Integrity Tester
Filter
Showing 31–54 of 54 results
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Model | MFT-900 |
| Detection Principle | Vacuum Decay / Differential Pressure |
| Compliance | USP <1207>, ASTM F2338-22, FDA Guidance on Container Closure Integrity Testing (CCIT) |
| Test Range | −0.1 to 1.0 MPa (selectable), Differential Pressure Sensor: ±2 kPa, Accuracy ≤0.5% F.S., Resolution: 0.1 Pa |
| Minimum Detectable Leak Equivalent Hole Diameter | 2 µm |
| Touchscreen | 10-inch Color TFT LCD |
| Power | AC 100–240 V, 50/60 Hz, ≤200 W |
| Audit Trail Retention | ≥5 years |
| User Access Control | 4-tier Role-Based Authentication |
| Data Export | USB drive, CSV format |
| Communication Interfaces | Ethernet, RS485, 4–20 mA analog output, 12 V alarm signal |
| Optional Accessories | Certified Standard Leak Reference Vials (NIST-traceable), Custom Test Chamber Kits, Method Validation Support Services |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Model | MFT-900 |
| Detection Principle | Vacuum Decay / Differential Pressure Method |
| Compliance | USP <1207>, ASTM F2338-22, FDA Guidance on Container Closure Integrity Testing (CCIT) |
| Automation Level | Semi-Automatic |
| Detection Sensitivity | Equivalent to ≥2 µm defect aperture |
| Pressure Range | −0.1 to 1.0 MPa (selectable ranges: −0.1–0.2 MPa / −0.1–0.7 MPa / −0.1–1.0 MPa) |
| Differential Pressure Sensor | ±2 kPa, Accuracy ≤0.5% FS, Resolution: 0.1 Pa |
| Test Pressure Sensor Accuracy | ≤±1% FS |
| Display | 10-inch color capacitive touchscreen |
| Data Storage | ≥5 years audit-trail-enabled event logs & test records |
| User Management | 4-tier role-based access control (RBAC), password-authenticated login |
| Regulatory Support | Full 21 CFR Part 11 compliance (electronic signatures, audit trail, data integrity) |
| Interface Options | Ethernet, USB, RS485, 4–20 mA analog output, 12 V alarm signal |
| Power Supply | AC 100–240 V, 50/60 Hz, ≤200 W |
| Data Export | USB flash drive backup |
| Optional Peripherals | External thermal printer |
| Brand | Neuronbc |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Authorized Distributor |
| Country of Origin | China |
| Model | NBC-1002 |
| Price | USD 28,000 (FOB Beijing) |
| Excitation Wavelength | 785 nm |
| Laser Power | ≤500 mW (adjustable) |
| Spectral Range | 200–3200 cm⁻¹ |
| Spectral Resolution | ≤8 cm⁻¹ |
| Detector | TE-cooled CCD |
| Battery Life | ≥6 hours (continuous operation) |
| Connectivity | Wi-Fi 5 (802.11ac), Bluetooth 5.0, USB-C |
| Data Storage | Internal 64 GB + cloud-sync capable |
| Compliance | Meets ISO/IEC 17025 calibration traceability requirements |
| Brand | NeuronBC |
|---|---|
| Model | V4.0 |
| Display | 5.7″ backlit digital LCD |
| Interface | Chinese-language GUI |
| Data Visualization | Real-time test curve and numerical display |
| Processing | Dual-core embedded controller |
| Software | PC-based upper-computer software with audit trail capability |
| User Management | Multi-level password-protected access control |
| Safety Protection | Automatic pressure-limiting and overpressure cutoff for filter safeguard |
| Test Methods | Forward flow (diffusion), bubble point, and water intrusion (for hydrophobic membranes) |
| Compliance Support | Configurable report generation aligned with FDA 21 CFR Part 11 and EU Annex 11 requirements |
| Reporting | Optional integrated thermal printer interface |
| Membrane Compatibility | Flat-sheet membranes, cartridge filters, and disc filters |
| Origin | Beijing, China |
| Price | USD 5,600 (FOB) |
| Origin | Beijing, China |
|---|---|
| Manufacturer Type | Distributor |
| Origin Category | Domestic |
| Model | V4.0 |
| Price | USD 5,300 (FOB) |
| Power Supply | 100–240 VAC, 50 Hz, 120 W |
| Operating Pressure Range | 100–8000 mbar |
| Test Accuracy | Bubble Point ±50 mbar |
| Bubble Point Range | 590–6900 mbar |
| Diffusion Flow Range | 1–180 mL/min |
| Water Intrusion Range | 0.01–30 mL/min |
| Display | 5.7″ monochrome TFT LCD with backlight |
| Data Storage | 500 test records |
| Interface | RS232 serial port |
| Language Support | English & Chinese |
| Operating Modes | Online & Offline |
| Environmental Class | ISO Class B or better |
| Weight | 7.5 kg |
| Dimensions (L×W×H₁×H₂) | 400 × 240 × 270 × 220 mm |
| Relative Humidity | 10–80% RH |
| Ambient Temperature | +5 °C to +40 °C |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Model | V6.5 Audit-Trail |
| Power Supply | 100–240 VAC, 50 Hz, 120 W |
| Operating Pressure Range | 100–10,000 mbar |
| Temperature Range | +5°C to +40°C |
| Relative Humidity | 10–80% |
| Dimensions (mm) | 400(D) × 240(W) × 270(H1) × 220(H2) |
| Weight | 8 kg |
| Bubble Point Accuracy | ±50 mbar |
| Diffusion Flow Accuracy | ±4% |
| Water Intrusion Accuracy | ±0.01 mL |
| Bubble Point Range | 100–8,000 mbar |
| Diffusion Flow Range | 1–600 mL/min |
| Water Intrusion Range | 0.01–100 mL/min |
| Pre-stored Protocols | 80 |
| Event Log Capacity | ≥16 auditable parameters with timestamp |
| User Access Levels | 4-tier role-based authentication with electronic signature support |
| Compliance | Fully aligned with FDA 21 CFR Part 11, ALCOA+ data integrity principles, and ISO 13485 / ASTM F838-22 / USP <1207> guidance for filter integrity testing |
| Brand | NeuronBC |
|---|---|
| Model | V6.5 |
| Origin | Beijing, China |
| Power Supply | 100–240 VAC, 50 Hz, 120 W |
| Operating Pressure Range | 100–10,000 mbar (145 psi) |
| Environmental Operating Conditions | 5–40 °C, 10–80% RH |
| Dimensions (mm) | 400 (D) × 240 (W) × 270/220 (H1/H2) |
| Weight | 8 kg |
| Bubble Point Accuracy | ±50 mbar |
| Diffusive Flow Accuracy | ±4% |
| Water Intrusion Accuracy | ±0.01 mL |
| Net Volume Test Accuracy | ±4% |
| Bubble Point Range | 100–8,000 mbar |
| Diffusive Flow Range | 1–600 mL/min |
| Water Intrusion Range | 0.01–100 mL/min |
| Pre-stored Protocols | 80 |
| Event Log Capacity | ≥16 auditable events with timestamped traceability |
| User Access Levels | 4-tier role-based authentication with electronic signature support |
| Data Storage | ≥5,000 test records, USB backup of full datasets including real-time curves |
| Display | 7" high-resolution color touchscreen |
| Interface Options | RS232, USB, optional industrial fieldbus (e.g., Modbus TCP, Profibus DP) and analog I/O |
| Language Support | English / Chinese |
| Compliance | FDA 21 CFR Part 11, ALCOA+ principles, Annex 11 (EU GMP), ISO 13408-2, ASTM F838, USP <1207> |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Model | V8.0 |
| Operating Pressure Range | 100–10,000 mbar (150 psi) |
| Pressure Units | mbar, kPa, psi, kgf/cm² |
| Power Supply | 100–240 VAC, 50 Hz, 120 W |
| Environmental Operating Conditions | 5–40 °C, 10–80% RH |
| Enclosure Rating | IP54 |
| Dimensions (L×W×H) | 480 × 300 × 190 mm |
| Weight | 10 kg |
| Compliance | CE Marked, FDA 21 CFR Part 11, GMP-Aligned Audit Trail & Electronic Signature Support |
| Display | 10-inch true-color touchscreen |
| Data Export | USB (raw test data, configuration files, audit logs) |
| Connectivity | RS232, USB |
| Sensor Accuracy | High-precision pressure transducers with low drift and extended pneumatic line compensation up to 100 m |
| Integrated Thermal Printer | Non-ink, particle-free, archival-grade print retention (>10 years) |
| User Management | Four-tier role-based access control with customizable permissions, mandatory electronic signatures, and password-protected login |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Model | WGT-1000 |
| Power Supply | DC 12 V, 20 W |
| Test Accuracy | Detects leaks equivalent to ≥300 µm aperture |
| Test Duration | 3–8 min |
| Pressure Range | 500–3000 Pa |
| Pressure Resolution | 0.1 Pa |
| Noise Level | ≤50 dB(A) |
| Communication Method | Wi-Fi |
| Gas Source | Built-in sanitary air pump |
| Maximum Sealing Pressure | ≤200 kPa |
| Data Storage Capacity | >12,000 test records |
| Glove Identification | RFID or manual input |
| Enclosure Rating | IP65 |
| Weight | 4.2 kg |
| Display | Color touchscreen interface |
| Compliance | FDA 21 CFR Part 11, GMP, ISO 14644-7:2004, GB/T 25915.7–2010 |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Authorized Distributor |
| Product Category | Domestic (China-manufactured) |
| Model | WGT-1200 |
| Price | Upon Request |
| Power Supply | DC 12 V, 20 W |
| Built-in Battery | 12 V Li-ion |
| Detection Accuracy | Minimum Detectable Hole Size: 300 µm |
| Test Duration | 3–8 min |
| Pressure Range | 500–3000 Pa |
| Pressure Resolution | 0.1 Pa |
| Noise Level | ≤50 dB(A) |
| Communication | Wi-Fi |
| Air Source | Integrated Sanitary-Grade Diaphragm Pump |
| Maximum Seal Pressure | ≤200 kPa |
| Data Storage Capacity | >12,000 test records |
| Data Storage Medium | Onboard Flash Memory |
| Glove Identification | RFID or Manual Entry |
| Unit Weight | 4.2 kg |
| Max. Controlled Test Units per Controller | 10 |
| Operating Time | Up to 8 h continuous / >48 h standby |
| User Accounts | 120 max |
| Predefined Test Protocols | 120 max |
| Access Control | Four-Tier Role-Based Permissions |
| Software Compliance | FDA 21 CFR Part 11 (Electronic Records & Signatures), GMP Annex 11, ALCOA+ Principles |
| Origin | Beijing, China |
|---|---|
| Manufacturer Type | Distributor |
| Origin Category | Domestic |
| Model | V4 |
| Price | USD 5,300 (approx.) |
| Power Supply | 100–240 VAC, 50 Hz, 120 W |
| Operating Pressure Range | 100–8000 mbar |
| Environmental Operating Conditions | 5–40 °C, 10–80% RH |
| Dimensions (mm) | 400 D × 240 W × 270 H₁ × 220 H₂ |
| Weight | 7.5 kg |
| Display | 5.7″ monochrome TFT LCD with backlight |
| Interface | RS232 serial port |
| Language Support | English & Chinese |
| Test Modes | Manual Bubble Point, Basic Bubble Point, Enhanced Bubble Point, Diffusion Flow, Water Intrusion |
| Accuracy | Net Volume ±4% |
| Measurement Ranges | Bubble Point 590–6900 mbar |
| Sample Compatibility | Symmetric & asymmetric membranes |
| Data Storage | 500 test records |
| Application Environment Classification | ISO Class B or better |
| Compliance Context | Designed for use in GMP-compliant pharmaceutical manufacturing and QC laboratories |
| Origin | Beijing, China |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Origin Category | Domestic (PRC) |
| Model | V1.2 |
| Power Supply | 100–240 VAC, 50 Hz, 120 W |
| Operating Pressure Range | 100–8000 mbar (max. 115 psi) |
| Bubble Point Accuracy | ±75 mbar |
| Diffusion Flow Accuracy | ±4% |
| Bubble Point Range | 590–6900 mbar |
| Diffusion Flow Range | 1–180 mL/min |
| Environmental Operating Conditions | 5–40 °C, 10–80% RH |
| Dimensions (L×W×H) | 400 × 240 × 270 mm (base) + 220 mm (height with display module) |
| Weight | 7.5 kg |
| Display | 3.2″ monochrome LCD |
| Data Storage | 20 test records |
| Language | Chinese only |
| Test Modes | Bubble Point (fully automated), Diffusion Flow (semi-automated), Hold Pressure (pressure decay) |
| Sample Compatibility | Symmetric & asymmetric membranes |
| Cleanroom Suitability | ISO Class B and above |
| Compliance | Designed to support execution of ASTM F838-22, USP <1207>, EP 2.6.27, and ISO 13408-2 integrity test protocols |
| Origin | Beijing, China |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Origin Category | Domestic (China) |
| Model | 6.0 |
| Price | USD 11,800 (approx.) |
| Power Supply | 100–240 VAC, 50 Hz, 120 W |
| Operating Pressure Range | 100–10,000 mbar (14.5–145 psi) |
| Environmental Operating Conditions | 5–40 °C, 10–80% RH |
| Dimensions (mm) | 400 D × 240 W × 270 H₁ × 220 H₂ |
| Weight | 8 kg |
| Bubble Point Accuracy | ±50 mbar |
| Diffusion Flow Accuracy | ±4% |
| Water Intrusion Accuracy | ±0.01 mL |
| Net Volume Test Accuracy | ±4% |
| Bubble Point Range | 100–8,000 mbar |
| Diffusion Flow Range | 1–600 mL/min |
| Water Intrusion Range | 0.01–100 mL/min |
| Pre-stored Protocols | 80 |
| Audit Trail Events | 16+ logged parameters (user, timestamp, method, pressure, flow, pass/fail status, etc.) |
| Data Storage Capacity | >5,000 test records (unlimited with USB backup) |
| Display | 7" high-resolution color touchscreen |
| Interface Options | RS232, USB, optional industrial bus (e.g., Modbus TCP, Profibus DP) and analog I/O |
| Language Support | English & Chinese |
| Compliance | Fully aligned with USP <1207>, EP 2.6.27, ISO 13408-2, and FDA 21 CFR Part 11 requirements for electronic records and signatures |
| [Brand | All Precision Instrument |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Model | BSQ-19 |
| Power Supply | 170–240 V AC, 50/60 Hz, 100 W |
| Max Operating Pressure | 7999 mbar |
| Minimum Inlet Pressure | 3000 mbar |
| Test Pressure Range | 500–6900 mbar (customizable sensor) |
| Pressure Sensitivity | ±1.0 mbar |
| Bubble Point Accuracy | ±60 mbar |
| Dimensions (W×D×H) | 240 × 320 × 260 mm |
| Display | 5.7–7-inch color TFT touchscreen, 256-color, GUI in English/Chinese |
| Data Storage | 32 GB internal flash memory (≥1,000,000 test records) |
| Test Duration | Bubble Point: 2 ± 2 min |
| Diffusion Flow | 7 ± 2 min |
| Water Intrusion | 7 ± 2 min |
| System Leak Check | 3 ± 1 min |
| Environmental Operating Range | 0–50 °C, 10–80% RH |
| Compliance | Fully supports GMP documentation requirements, audit trail-ready configuration, user-level access control (Operator/Supervisor/Admin), 21 CFR Part 11–compatible data handling with electronic signature support] |
| Brand | Red Sun Instrument |
|---|---|
| Origin | Hebei, China |
| Manufacturer Type | Authorized Distributor |
| Origin Category | Domestic (China) |
| Model | HR-W2400 |
| Pressure Range | 200–5000 Pa (adjustable) |
| Test Duration | 1–10 min (adjustable) |
| Pressure Resolution | 1 Pa |
| Acoustic Noise Level | ≤50 dB(A) |
| Operating Mode | Continuous |
| Bubble Point Accuracy | ±10 mbar |
| Diffusion Flow Resolution | 0.1 mL/min |
| Water Intrusion Flow Resolution | 0.01 mL/min |
| System Sensitivity | ±1.0 mbar |
| Max Test Pressure | 9999 mbar |
| Min Supply Pressure | 100 mbar |
| Gas Supply Requirement | Clean, dry air or nitrogen (4000–6999 mbar) |
| Power Supply | 100–240 V AC, 50/60 Hz, 100 W |
| IP Rating | IP54 |
| Data Storage Capacity | 10,000 test records |
| Display | 10-inch full-color TFT touchscreen |
| Communication Interfaces | RS232, USB, optional industrial bus (e.g., Modbus RTU), wireless module |
| Printer Options | Built-in thermal or dot-matrix printer |
| User Management | Up to 9-level permission hierarchy with electronic signature support |
| Audit Trail | Time-stamped, immutable event log (user login/logout, parameter changes, test execution, result modification, calibration, system errors) |
| Compliance | Fully aligned with FDA 21 CFR Part 11, EU Annex 11, and ISO 13485 |
| Brand | Red Sun Instrument |
|---|---|
| Origin | Hebei, China |
| Model | HR-WZX2600 |
| Price | USD 8,200 (FOB) |
| Application | Pharmaceutical Sterile Filtration Validation |
| Compliance Basis | ASTM F838-22, ISO 13408-2, USP <1207>, EU Annex 1 (2022), FDA Guidance for Industry – Sterile Drug Products (2004) |
| Brand | Red Sun Instrument |
|---|---|
| Origin | Hebei, China |
| Model | HRWZ-2600 |
| Power Supply | 100–240 V AC, 50/60 Hz, 120 W |
| Maximum Operating Pressure | 10,000 mbar (150 psi) |
| Inlet Gas Pressure Range | 100 mbar (1.5 psi) |
| Bubble Point Test Range | 100–8,000 mbar |
| Diffusion Flow Measurement Range | 1–600 mL/min |
| Water Intrusion Range | 0.01–100 mL/min |
| Test Accuracy | Bubble Point ±50 mbar |
| Dimensions (L×W×H₁×H₂) | 400 × 240 × 270 × 220 mm |
| Weight | 8 kg |
| Ambient Operating Conditions | 5–40 °C, 10–80% RH |
| Display | 7-inch high-resolution color touchscreen |
| Connectivity | RS232, USB, 4–20 mA analog output, RS485, 12 V alarm output |
| Data Storage | ≥10,000 test records |
| Pre-stored Methods | 120 |
| User Access Control | Four-tier permission system with username/password login |
| Audit Trail | Full event logging compliant with FDA 21 CFR Part 11 requirements |
| Test Modes | Manual bubble point, standard bubble point, enhanced bubble point, pressure hold, diffusion flow, water intrusion, ultrafiltration module testing |
| Brand | Red Sun Instrument |
|---|---|
| Origin | Hebei, China |
| Manufacturer Type | Authorized Distributor |
| Country of Origin | China |
| Model | HRYQ-2600 |
| Price | ¥72,000 CNY |
| Power Supply | 170–240 V AC, 50/60 Hz, 100 W |
| Max Operating Pressure | 7999 mbar |
| Minimum Inlet Pressure | 3000 mbar |
| Dimensions (L×W×H) | 500 × 200 × 180 mm |
| Test Pressure Range | 500–6900 mbar |
| Pressure Sensitivity | ±1.0 mbar |
| Bubble Point Accuracy | ±60 mbar |
| Operating Temperature | 0–50 °C |
| Relative Humidity | 10–80 % RH |
| Test Methods Supported | Pressure Hold (Diffusion Flow), Bubble Point (Standard & Enhanced), Water Intrusion, Manual Bubble Point |
| Brand | Tailin |
|---|---|
| Origin | Zhejiang, China |
| Manufacturer Type | Direct Manufacturer |
| Product Origin | Domestic (China) |
| Model | HTY-FT223Pro |
| Pricing | Upon Request |
| Forward Flow Range | 0.1–600 mL/min |
| Water Intrusion Range | 0.01–100 mL/min |
| Bubble Point Range | 400–8000 mbar |
| Pressure Hold Range | 100–8000 mbar |
| Forward Flow Accuracy | ±4% or ±0.1 mL/min (whichever is greater) |
| Water Intrusion Accuracy | ±4% or ±0.02 mL/min (whichever is greater) |
| Bubble Point Accuracy | ±50 mbar |
| Pressure Hold Resolution | 1 mbar |
| Dimensions | 360 × 280 × 280 mm |
| Brand | Tailin |
|---|---|
| Origin | Zhejiang, China |
| Manufacturer Type | Direct Manufacturer |
| Product Origin | Domestic (China) |
| Model | HTY-FT224 |
| Pricing | Available Upon Request |
| Forward Flow Range | 0.1–1000 mL/min |
| Water Intrusion Range | 0.01–100 mL/min |
| Bubble Point Range | 100–8000 mbar |
| Pressure Hold Range | 100–8000 mbar |
| Forward Flow Accuracy | ±4% or ±0.1 mL/min (whichever is greater) |
| Water Intrusion Accuracy | ±4% or ±0.02 mL/min (whichever is greater) |
| Bubble Point Accuracy | ±50 mbar |
| Pressure Hold Resolution | 1 mbar |
| Dimensions | 460 mm × 270 mm × 345 mm |
| Brand | Tailin |
|---|---|
| Origin | Zhejiang, China |
| Manufacturer Type | Direct Manufacturer |
| Origin Category | Domestic (China) |
| Model | HTY-LT100P |
| Pricing | Upon Request |
| Test Methods | Vacuum Decay & Pressure Decay |
| Detection Sensitivity | 1–2 µm (rigid containers), 5–10 µm (flexible packaging) |
| Differential Pressure Resolution | 0.01 Pa |
| Operating Temperature Range | 10–30 °C |
| Relative Humidity Limit | <65 % RH |
| Dimensions | 550 mm × 320 mm × 320 mm |
| Throughput | 2–3 samples/min |
| Chamber Configuration | Dual-symmetric, customizable test chambers |
| Brand | Tailin |
|---|---|
| Origin | Zhejiang, China |
| Manufacturer Type | Manufacturer |
| Region of Origin | Domestic (China) |
| Model | HTY-MBP |
| Pricing | Upon Request |
| Bubble Point Range | 400–8000 mbar |
| Pressure Hold Range | 100–8000 mbar |
| Bubble Point Sensitivity | 50 mbar |
| Pressure Hold Sensitivity | 1 mbar |
| Bubble Point Accuracy | ±50 mbar |
| Pressure Hold Accuracy | ±10 mbar |
| Dimensions | 300 mm × 200 mm × 160 mm |
| Communication Interface | Fieldbus (e.g., Profibus, Modbus RTU/TCP), PLC-integrated I/O (Start/Stop/Status/Result), SCADA-ready |
| Brand | Tailin |
|---|---|
| Origin | Zhejiang, China |
| Manufacturer Type | Manufacturer |
| Product Category | Domestic |
| Model | RTP |
| Pricing | Upon Request |
| Brand | TECHLEAD® |
|---|---|
| Manufacturer | Tailin Biotechnology |
| Origin | Zhejiang, China |
| Model | BIT-L01S |
| Type | Microbial Challenge & Dye Penetration Integrity Tester |
| Compliance Design | IP69-rated stainless steel test chamber, steam-sterilizable (125°C), CFR Part 11–ready data logging |
| Application Scope | Parenteral packaging, blister packs, vials, syringes, IV bags, medical device pouches |