
| Origin | USA |
| Manufacturer | Thermo Fisher Scientific |
| Product Type | Benchtop Box-Type Muffle Furnace |
| Models | FB 1300 / FB 1400 |
| Max Temperature | 1100°C |
| Temperature Uniformity (at 1000°C) | ±7.8°C (FB 1300) / ±5.0°C (FB 1400) |
| Temperature Stability (at 1000°C) | ±0.3°C (FB 1300) / ±0.5°C (FB 1400) |
| Control System | Digital Single-Setpoint (Class A1) |
| Heating Method | Resistive Wire (Ni-Cr Alloy) |
| Chamber Dimensions (W×H×D, mm) | 103×98×130 (FB 1300) / 127×108×152 (FB 1400) |
| External Dimensions (W×H×D, mm) | 230×360×216 / 330* (FB 1300, door open) / 250×370×282 / 400* (FB 1400, door open) |
| Chamber Volume | 1.3 L (FB 1300) / 2.1 L (FB 1400) |
| Power Rating | 1060 W |
| Input Voltage/Frequency | 240 V, 50/60 Hz |
| Current Draw | 4.4 A (FB 1300) / 6.3 A (FB 1400) |
| Ramp Rate to Max Temp | 6–12°C/min |
| Thermal Insulation | High-Density Ceramic Fiber |
| Safety Features | Door-Interlocked Power Cut-off, Thermocouple Fault Protection, Rear 9.5-mm Process Port |
| Net Weight | 9 kg (FB 1300) / 12.7 kg (FB 1400) |
| Brand | Thermo Fisher |
|---|---|
| Origin | USA |
| Model | FB1300 / FB1400 |
| Temperature Range | 100–1100 °C |
| Chamber Volume | 1.3 L / 2.1 L |
| Insulation | Ceramic Fiber |
| Heating Element Layout | Top + Dual Side Mounting |
| Door Operation | Downward-Opening Hinged Door |
| Safety Interlock | Integrated Door-Activated Power Cut-off |
| Thermocouple Fault Protection | Automatic Heater Disengagement |
| Rear Port | 0.95 cm Diameter for External Sensor Integration |
| Control System | Digital Single-Setpoint Controller with Dual-Mode Display (Actual/Setpoint Temperature) |
| Brand | Thermo Fisher |
|---|---|
| Origin | USA |
| Manufacturer | Thermo Fisher Scientific |
| Product Type | Imported |
| Model | FH 40 G NBR |
| Instrument Category | Dose Rate Meter |
| Form Factor | Portable |
| Radiation Type | X-ray and Gamma (γ) |
| Detector Architecture | Dual-detector (NaI(Tl) scintillator + plastic scintillator) with NBR (Natural Background Reduction) patented technology |
| Measurement Range | 1 nSv/h – 100 µSv/h |
| Sensitivity | 2000 s⁻¹/µSv/h |
| Energy Response Range (Probe FHZ 672E-10) | 48 keV – 6 MeV |
| Detection Limit | <20% of typical natural background dose rate |
| Host Unit (FH 40 GL-10) | Energy Range: 30 keV – 4.4 MeV |
| Host Measurement Range | 10 nSv/h – 100 mSv/h |
| Host Sensitivity | 2 s⁻¹/µSv/h |
| Host Weight (excl. batteries) | 410 g |
| Battery Life | >250 h (standalone), >30 h (with FHZ 672E-10 connected) |
| Data Handling | Onboard storage, USB interface, PC configuration & data download capability |
| Signal Processing | ADF (Advanced Digital Filtering) adaptive algorithm |
| Brand | Thermo Fisher |
|---|---|
| Origin | Germany |
| Manufacturer | Thermo Fisher Scientific |
| Product Type | Portable Dose Rate and Integrated Dose Meter |
| Detector Compatibility | Interchangeable (NaI(Tl), He-3 neutron, GM, scintillation, or proportional detectors) |
| Measurement Range Extension | Up to 4 m via modular fiberglass telescopic pole |
| Weight | 2.6 kg |
| Compliance | Meets IEC 60846-1:2014, IEC 61526:2010, and EURATOM Basic Safety Standards Directive 2013/59 |
| Software Interface | USB and Bluetooth connectivity for real-time data logging and remote monitoring via Thermo Fisher RadEye Suite |
| Brand | Thermo Fisher |
|---|---|
| Origin | USA |
| Manufacturer | Thermo Fisher Scientific |
| Product Type | Imported |
| Model | FHT 1377 PackEye |
| Instrument Category | Dose Rate Meter |
| Portability | Portable |
| Measured Radiation Types | Gamma and Neutron |
| Detector Configuration | Dual Plastic Scintillator + Two He-3 Proportional Counters |
| Dose Rate Range | 50 nSv/h – 2 mSv/h |
| Sensitivity | 50 nSv/h |
| Power Consumption | 350 mW |
| Battery Life | ~70 h (7.2 V, 3600 mAh rechargeable Li-ion) |
| Dimensions | 58 cm (H) × 30 cm (W) × 18 cm (D) |
| Weight | 6 kg |
| Data Update Interval | 100 ms |
| LED Indication | 10-level gamma / 10-level neutron |
| Communication | Bluetooth + GPS interface |
| Compliance | IAEA-recommended system for orphan source search |
| Key Technology | Patented Natural Background Rejection (NBR), Digital Filtering, Low-Power Embedded Electronics |
| Brand | Thermo Fisher |
|---|---|
| Origin | Germany |
| Manufacturer | Thermo Fisher Scientific |
| Product Type | Portable Dose Rate Meter |
| Model | FHT 40NBR |
| Detector Configuration | Dual-detector (NaI(Tl) scintillation crystal + plastic scintillator) |
| Measurement Principle | Gamma photon energy discrimination via pulse-height analysis and NBR (Natural Background Rejection) algorithm |
| Compliance | Designed for field-deployable radiological survey per IEC 60846-1:2014, ANSI N42.33, and EURATOM Directive 2013/59/Euratom |
| Data Interface | USB and Bluetooth® 4.0 |
| Storage Capacity | Up to 10,000 timestamped dose rate records |
| Operating Temperature | –10 °C to +50 °C |
| IP Rating | IP67 |
| Battery Life | >100 hours (alkaline) or >200 hours (rechargeable Li-ion) |
| Brand | Thermo Fisher |
|---|---|
| Origin | USA |
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | Finnigan Surveyor |
| Product Category | Chromatography |
| Operational Age | 3–4 years |
| Warranty | 6 months |
| Instrument Model | Finnigan Surveyor |
| Instrument Usage Duration | 3 years |
| Brand | Thermo Fisher |
|---|---|
| Origin | USA |
| Model | Finnigan Surveyor Plus |
| Category | Liquid Chromatograph |
| Usage Duration | 3–4 years |
| Warranty | 6 months |
| Compliance | FDA 21 CFR Part 11 (via ChromQuest CDS) |
| Detector Configuration | UV/Vis, Refractive Index, PDA with LightPipe™ technology |
| Pump Type | Quaternary LC pump |
| Integrated Column Oven | Yes |
| Software | ChromQuest Chromatography Data System |
| Brand | Thermo Fisher |
|---|---|
| Origin | USA |
| Manufacturer Type | Authorized Distributor |
| Product Origin | Imported |
| Model | 4640040 |
| Instrument Type | Single-Channel Pipette |
| Automation Level | Manual |
| Channels | 1 |
| Volume Range | 10–100 µL |
| Accuracy | ±3.0% to ±0.8% (volume-dependent, per ISO 8655) |
| Brand | Thermo Fisher |
|---|---|
| Origin | Italy |
| Model | FlashSmart CHN |
| Measurement Elements | C, H, N (expandable to CHNSO, CHNS, NCS, NC, N, S, O) |
| Analysis Time | 5–10 min (sample-weight and matrix dependent) |
| Detection Limit | <40 ppm |
| Accuracy | ≤0.05% |
| Sample Volume | up to 500 µL |
| Detectors | Thermal Conductivity Detector (TCD), Flame Photometric Detector (FPD), Infrared Detector (IR), Isotope Ratio Mass Spectrometer (IRMS) interface |
| Combustion Temperature | up to 1800 °C |